500 research outputs found
Fifteen years follow-up of photorefractive keratectomy up to 10 D of myopia: outcomes and analysis of the refractive regression
PURPOSE:
To evaluate outcomes of photorefractive keratectomy up to -10.00 D of myopia and -4.50 of astigmatism and to develop a predictive model for the refractive changes in the long term.
SETTING:
Vissum Corporation and Miguel Hernandez University (Alicante, Spain).
DESIGN:
Retrospective-prospective observational series of cases.
METHODS:
This study included 33 eyes of 33 patients aged 46.79±7.04 years (range 40-57) operated with the VISX 20/20 excimer laser with optical zones of 6 mm. No mitomycin C was used in any of these cases. The minimum follow-up was 15 years. The main outcome measures were: uncorrected and corrected distance visual acuity, manifest refraction and corneal topography. Linear regression models were developed from the observed refractive changes over time.
RESULTS:
Safety and efficacy indexes at 15 years were 1.18 and 0.83, respectively. No statistically significant differences were detected for any keratometric variable during the follow-up (p≥0.103). 15 years after the surgery 54.55% of the eyes were within ±1.00 D of spherical equivalent and 84.85% within ±2.00 D. The uncorrected distance visual acuity at 15 years was 20/25 or better in 60.6% of the eyes and 20/40 or better in 72.73% of the eyes. The correlation between the attempted and the achieved refractions was r=0.948 (p<0.001) at 1 year, and r=0.821 (p<0.001) at 15 years. No corneal ectasia was detected in any case during the follow-up.
CONCLUSIONS:
Photorefractive keratectomy is a safe refractive procedure in the long term within the range of myopia currently considered suitable for its use, although its efficacy decreases with time, especially, in high myopia. The model developed predicts a myopic regression of 2.00 D at 15 years for an ablation depth of 130 µm
Morphogeometric analysis for characterization of keratoconus considering the spatial localization and projection of apex and minimum corneal thickness point
This work evaluates changes in new morphogeometric indices developed considering the position of anterior and posterior corneal apex and minimum corneal thickness (MCT) point in keratoconus. This prospective comparative study included 440 eyes of 440 patients (age, 7–99 years): control (124 eyes) and keratoconus (KC) groups (316 eyes). Tomographic information (Sirius®, Costruzione Strumenti Oftalmici, Italy) was treated with SolidWorks v2013, creating the following morphogeometric parameters: geometric axis–apex line angle (GA–AP), geometric axis–MCT line angle (GA–MCT, apex line–MCT line angle (AP–MCT), and distances between apex and MCT points on the anterior (anterior AP–MCTd) and posterior corneal surface (posterior AP–MCTd). Statistically significant higher values of GA–AP, GA–MCT, AP–MCT and anterior AP–MCTd were found in the keratoconus group (p ≤ 0.001). Moderate significant correlations of corneal aberrations (r ≥ 0.587, p < 0.001) and corneal thickness parameters (r ≤ −0.414, p < 0.001) with GA–AP and AP–MCT were found. Anterior asphericity was found to be significantly correlated with anterior and posterior AP–MCTd (r ≥ 0.430, p < 0.001). Likewise, GA–AP and AP–MCT showed a good diagnostic ability for the detection of keratoconus, with optimal cutoff values of 9.61° (sensitivity 85.5%, specificity 80.3%) and 18.08° (sensitivity 80.5%, specificity 78.7%), respectively. These new morphogeometric indices allow a clinical characterization of the 3-D structural alteration occurring in keratoconus, with less coincidence in the spatial projection of the apex and MCT points of both corneal surfaces. Future studies should confirm the potential impact on the precision of these indices of the variability of posterior corneal surface measurements obtained with Scheimpflug imaging technology.This publication has been carried out in the framework of the Thematic Network for Co-Operative Research in Health (RETICS), reference number RD16/0008/0012, financed by the Carlos III Health Institute–General Subdirection of Networks and Cooperative Investigation Centers (R&D&I National Plan 2013–2016) and the European Regional Development Fund (FEDER). The author David P. Piñero has been supported by the Ministry of Economy, Industry and Competitiveness of Spain within the program Ramón y Cajal, RYC-2016-20471
Procedimiento de reconstrucción de la topografía corneal a partir de datos altímetros o de curvatura
Número de publicación: ES2392619 A1 (12.12.2012) También publicado como: ES2392619 B1 (22.10.2013) Número de Solicitud: Consulta de Expedientes OEPM (C.E.O.) P201000842(08.06.2010)Procedimiento de reconstrucción de la topografía corneal a partir de datos altimétricos o de curvatura. La invención consiste en un método de reconstrucción de la superficie de la cara anterior de la córnea, a partir de los datos medidos en un conjunto discreto de puntos por medio de un topógrafo corneal o equipo equivalente. Se trata de un procedimiento que obtiene una expresión analítica de la superficie, combinando un ajuste por polinomios de Zernike o con esfera de mejor ajuste, con una reconstrucción por funciones de base radial gaussianas. Se logra obtener una descripción detallada de la superficie corneal, permitiendo un diagnóstico más fiable de patologías, o la implementación de tratamientos customizados. Este procedimiento es fácilmente implementable en cualquier topógrafo corneal, tomógrafo de coherencia óptica, equipos de lámpara de hendidura y equivalentes, de los existentes en el mercado, como sustituto del método estándar basado en polinomios de Zernike.Universidad de Almerí
An adaptive algorithm for the cornea modeling from keratometric data
In this paper we describe an adaptive and multi-scale algorithm for the parsimonious t of the corneal surface data that allows to adapt the number of functions used in the reconstruction to the conditions of each cornea. The method implements also a dynamical selection of the parameters and the management of noise. It can be used for the real-time reconstruction of both altimetric data and corneal power maps from the data collected by keratoscopes, such as the Placido rings based topographers, decisive for an early detection of corneal diseases such as keratoconus. Numerical experiments show that the algorithm exhibits a steady exponential error decay, independently of the level of aberration of the cornea. The complexity of each anisotropic gaussian basis functions in the functional representation is the same, but their parameters vary to fit the current scale. This scale is determined only by the residual errors and not by the number of the iteration. Finally, the position and clustering of their centers, as well as the size of the shape parameters, provides an additional spatial information about the regions of higher irregularity. These results are compared with the standard approximation procedures based on the Zernike polynomials expansions
Comparative Analysis of Some Modal Reconstruction Methods of the Shape of the Cornea from Corneal Elevation Data
Purpose: A comparative study of the ability of some modal schemes to reproduce corneal shapes of varying complexity was performed, by using both standard radial polynomials and radial basis functions (RBFs). The hypothesis was that the correct approach in the case of highly irregular corneas should combine several bases.
