277 research outputs found
Leadless Cardiac Pacemakers Back to the Future
AbstractDespite significant advances in battery longevity, lead performance, and programming features since the first implanted permanent pacemaker was developed, the basic design of cardiac pacemakers has remained relatively unchanged over the past 50 years. Because of inherent limitations in their design, conventional (transvenous) pacemakers are prone to multiple potential short- and long-term complications. Accordingly, there has been intense interest in a system able to provide the symptomatic and potentially lifesaving therapies of cardiac pacemakers while mitigating many of the risks associated with their weakest link—the transvenous lead. Leadless cardiac pacing represents the future of cardiac pacing systems, similar to the transition that occurred from the use of epicardial pacing systems to the familiar transvenous systems of today. This review summarizes the current evidence and potential benefits of leadless pacing systems, which are either commercially available (in Europe) or under clinical investigation
Left atrial appendage closure for stroke prevention in atrial fibrillation: current status and perspectives.
Atrial fibrillation (AF) is associated with an increased risk of stroke and systemic embolism, and the left atrial appendage (LAA) has been identified as a principal source of thromboembolism in these patients. While oral anticoagulation is the current standard of care, LAA closure (LAAC) emerges as an alternative or complementary treatment approach to reduce the risk of stroke or systemic embolism in patients with AF. Moderate-sized randomized clinical studies have provided data for the efficacy and safety of catheter-based LAAC, largely compared with vitamin K antagonists. LAA device iterations, advances in pre- and peri-procedural imaging, and implantation techniques continue to increase the efficacy and safety of LAAC. More data about efficacy and safety of LAAC have been collected, and several randomized clinical trials are currently underway to compare LAAC with best medical care (including non-vitamin K antagonist oral anticoagulants) in different clinical settings. Surgical LAAC in patients with AF undergoing cardiac surgery reduced the risk of stroke on background of anticoagulation therapy in the LAAOS III study. In this review, we describe the rapidly evolving field of LAAC and discuss recent clinical data, ongoing studies, open questions, and current limitations of LAAC
Radiofrequency Ablation Using a Novel Insulated-Tip Ablation Catheter Can Create Uniform Lesions Comparable in Size to Conventional Irrigated Ablation Catheters While Using a Fraction of the Energy and Irrigation
[EN] Introduction: During radiofrequency ablation (RFA) using conventional RFA catheters (RFC), similar to 90% of the energy dissipates into the bloodstream/surrounding tissue. We hypothesized that a novel insulated-tip ablation catheter (SMT) capable of blocking the radiofrequency path may focus most of the energy into the targeted tissue while utilizing reduced power and irrigation.
Methods: This study evaluated the outcomes of RFA using SMT versus an RFC in silico, ex vivo, and in vivo. Radiofrequency applications were delivered over porcine myocardium (ex vivo) and porcine thigh muscle preparations superfused with heparinized blood (in vivo). Altogether, 274 radiofrequency applications were delivered using SMT (4-15 W, 2 or 20 ml/min) and 74 applications using RFC (30 W, 30 ml/min).
Results: RFA using SMT proved capable of directing 66.8% of the radiofrequency energy into the targeted tissue. Accordingly, low power-low irrigation RFA using SMT (8-12 W, 2 ml/min) yielded lesion sizes comparable with RFC, whereas high power-high irrigation (15 W, 20 ml/min) RFA with SMT yielded lesions larger than RFC (p < .05). Although SMT was associated with greater impedance drops ex vivo and in vivo, ablation using RFC was associated with increased charring/steam pop/ tissue cavitation (p < .05). Lastly, lesions created with SMT were more homogeneous than RFC (p < .001).
