113 research outputs found
Concept review of dry powder inhalers : correct interpretation of published data
Peer reviewedPublisher PD
The Inhalation Characteristics of Patients When They Use Different Dry Powder Inhalers
Background: The characteristics of each inhalation maneuver when patients use dry powder inhalers (DPIs) are important, because they control the quality of the emitted dose.
Methods: We have measured the inhalation profiles of asthmatic children [CHILD; n=16, mean forced expiratory volume in 1 sec (FEV1) 79% predicted], asthmatic adults (ADULT; n=53, mean predicted FEV1 72%), and chronic obstructive pulmonary disease (COPD; n=29, mean predicted FEV1 42%) patients when they inhaled through an Aerolizer, Diskus, Turbuhaler, and Easyhaler using their “real-life” DPI inhalation technique. These are low-, medium-, medium/high-, and high-resistance DPIs, respectively. The inhalation flow against time was recorded to provide the peak inhalation flow (PIF; in L/min), the maximum pressure change (ΔP; in kPa), acceleration rates (ACCEL; in kPa/sec), time to maximum inhalation, the length of each inhalation (in sec), and the inhalation volume (IV; in liters) of each inhalation maneuver.
Results: PIF, ΔP, and ACCEL values were consistent with the order of the inhaler's resistance. For each device, the inhalation characteristics were in the order ADULT>COPD>CHILD for PIF, ΔP, and ACCEL (p4 L and ΔP >4 kPa.
Conclusion: The large variability of these inhalation characteristics and their range highlights that if inhalation profiles were used with compendial in vitro dose emission measurements, then the results would provide useful information about the dose patients inhale during routine use. The inhalation characteristics highlight that adults with asthma have greater inspiratory capacity than patients with COPD, whereas children with asthma have the lowest. The significance of the inhaled volume to empty doses from each device requires investigation
Recommended from our members
In vitro Performance Assessment of Recent Nebuliser Delivery Systems for Nebulisation of Approved Aerosolised Tobramycin (TOBI)®
YesTOBI® is a recently marketed preservative and sulphate free tobramycin formulation approved by FDA for
maintenance therapy for patient with cystic fibrosis. The performance of selected recent nebuliser delivery systems has
been assessed using the developed method to determine the optimum combinations to deliver approved tobramycin
inhaled solution (TOBI)®. A simple, sensitive and specific high performance liquid chromatographic method has been
developed and used to quantitative determination of the aminoglycoside tobramycin following pre-column derivatisation
with phenylisocyanate (PIC). The reaction time was 10 min at 80º C and the resulting derivative was stable for five days
at room temperature. The quantitative performance of the assay was further improved by using another aminoglycoside
(neomycin) as internal standard. The stable resulting PIC-tobramycin derivative was separated using a HPLC 5μm
Columbus C18 column (150x4.60 mm i.d, Phenomenex). The mobile phase was consisted of acetonitrile-glacial acetic
acid-water (450:5:545, v/v/v) and ultraviolet detection at (240 nm). The proposed method showed good validation data.
The standard curve was linear (n=5) at seven different concentrations, ranging from 20 to 140μg/ml and the correlation
coefficient (R2) of the regression line was 0.9995. The limit of detection (LOD) and limit of quantitation (LOQ) were
0.86μg/ml and 2.62μg/ml, respectively. The relative standard deviation (RSD %) was less than 0.6% for intra-day assay
(n=5) and 2.5% for inter-day assay (n=5). A number of nebuliser performance comparison studies have been
demonstrated for aerosolise TOBI® to choice the optimum combination produces high repirable inhaled mass of
tobramycin. The objective of this study was to evaluate the performance of recent nebuliser delivery systems to nebulise
approved tobramycin inhaled solution (TOBI)®
Inhalation characteristics of asthma patients, COPD patients and healthy volunteers with the Spiromax® and Turbuhaler® devices: a randomised, cross-over study.