Methods: Standard approaches of reconstruction by Zernike and other types of radial polynomials were compared with the discrete least-squares fit (LSF) by the RBF in three theoretical surfaces, synthetically generated by computer algorithms in the absence of measurement noise. For the reconstruction by polynomials, the maximal radial order 6 was chosen, which corresponds to the first 28 Zernike polynomials or the first 49 Bhatia-Wolf polynomials. The fit with the RBF was performed by using a regular grid of centers.
Results: The quality of fit was assessed by computing for each surface the mean square errors (MSEs) of the reconstruction by LSF, measured at the same nodes where the heights were collected. Another criterion of the fit quality used was the accuracy in recovery of the Zernike coefficients, especially in the case of incomplete data.
Conclusions: The Zernike (and especially, the Bhatia-Wolf) polynomials constitute a reliable reconstruction method of a nonseverely aberrated surface with a small surface regularity index (SRI). However, they fail to capture small deformations of the anterior surface of a synthetic cornea. The most promising approach is a combined one that balances the robustness of the Zernike fit with the localization of the RBF
Intraocular lenses in age-related macular degeneration
Purpose:
The aim of this work is to review the lenses, assessing their advantages and disadvantages. We describe a total of seven types of intraocular lenses (IOLs) recommended for age-related macular degeneration (AMD).
Methods:
We used the PubMed web platform to search for implantable devices in various stages of AMD. We searched for both prospective and retrospective studies and also case reports.
Results:
Clinical results in AMD patients have been described for a total of seven types of IOLs recommended for AMD: an implantable miniature telescope (IMT), IOL-VIP System, Lipshitz macular implant (LMI), sulcus-implanted Lipshitz macular implant, LMI-SI, Fresnel Prism Intraocular Lens, iolAMD and Scharioth Macula Lens.
Conclusions:
We conclude that to objectively ascertain the effectiveness and safety of these lenses, further independent clinical studies with longer follow-up data are necessary prior to the general use of these optical devices
Radiofrequency Heating of the Cornea: An Engineering Review of Electrodes and Applicators
This paper reviews the different applicators and electrodes employed to create localized heating in the cornea by means of the application of radiofrequency (RF) currents. Thermokeratoplasty (TKP) is probably the best known of these techniques and is based on the principle that heating corneal tissue (particularly the central part of the corneal tissue, i.e. the central stroma) causes collagen to shrink, and hence changes the corneal curvature. Firstly, we point out that TKP techniques are a complex challenge from the engineering point of view, due to the fact that it is necessary to create very localized heating in a precise location (central stroma), within a narrow temperature range (from 58 to 76ºC). Secondly, we describe the different applicator designs (i.e. RF electrodes) proposed and tested to date. This review is planned from a technical point of view, i.e. the technical developments are classified and described taking into consideration technical criteria, such as energy delivery mode (monopolar versus bipolar), thermal conditions (dry versus cooled electrodes), lesion pattern (focal versus circular lesions), and application placement (surface versus intrastromal)
Search for squarks and gluinos in events with isolated leptons, jets and missing transverse momentum at s√=8 TeV with the ATLAS detector
The results of a search for supersymmetry in final states containing at least one isolated lepton (electron or muon), jets and large missing transverse momentum with the ATLAS detector at the Large Hadron Collider are reported. The search is based on proton-proton collision data at a centre-of-mass energy s√=8 TeV collected in 2012, corresponding to an integrated luminosity of 20 fb−1. No significant excess above the Standard Model expectation is observed. Limits are set on supersymmetric particle masses for various supersymmetric models. Depending on the model, the search excludes gluino masses up to 1.32 TeV and squark masses up to 840 GeV. Limits are also set on the parameters of a minimal universal extra dimension model, excluding a compactification radius of 1/R c = 950 GeV for a cut-off scale times radius (ΛR c) of approximately 30
- …