Conclusion: Low power-low irrigation (8-12 W, 2 ml/min) RFA using the novel SMT ablation catheter can create more uniform, but comparable-sized lesions as RFC with reduced charring/steam pop/tissue cavitation. High power-high irrigation (15 W, 20 ml/min) RFA with SMT yields lesions larger than RFC.Sirona Medical Technologies, Inc; The Proyecto UNAJ Investiga 2017, Grant/Award Number: 80020170100019UJ; The Spanish Ministerio de Ciencia, Innovacion y Universidades/Agencia Estatal de Investigacion (MCIN/AEI/10.13039/501100011033), Grant/Award Number: RTI2018-094357B-C21Aryana, A.; Irastorza, RM.; Berjano, E.; Cohen, RJ.; Kraus, J.; Haghighi-Mood, A.; Reddy, VY.... (2022). Radiofrequency Ablation Using a Novel Insulated-Tip Ablation Catheter Can Create Uniform Lesions Comparable in Size to Conventional Irrigated Ablation Catheters While Using a Fraction of the Energy and Irrigation. Journal of Cardiovascular Electrophysiology. 33(6):1146-1156. https://doi.org/10.1111/jce.154611146115633
Incidence, predictors, and evolution of conduction disorders and atrial arrhythmias after contemporary mitral valve repair
Background: Conduction disorders (CD) and atrial arrhythmias (AA) in the postoperative period of cardiac surgery impede prompt clinical recovery and prolong hospitalization. Mitral valve repair (MVR) has become the treatment of choice for patients with significant valvular regurgitation, but information on CD and AA in this population is scarce.Methods: Records of consecutive patients undergoing MVR at a single center were reviewed. Patients with a preoperative pacemaker, CD, prior cardiac surgery or concomitant MVR were excluded. A total of 290 patients were included in the final analysis. Electrocardiograms pre andpost-operatively were analyzed for CD and AA.Results: CD occurred in 69 (23.7%) patients: 47 (16.2%) had first degree atrio-ventricular block (AVB), 10 (3.4%) had Mobitz I, 3 (1.03%) had Mobitz II, and 9 (3.1%) complete AVB. Only 6 (2.0%) patients required pacemakers. Univariate predictors of AVB were age, preexisting right bundle branch block (RBBB), mitral valve ring size, and bypass time. The only multivariate predictors of AVB were bypass time and preexisting RBBB (OR 3.23 and 1.98, respectively). The most common AA was atrial fibrillation 13.1% (38 patients) followed by atrial flutter 2.7% (8 patients). Multivariate predictors of AA were age and left atrial size (OR 1.85 and 4.2, respectively). Length of stay in patients with CD or AA was prolonged 2.2 ± 2 days compared to controls (p < 0.05).Conclusions: In this large sample of patients undergoing MVR, we found that bypass time and preexisting RBBB were independent predictors of CD; age and left atrial size were independent predictors of AA.
Isolation of pulmonary veins using a thermoreactive implantable device with external energy transfer : evaluation in a porcine model
Background: Pulmonary vein isolation (PVI) is a well-established method for the treatment of symptomatic paroxysmal atrial fibrillation, but is only partly successful with a high rate of electrical reconnection. We introduce a novel technique in which PVI is accomplished by noninvasive heating of a dedicated thermoresponse implant inserted into the pulmonary veins (PV), demonstrated in a porcine model.
Methods: A self-expanding nitinol-based implant was positioned in the common inferior PV of 11 pigs, using a fluoroscopy-guided transatrial appendage approach. Ablation was performed through contactless energy transfer from a primary extracorporal coil to a secondary heat ring (HR) embedded in the proximal part of the implant. Electrophysiological conduction was assessed prior to and postablation, and at 3 months. Histological samples were obtained acutely (n=4) and after 3 months (n=7).
Results: In total, 13 PV implants were successfully positioned in the inferior PVs of 11 animals. Ablation was performed without injury of adjacent structures. PVI and bidirectional block was electrophysiologically confirmed in all cases immediately at the time of implantation and 3 months later in seven chronic animals in whom testing was repeated. Marked evidence of ablation around the proximal HR was evident at 3 months postprocedure, with scar tissue formation and only mild neointimal proliferation.
