103 research outputs found
A FABRY-PEROT CAVITY FOR COMPTON POLARIMETRY
A new kind of Compton polarimeter using a resonant Fabry—Pe« rot cavity as a power buildup for the photon beam is proposed. A prototype of such a cavity is described, along with the results obtained in terms of source to be used in a Compton scattering polarimeter. ( 1998 Elsevier Science B.V. All rights reserved
SFE16, a low noise front-end integrated circuit dedicated to the read-out of large Micromegas detectors
Abstract A front-end BiCMOS ASIC was specially developed for the Micromegas detectors to be used in the Small Angle Tracker of the COMPASS experiment at CERN. Each of the 16 channels of this integrated circuit contains a low noise preamplifier with a 100 ns peaking time filter and a discriminator driving a low-level differential digital buffer. The design of the preamplifier and the choice of the shaping have been tuned to the detector signal shape in order to allow the operation of Micromegas even for very low multiplication gain values. Noise measurements show an equivalent noise charge of less than 1500 e-rms for a detector capacitance of 40pF. The measured performances of this ASIC associated or not with the detector are fully described in this paper
The Fast Read-out System for the MAPMTs of COMPASS RICH-1
A fast readout system for the upgrade of the COMPASS RICH detector has been
developed and successfully used for data taking in 2006 and 2007. The new
readout system for the multi-anode PMTs in the central part of the photon
detector of the RICH is based on the high-sensitivity MAD4
preamplifier-discriminator and the dead-time free F1-TDC chip characterized by
high-resolution. The readout electronics has been designed taking into account
the high photon flux in the central part of the detector and the requirement to
run at high trigger rates of up to 100 kHz with negligible dead-time. The
system is designed as a very compact setup and is mounted directly behind the
multi-anode photomultipliers. The data are digitized on the frontend boards and
transferred via optical links to the readout system. The read-out electronics
system is described in detail together with its measured performances.Comment: Proceeding of RICH2007 Conference, Trieste, Oct. 2007. v2: minor
change
Fast Photon Detection for Particle Identification with COMPASS RICH-1
Particle identification at high rates is an important challenge for many
current and future high-energy physics experiments. The upgrade of the COMPASS
RICH-1 detector requires a new technique for Cherenkov photon detection at
count rates of several per channel in the central detector region, and a
read-out system allowing for trigger rates of up to 100 kHz. To cope with these
requirements, the photon detectors in the central region have been replaced
with the detection system described in this paper. In the peripheral regions,
the existing multi-wire proportional chambers with CsI photocathode are now
read out via a new system employing APV pre-amplifiers and flash ADC chips. The
new detection system consists of multi-anode photomultiplier tubes (MAPMT) and
fast read-out electronics based on the MAD4 discriminator and the F1-TDC chip.
The RICH-1 is in operation in its upgraded version for the 2006 CERN SPS run.
We present the photon detection design, constructive aspects and the first
Cherenkov light in the detector.Comment: Proceedings of the Imaging 2006 conference, Stockholm, Sweden, 27-30
June 2006, 5 pages, 6 figures, to appear in NIM A; corrected typo in caption
of Fig.
Fast photon detection for the COMPASS RICH detector
The COMPASS experiment at the SPS accelerator at CERN uses a large scale Ring
Imaging CHerenkov detector (RICH) to identify pions, kaons and protons in a
wide momentum range. For the data taking in 2006, the COMPASS RICH has been
upgraded in the central photon detection area (25% of the surface) with a new
technology to detect Cherenkov photons at very high count rates of several 10^6
per second and channel and a new dead-time free read-out system, which allows
trigger rates up to 100 kHz. The Cherenkov photons are detected by an array of
576 visible and ultra-violet sensitive multi-anode photomultipliers with 16
channels each. The upgraded detector showed an excellent performance during the
2006 data taking.Comment: Proceeding of the IPRD06 conference (Siena, Okt. 06
Recommended from our members
A New Type of Compton Polarimeter for Jefferson Lab Hall A
In this paper, we present the principle and the studies for a new type of Compton polarimeter based on the use of power buildup Fabry-Perot cavity to get a 3% measurement of the Jefferson Lab electron polarization within minutes
First Measurement of the Transverse Spin Asymmetries of the Deuteron in Semi-Inclusive Deep Inelastic Scattering
First measurements of the Collins and Sivers asymmetries of charged hadrons
produced in deep-inelastic scattering of muons on a transversely polarized
6-LiD target are presented. The data were taken in 2002 with the COMPASS
spectrometer using the muon beam of the CERN SPS at 160 GeV/c. The Collins
asymmetry turns out to be compatible with zero, as does the measured Sivers
asymmetry within the present statistical errors.Comment: 6 pages, 2 figure
The COMPASS Experiment at CERN
The COMPASS experiment makes use of the CERN SPS high-intensitymuon and
hadron beams for the investigation of the nucleon spin structure and the
spectroscopy of hadrons. One or more outgoing particles are detected in
coincidence with the incoming muon or hadron. A large polarized target inside a
superconducting solenoid is used for the measurements with the muon beam.
Outgoing particles are detected by a two-stage, large angle and large momentum
range spectrometer. The setup is built using several types of tracking
detectors, according to the expected incident rate, required space resolution
and the solid angle to be covered. Particle identification is achieved using a
RICH counter and both hadron and electromagnetic calorimeters. The setup has
been successfully operated from 2002 onwards using a muon beam. Data with a
hadron beam were also collected in 2004. This article describes the main
features and performances of the spectrometer in 2004; a short summary of the
2006 upgrade is also given.Comment: 84 papes, 74 figure
- …