19 research outputs found

    Intercomparison of air ion spectrometers: An evaluation of results in varying conditions

    Get PDF
    We evaluated 11 air ion spectrometers from Airel Ltd. after they had spent one year in field measurements as a part of the EUCAARI project: 5 Air Ion Spectrometers (AIS), 5 Neutral cluster and Air Ion Spectrometers (NAIS) and one Airborne NAIS (ANAIS). This is the first time that an ANAIS is evaluated and compared so extensively. The ion spectrometers' mobility and concentration accuracy was evaluated. Their measurements of ambient air were compared between themselves and to reference instruments: a Differential Mobility Particle Sizer (DMPS), a Balanced Scanning Mobility Analyzer (BSMA), and an Ion-DMPS. We report on the simultaneous measurement of a new particle formation (NPF) event by all 11 instruments and the 3 reference instruments. To our knowledge, it is the first time that the size distribution of ions and particles is measured by so many ion spectrometers during a NPF event. The new particle formation rates (~0.2 cm−3 s−1 for ions and ~2 cm−3 s−1 for particles) and growth rates (~25 nm h−1 in the 3–7 nm size range) were calculated for all the instruments. The NAISs and the ANAIS gave higher concentrations and formation rates than the AISs. For example, the AISs agreed with the BSMA within 11 % and 28 % for negative and positive ion concentration respectively, whereas the NAISs agreed within 23 % and 29 %. Finally, based on the results presented here, we give guidelines for data evaluation, when data from different individual ion spectrometers are compared

    Technical Note : Using DEG-CPCs at upper tropospheric temperatures

    Get PDF
    Cited By :6 Export Date: 30 January 2018Over the last few years, several condensation particle counters (CPCs) capable of measuring in the sub-3 nm size range have been developed. Here we study the performance of CPCs based on diethylene glycol (DEG) at different temperatures during Cosmics Leaving OUtdoor Droplets (CLOUD) measurements at CERN. The data shown here are the first set of verification measurements for sub-3 nm CPCs under upper tropospheric temperatures using atmospherically relevant aerosol particles. To put the results in perspective we calibrated the DEG-CPC at room temperature, resulting in a cut-off diameter of 1.4 nm. All diameters refer to mobility equivalent diameters in this paper. At upper tropospheric temperatures ranging from 246.15 K to 207.15 K, we found cut-off sizes relative to a particle size magnifier in the range of 2.5 to 2.8 nm. Due to low number concentration after size classification, the cut-off diameters have a high uncertainty (+/- 0.3 nm) associated with them. Operating two laminar flow DEG-CPCs with different cut-off sizes together with other aerosol instruments, we looked at the growth rates of aerosol population in the CLOUD chamber for particles smaller than 10 nm at different temperatures. A more consistent picture emerged when we normalized the growth rates to a fixed gas-phase sulfuric acid concentration. All of the instruments detected larger growth rates at lower temperatures, and the observed growth rates decreased as a function of temperature, showing a similar trend for all instruments. The theoretical calculations had a similar but much smaller temperature dependency.Peer reviewe

    EUCAARI ion spectrometer measurements at 12 European sites – analysis of new particle formation events

    Get PDF
    We present comprehensive results on continuous atmospheric cluster and particle measurements in the size range ~1–42 nm within the European Integrated project on Aerosol Cloud Climate and Air Quality interactions (EUCAARI) project. We focused on characterizing the spatial and temporal variation of new particle formation events and relevant particle formation parameters across Europe. Different types of air ion and cluster mobility spectrometers were deployed at 12 field sites across Europe from March 2008 to May 2009. The measurements were conducted in a wide variety of environments, including coastal and continental locations as well as sites at different altitudes (both in the boundary layer and the free troposphere). New particle formation events were detected at all of the 12 field sites during the year-long measurement period. From the data, nucleation and growth rates of newly formed particles were determined for each environment. In a case of parallel ion and neutral cluster measurements, we could also estimate the relative contribution of ion-induced and neutral nucleation to the total particle formation. The formation rates of charged particles at 2 nm accounted for 1–30% of the corresponding total particle formation rates. As a significant new result, we found out that the total particle formation rate varied much more between the different sites than the formation rate of charged particles. This work presents, so far, the most comprehensive effort to experimentally characterize nucleation and growth of atmospheric molecular clusters and nanoparticles at ground-based observation sites on a continental scale

