The Ramsey method in high-precision mass spectrometry with Penning traps: Experimental results
The highest precision in direct mass measurements is obtained with Penning
trap mass spectrometry. Most experiments use the interconversion of the
magnetron and cyclotron motional modes of the stored ion due to excitation by
external radiofrequency-quadrupole fields. In this work a new excitation
scheme, Ramsey's method of time-separated oscillatory fields, has been
successfully tested. It has been shown to reduce significantly the uncertainty
in the determination of the cyclotron frequency and thus of the ion mass of
interest. The theoretical description of the ion motion excited with Ramsey's
method in a Penning trap and subsequently the calculation of the resonance line
shapes for different excitation times, pulse structures, and detunings of the
quadrupole field has been carried out in a quantum mechanical framework and is
discussed in detail in the preceding article in this journal by M. Kretzschmar.
Here, the new excitation technique has been applied with the ISOLTRAP mass
spectrometer at ISOLDE/CERN for mass measurements on stable as well as
short-lived nuclides. The experimental resonances are in agreement with the
theoretical predictions and a precision gain close to a factor of four was
achieved compared to the use of the conventional excitation technique.Comment: 12 pages, 14 figures, 2 table
Evidence for a breakdown of the Isobaric Multiplet Mass Equation: A study of the A=35, T=3/2 isospin quartet
Mass measurements on radionuclides along the potassium isotope chain have
been performed with the ISOLTRAP Penning trap mass spectrometer. For 35K
T1/2=178ms) to 46K (T1/2=105s) relative mass uncertainties of 2x10-8 and better
have been achieved. The accurate mass determination of 35K (dm=0.54keV) has
been exploited to test the Isobaric Multiplet Mass Equation (IMME) for the
A=35, T=3/2 isospinquartet. The experimental results indicate a deviation from
the generally adopted quadratic form.Comment: 8 pages, 4 figure
Separated Oscillatory Fields for High-Precision Penning Trap Mass Spectrometry
Ramsey's method of separated oscillatory fields is applied to the excitation
of the cyclotron motion of short-lived ions in a Penning trap to improve the
precision of their measured mass. The theoretical description of the extracted
ion-cyclotron-resonance line shape is derived out and its correctness
demonstrated experimentally by measuring the mass of the short-lived Ca
nuclide with an uncertainty of using the ISOLTRAP Penning
trap mass spectrometer at CERN. The mass value of the superallowed beta-emitter
Ca is an important contribution for testing the conserved-vector-current
hypothesis of the electroweak interaction. It is shown that the Ramsey method
applied to mass measurements yields a statistical uncertainty similar to that
obtained by the conventional technique ten times faster.Comment: 5 pages, 4 figures, 0 table
Magnetic field stabilization for high-accuracy mass measurements on exotic nuclides
The magnetic-field stability of a mass spectrometer plays a crucial role in
precision mass measurements. In the case of mass determination of short-lived
nuclides with a Penning trap, major causes of instabilities are temperature
fluctuations in the vicinity of the trap and pressure fluctuations in the
liquid helium cryostat of the superconducting magnet. Thus systems for the
temperature and pressure stabilization of the Penning trap mass spectrometer
ISOLTRAP at the ISOLDE facility at CERN have been installed. A reduction of the
fluctuations by at least one order of magnitude downto dT=+/-5mK and
dp=+/-50mtorr has been achieved, which corresponds to a relative frequency
change of 2.7x10^{-9} and 1.5x10^{-10}, respectively. With this stabilization
the frequency determination with the Penning trap only shows a linear temporal
drift over several hours on the 10 ppb level due to the finite resistance of
the superconducting magnet coils.Comment: 23 pages, 13 figure
Production and trapping of carbon clusters for absolute mass measurements at ISOLTRAP
Singly-charged carbon clusters C/sub n//sup +/ (n >or= 1) have been produced by laser-induced desorption and fragmentation of C/sub 60/ fullerenes and have been injected into and stored in the Penning trap system of the ISOLTRAP mass spectrometer at ISOLDE/CERN. The present study is the first step to extend the until now direct mass measurements at ISOLTRAP to absolute mass measurements by using clusters of /sup 12/C. (10 refs)
Electric and magnetic field optimization procedure for Penning trap mass spectrometers
Significant systematic errors in high-precision Penning trap mass
spectrometry can result from electric and magnetic field imperfections. An
experimental procedure to minimize these uncertainties is presented for the
on-line Penning trap mass spectrometer ISOLTRAP, located at ISOLDE/CERN. The
deviations from the ideal magnetic and electric fields are probed by measuring
the cyclotron frequency and the reduced cyclotron frequency, respectively, of
stored ions as a function of the time between the ejection of ions from the
preparation trap and their capture in the precision trap, which influences the
energy of their axial motion. The correction parameters are adjusted to
minimize the frequency shifts.Comment: 25 pages, 9 figure
Transport and cooling of singly-charged noble gas ion beams
The transport and cooling of noble gas singly-charged ion beams by means of a
Radio Frequency Quadrupole Cooler Buncher (RFQCB) have been studied at the
LIMBE low energy beam line of the GANIL facility. Ions as light as
have been cooled and stored before their extraction in bunches using as
buffer gas. Bunches characteristics have been studied as a function of the
parameters of the device. Sizeable transmissions of up to 10 have been
obtained. A detailed study of the lifetime of ions inside the buncher has been
performed giving an estimate of the charge exchange cross-section. Results of a
microscopic Monte-Carlo transport code show reasonable agreement with
experimental data.Comment: 13 figure
From direct to absolute mass measurements: a study of the accuracy of ISOLTRAP
For a detailed study of the accuracy of the Penning trap mass spectrometer ISOLTRAP all expected sources of uncertainty were investigated with respect to their contributions to the uncertainty of the final result. In the course of these investigations, cross-reference measurements with singly charged carbon clusters C were carried out. The carbon cluster ions were produced by use of laser-induced desorption, fragmentation, and ionization of C fullerenes and injected into and stored in the Penning trap system. The comparison of the cyclotron frequencies of different carbon clusters has provided detailed insight into the residual systematic uncertainty of \acro{ISOLTRAP} and yielded a value of . This also represents the current limit of mass accuracy of the apparatus. Since the unified atomic mass unit is defined as 1/12 of the mass of the C atom, it will be possible to carry out absolute mass measurements with \acro{ISOLTRAP} in the future.\\[1\baselineskip] PACS: 07.75.+h (Mass spectrometers and related techniques), 21.10.Dr (Binding energies and masses), 32.10.Bi (Atomic masses, mass spectra, abundances, and isotopes), 36.40.Wa (Charged clusters)
- …