    Critical Assessment of Single-Use Ureteroscopes in an In Vivo Porcine Model

    Get PDF
    Methods A female pig was placed under general anesthesia and positioned supine, and retrograde access to the renal collecting system was obtained. The LithoVue (Boston Scientific) and Uscope (Pusen Medical) were evaluated by three experienced surgeons, and each surgeon started with a new scope. The following parameters were compared between each ureteroscope: time for navigation to upper and lower pole calyces with and without implements (1.9 F basket, 200 μm laser fiber, and 365 μm laser fiber for upper only) in the working channel and subjective evaluations of maneuverability, irrigant flow through the scope, lever force, ergonomics, and scope optics. Results Navigation to the lower pole calyx was significantly faster with LithoVue compared to Uscope when the working channel was empty (24.3 vs. 49.4 seconds, p < 0.01) and with a 200 μm fiber (63.6 vs. 94.4 seconds, p=0.04), but not with the 1.9 F basket. Navigation to the upper pole calyx was similar for all categories except faster with LithoVue containing the 365 μm fiber (67.1 vs. 99.7 seconds, p=0.02). Subjective assessments of scope maneuverability to upper and lower pole calyces when the scope was empty and with implements favored LithoVue in all categories, as did assessments of irrigant flow, illumination, image quality, and field of view. Both scopes had similar scores of lever force and ergonomics. Conclusions In an in vivo porcine model, the type of single-use ureteroscope employed affected the navigation times and subjective assessments of maneuverability and visualization. In all cases, LithoVue provided either equivalent or superior metrics than Uscope. Further clinical studies are necessary to determine the implications of these findings

    In vitro head-to-head comparison of the durability, versatility and efficacy of the NGage and novel Dakota stone retrieval baskets

    No full text
    BackgroundTo compare head to head two end-engaging nitinol stone retrieval devices available to urologists, in terms of durability, versatility and efficacy.MethodsFor durability testing, 30 NGage and Dakota baskets were cycled 20 times between grasping and releasing synthetic stone models and evaluated for damage or device failure. For versatility and efficacy testing, baskets were assessed in their ability to capture and release stone models from 1 to 11 mm. Each stone was raised above the capture site and the basket was opened to passively release the stone. If the stone did not release, the basket handle was shaken and the OpenSure feature employed if needed. Manual release was used as a last resort.ResultsDurability-the Cook NGage demonstrated a statistically significant increased rate of visible device breakdown (P=0.0046) in 8 of 30 (26.7%) devices vs. 0 of 30 Dakota devices, with mean damage at 13.5 cycles. Versatility and efficacy-both 8 mm baskets successfully captured stones from 1-8 mm. The Dakota more effectively released 7-8 mm stones (P&lt;0.0001). NGage required manual release of 8 mm stones in 13 cases compared to none with Dakota. For 11 mm baskets, the Dakota released all stones up to 10 mm with simple opening, while the NGage released 10 of 15 (67%) of 9 mm stones and 1 of 15 (7%) of 10 mm stones by simple opening. For 11 mm stones, the Dakota captured 100% whereas NGage could not capture any.ConclusionsBoth baskets showed similar durability characteristics. The Dakota basket more effectively captured and released stones over 7 mm, as compared to the NGage basket. The OpenSure aspect conferred an advantage in handling and release of larger stones. These in vitro results demonstrate potential versatility, durability and efficacy of the Dakota basket

    Reply to M. Froehner

    No full text
    corecore