Conclusions: Successful PVI can be obtained by external electromagnetic heat transfer to a novel pulmonary vein implant
A Randomized Trial of Electrographic Flow-Guided Redo Ablation for Nonparoxysmal Atrial Fibrillation (FLOW-AF)
Background: Electrographic flow (EGF) mapping enables full spatiotemporal reconstruction of organized wavefront propagation to identify extrapulmonary vein sources of atrial fibrillation (AF). Objectives: FLOW-AF (A Randomized Controlled Study to Evaluate the Reliability of the Ablacon Electrographic FLOW [EGF] Algorithm Technology [Ablamap Software] to Identify AF Sources and Guide Ablation Therapy in Patients With Persistent Atrial Fibrillation) was multicenter, randomized controlled study of EGF mapping to: 1) stratify a nonparoxysmal AF population undergoing redo ablation; 2) guide ablation of these extrapulmonary vein AF sources; and 3) improve AF recurrence outcomes. Methods: FLOW-AF enrolled persistent atrial fibrillation (PerAF)/long-standing PerAF patients undergoing redo ablation at 4 centers. One-minute EGF maps were recorded from standardized biatrial basket positions. Patients with source activity ≥26.5% were randomized 1:1 to PVI + EGF-guided ablation vs PVI only; patients without sources ≥26.5% threshold were not randomized. Follow-up and electrocardiographic monitoring occurred at 3, 6, and 12 months. Results: We enrolled 85 patients (age 65.6 ± 9.3 years, 37% female, 24% long-standing PerAF). Thirty-four (40%) patients had no sources greater than threshold; at least 1 source greater than threshold was present in 46 (60%) (EGF-guided ablation, n = 22; control group, n = 26). Patients with sources were older (68.2 vs 62.6 years; P = 0.005) with higher CHA2DS2-VASc scores (2.8 vs 1.9; P = 0.001). The freedom from safety events was 97.2%, and 95% of EGF-identified sources were successfully ablated. In randomized patients, AF-free survival at 12 months was 68% for EGF-guided ablation vs 17% for the control group (P = 0.042); freedom from AF/atrial tachycardia/atrial flutter at 12 months was 51% vs 14% (P = 0.103), respectively. Conclusions: In nonparoxysmal AF patients undergoing redo ablation, EGF mapping identified AF sources in 60% of patients, and could be successfully ablated in 95%. Compared with PVI alone, PVI + source ablation improved AF-free survival by 51% on an absolute basis.</p
Initial Independent Outcomes from Focal Impulse and Rotor Modulation Ablation for Atrial Fibrillation: Multicenter FIRM Registry
Introduction
The success of pulmonary vein isolation (PVI) for atrial fibrillation (AF) may be improved if stable AF sources identified by Focal Impulse and Rotor Mapping (FIRM) are also eliminated. The long-term results of this approach are unclear outside the centers where FIRM was developed; thus, we assessed outcomes of FIRM-guided AF ablation in the first cases at 10 experienced centers.
Methods
We prospectively enrolled n = 78 consecutive patients (61 ± 10 years) undergoing FIRM guided ablation for persistent (n = 48), longstanding persistent (n = 7), or paroxysmal (n = 23) AF. AF recordings from both atria with a 64-pole basket catheter were analyzed using a novel mapping system (Rhythm View™; Topera Inc., CA, USA). Identified rotors/focal sources were ablated, followed by PVI.
Results
Each institution recruited a median of 6 patients, each of whom showed 2.3 ± 0.9 AF rotors/focal sources in diverse locations. 25.3% of all sources were right atrial (RA), and 50.0% of patients had ≥1 RA source. Ablation of all sources required a total of 16.6 ± 11.7 minutes, followed by PVI. On >1 year follow-up with a 3-month blanking period, 1 patient lost to follow-up (median time to 1st recurrence: 245 days, IQR 145–354), single-procedure freedom from AF was 87.5% (patients without prior ablation; 35/40) and 80.5% (all patients; 62/77) and similar for persistent and paroxysmal AF (P = 0.89).
Conclusions
Elimination of patient-specific AF rotors/focal sources produced freedom-from-AF of ≈80% at 1 year at centers new to FIRM. FIRM-guided ablation has a rapid learning curve, yielding similar results to original FIRM reports in each center’s first cases
- …