171 research outputs found

    Executive summary of AAPM Report Task Group 113: Guidance for the physics aspects of clinical trials

    Full text link
    The charge of AAPM Task Group 113 is to provide guidance for the physics aspects of clinical trials to minimize variability in planning and dose delivery for external beam trials involving photons and electrons. Several studies have demonstrated the importance of protocol compliance on patient outcome. Minimizing variability for treatments at different centers improves the quality and efficiency of clinical trials. Attention is focused on areas where variability can be minimized through standardization of protocols and processes through all aspects of clinical trials. Recommendations are presented for clinical trial designers, physicists supporting clinical trials at their individual clinics, quality assurance centers, and manufacturers.Peer Reviewedhttps://deepblue.lib.umich.edu/bitstream/2027.42/146453/1/acm212384_am.pdfhttps://deepblue.lib.umich.edu/bitstream/2027.42/146453/2/acm212384.pd

    Tolerance limits and methodologies for IMRT measurement‐based verification QA: Recommendations of AAPM Task Group No. 218

    Full text link
    Peer Reviewedhttps://deepblue.lib.umich.edu/bitstream/2027.42/143649/1/mp12810_am.pdfhttps://deepblue.lib.umich.edu/bitstream/2027.42/143649/2/mp12810.pd

    AAPM Medical Physics Practice Guideline 2.a: Commissioning and quality assurance of X-ray–based image-guided radiotherapy systems

    Get PDF
    The American Association of Physicists in Medicine (AAPM) is a nonprofit professional society whose primary purposes are to advance the science, education, and professional practice of medical physics. The AAPM has more than 8,000 members and is the principal organization of medical physicists in the United States. The AAPM will periodically define new practice guidelines for medical physics practice to help advance the science of medical physics and to improve the quality of service to patients throughout the United States. Existing medical physics practice guidelines will be reviewed for the purpose of revision or renewal, as appropriate, on their fifth anniversary or sooner. Each medical physics practice guideline represents a policy statement by the AAPM, has undergone a thorough consensus process in which it has been subjected to extensive review, and requires the approval of the Professional Council. The medical physics practice guidelines recognize that the safe and effective use of diagnostic and therapeutic radiology requires specific training, skills, and techniques, as described in each document. Reproduction or modification of the published practice guidelines and technical standards by those entities not providing these services is not authorized. 1

    IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119.

    Get PDF
    AAPM Task Group 119 has produced quantitative confidence limits as baseline expectation values for IMRT commissioning. A set of test cases was developed to assess the overall accuracy of planning and delivery of IMRT treatments. Each test uses contours of targets and avoidance structures drawn within rectangular phantoms. These tests were planned, delivered, measured, and analyzed by nine facilities using a variety of IMRT planning and delivery systems. Each facility had passed the Radiological Physics Center credentialing tests for IMRT. The agreement between the planned and measured doses was determined using ion chamber dosimetry in high and low dose regions, film dosimetry on coronal planes in the phantom with all fields delivered, and planar dosimetry for each field measured perpendicular to the central axis. The planar dose distributions were assessed using gamma criteria of 3%/3 mm. The mean values and standard deviations were used to develop confidence limits for the test results using the concept confidence limit = /mean/ + 1.96sigma. Other facilities can use the test protocol and results as a basis for comparison to this group. Locally derived confidence limits that substantially exceed these baseline values may indicate the need for improved IMRT commissioning

    Dosimetry tools and techniques for IMRT

    Full text link
    Peer Reviewedhttp://deepblue.lib.umich.edu/bitstream/2027.42/98734/1/MPH001313.pd

    Low cost/no cost quality of life programs in non-profit theatre

    No full text
    In this paper, I will argue for quality-of-life programs as an effective recruiting tool for nonprofit theatres. Through review of the current literature and interviews, I will explore impediments to and best practice implementation strategies, while examining how theatres are grappling with changing priorities and attitudes of non-profit theatre leadership. Specific focus is given to low cost/no cost quality of life programs that theaters can employ to entice young leaders to enter and remain in the field. Finally, this paper includes recommendations for California Repertory Company (Cal Rep), the graduate arm of the Theatre Department at California State University, Long Beach. The strategies laid out in this paper could give Cal Rep additional tools with which to recruit and retain the best possible students for its MFA/MBA program