    Experimental investigation of ion-ion recombination under atmospheric conditions

    Get PDF
    Cited By :16 Export Date: 30 January 2018We present the results of laboratory measurements of the ion-ion recombination coefficient at different temperatures, relative humidities and concentrations of ozone and sulfur dioxide. The experiments were carried out using the Cosmics Leaving OUtdoor Droplets (CLOUD) chamber at CERN, the walls of which are made of conductive material, making it possible to measure small ions. We produced ions in the chamber using a 3.5 GeV c(-1) beam of positively charged pions (pi(+)) generated by the CERN Proton Synchrotron (PS). When the PS was switched off, galactic cosmic rays were the only ionization source in the chamber. The range of the ion production rate varied from 2 to 100 cm(-3) s(-1), covering the typical range of ionization throughout the troposphere. The temperature ranged from -55 to 20 degrees C, the relative humidity (RH) from 0 to 70 %, the SO2 concentration from 0 to 40 ppb, and the ozone concentration from 200 to 700 ppb. The best agreement of the retrieved ion-ion recombination coefficient with the commonly used literature value of 1.6 x 10(-6) cm(3) s(-1) was found at a temperature of 5 degrees C and a RH of 40% (1.5 +/- 0.6) x 10(-6) cm(3) s(-1). At 20 degrees C and 40% RH, the retrieved ion-ion recombination coefficient was instead (2.3 +/- 0.7) x 10(-6) cm(3) s(-1). We observed no dependency of the ion-ion recombination coefficient on ozone concentration and a weak variation with sulfur dioxide concentration. However, we observed a more than fourfold increase in the ion-ion recombination coefficient with decreasing temperature. We compared our results with three different models and found an overall agreement for temperatures above 0 degrees C, but a disagreement at lower temperatures. We observed a strong increase in the recombination coefficient for decreasing relative humidities, which has not been reported previously.Peer reviewe

    Experimental investigation of ion-ion recombination under atmospheric conditions

    Get PDF
    Cited By :16 Export Date: 30 January 2018We present the results of laboratory measurements of the ion-ion recombination coefficient at different temperatures, relative humidities and concentrations of ozone and sulfur dioxide. The experiments were carried out using the Cosmics Leaving OUtdoor Droplets (CLOUD) chamber at CERN, the walls of which are made of conductive material, making it possible to measure small ions. We produced ions in the chamber using a 3.5 GeV c(-1) beam of positively charged pions (pi(+)) generated by the CERN Proton Synchrotron (PS). When the PS was switched off, galactic cosmic rays were the only ionization source in the chamber. The range of the ion production rate varied from 2 to 100 cm(-3) s(-1), covering the typical range of ionization throughout the troposphere. The temperature ranged from -55 to 20 degrees C, the relative humidity (RH) from 0 to 70 %, the SO2 concentration from 0 to 40 ppb, and the ozone concentration from 200 to 700 ppb. The best agreement of the retrieved ion-ion recombination coefficient with the commonly used literature value of 1.6 x 10(-6) cm(3) s(-1) was found at a temperature of 5 degrees C and a RH of 40% (1.5 +/- 0.6) x 10(-6) cm(3) s(-1). At 20 degrees C and 40% RH, the retrieved ion-ion recombination coefficient was instead (2.3 +/- 0.7) x 10(-6) cm(3) s(-1). We observed no dependency of the ion-ion recombination coefficient on ozone concentration and a weak variation with sulfur dioxide concentration. However, we observed a more than fourfold increase in the ion-ion recombination coefficient with decreasing temperature. We compared our results with three different models and found an overall agreement for temperatures above 0 degrees C, but a disagreement at lower temperatures. We observed a strong increase in the recombination coefficient for decreasing relative humidities, which has not been reported previously.Peer reviewe

    Atmospheric ions and nucleation: a review of observations

    Get PDF
    This review is based on ca. 260 publications, 93 of which included data on the temporal and spatial variation of the concentration of small ions (<1.6 nm in diameter) especially in the lower troposphere, chemical composition, or formation and growth rates of sub-3 nm ions. This information was collected on tables and figures. The small ions exist all the time in the atmosphere, and the average concentrations of positive and negative small ions are typically 200-2500 cm(-3). However, concentrations up to 5000 cm(-3) have been observed. The results are in agreement with observations of ion production rates in the atmosphere. We also summarised observations on the conversion of small ions to intermediate ions, which can act as embryos for new atmospheric aerosol particles. Those observations include the formation rates (J(2)[ion]) of 2-nm intermediate ions, growth rates (GR[ion]) of sub-3 nm ions, and information on the chemical composition of the ions. Unfortunately, there were only a few studies which presented J(2)[ion] and GR[ion]. Based on the publications, the formation rates of 2-nm ions were 0-1.1 cm(-3) s(-1), while the total 2-nm particle formation rates varied between 0.001 and 60 cm(-3) s(-1). Due to small changes in J(2)[ion], the relative importance of ions in 2-nm particle formation was determined by the large changes in J(2)[tot], and, accordingly the contribution of ions increased with decreasing J(2)[tot]. Furthermore, small ions were observed to activate for growth earlier than neutral nanometer-sized particles and at lower saturation ratio of condensing vapours

    Observations of turbulence-induced new particle formation in the residual layer

    Get PDF
    Aerosol particle measurements in the atmospheric boundary layer performed by a helicopter-borne measurement payload and by a lidar system from a case study during the IMPACT field campaign in Cabauw (NL) are presented. Layers of increased number concentrations of ultrafine particles were observed in the residual layer, indicating relatively recent new-particle formation. These layers were characterized by a sub-critical Richardson number and concomitant increased turbulence. Turbulent mixing is likely to lead to local supersaturation of possible precursor gases which are essential for new particle formation. Observed peaks in the number concentrations of ultrafine particles at ground level are connected to the new particle formation in the residual layer by boundary layer development and vertical mixing
    corecore