BACKGROUND: Spiromax® is a novel dry-powder inhaler containing formulations of budesonide plus formoterol (BF). The device is intended to provide dose equivalence with enhanced user-friendliness compared to BF Turbuhaler® in asthma and chronic obstructive pulmonary disease (COPD). The present study was performed to compare inhalation parameters with empty versions of the two devices, and to investigate the effects of enhanced training designed to encourage faster inhalation. METHODS: This randomised, open-label, cross-over study included children with asthma (n = 23), adolescents with asthma (n = 27), adults with asthma (n = 50), adults with COPD (n = 50) and healthy adult volunteers (n = 50). Inhalation manoeuvres were recorded with each device after training with the patient information leaflet (PIL) and after enhanced training using an In-Check Dial device. RESULTS: After PIL training, peak inspiratory flow (PIF), maximum change in pressure (∆P) and the inhalation volume (IV) were significantly higher with Spiromax than with the Turbuhaler device (p values were at least <0.05 in all patient groups). After enhanced training, numerically or significantly higher values for PIF, ∆P, IV and acceleration remained with Spiromax versus Turbuhaler, except for ∆P in COPD patients. After PIL training, one adult asthma patient and one COPD patient inhaled <30 L/min through the Spiromax compared to one adult asthma patient and five COPD patients with the Turbuhaler. All patients achieved PIF values of at least 30 L/min after enhanced training. CONCLUSIONS: The two inhalers have similar resistance so inhalation flows and pressure changes would be expected to be similar. The higher flow-related values noted for Spiromax versus Turbuhaler after PIL training suggest that Spiromax might have human factor advantages in real-world use. After enhanced training, the flow-related differences between devices persisted; increased flow rates were achieved with both devices, and all patients achieved the minimal flow required for adequate drug delivery. Enhanced training could be useful, especially in COPD patients
Recommended from our members
In vitro aerodynamic characterization of the dose emitted during nebulization of tobramycin high strength solution by novel and jet nebulizer delivery systems
YesBackground: Chronic infections with Pseudomonas aeruginosa are a leading cause of morbidity in patients
with cystic fibrosis (CF). The aim of tobramycin inhalation therapy in CF patients with chronic pulmonary
infection is to deliver high amounts of drug directly to the site of infection. TOBI® is a tobramycin
nebulizer solution (300 mg/5 ml) approved by FDA for maintenance therapy for patient with CF. The 20%
tobramycin sulfate solution was reported as the optimal and maximal concentration.
Methods: Nebulization of high strength tobramycin solution (20% tobramycin sulfate) (HSTS) has been
assessed in this study by using different selected high performance nebulizer delivery systems: two
different designs of jet nebulizers, and three new nebulizers based on vibrating mesh technology. The
aerosol particle size distribution and output characteristics were measured for in vitro performance
assessment of the nebulizer systems. The methodology was adapted from the current European standard,
EN 13544-1:2001E.
Results: The particle size distribution characteristic measurements showed that all tested nebulizers may
be suitable for inhalation of HSTS. The mean (SD) of highest percentage of fine particles (<5 mm) was
77.64 (2.3) % for Sidestream®, at flow rate 16 L/min. The highest respirable inhaled mass was for Pari LC
Plus® combined with PariBoyN® compressor, with mean (SD) 90.85 (8.6) mg. The mean (SD) of highest
drug wastage percentage was 63.9 (3.9) % for Sidestream® jet nebulizer combined with compressed air
cylinder at flow rate 16 L/min, while the lowest was 2.3 (0.26) % for NE-U22 Omron® (high frequency).
Conclusions: The HSTS can be nebulized by all tested nebulisers but the high frequency NE-U22 Omron®
and Aeroneb Go® are more efficient. When the HSTS compared to TOBI®, the respirable inhaled dose was
increased to more than 73%
“Trying, But Failing” : The Role of Inhaler Technique and Mode of Delivery in Respiratory Medication Adherence
Recognition of Commercial Support: The Expert Adherence Panel Meeting from which the concepts presented in this article were first discussed and the manuscript submission costs were supported by the Respiratory Effectiveness Group.Peer reviewedPublisher PD
Effectiveness of inhaler devices in adult asthma and COPD
Peer reviewedPublisher PD
Inhaler technique mastery and maintenance in healthcare professionals trained on different devices
Peer reviewedPostprin
Use of low-dose oral theophylline as an adjunct to inhaled corticosteroids in preventing exacerbations of chronic obstructive pulmonary disease: study protocol for a randomised controlled trial.
BACKGROUND: Chronic obstructive pulmonary disease (COPD) is associated with high morbidity, mortality, and health-care costs. An incomplete response to the anti-inflammatory effects of inhaled corticosteroids is present in COPD. Preclinical work indicates that 'low dose' theophylline improves steroid responsiveness. The Theophylline With Inhaled Corticosteroids (TWICS) trial investigates whether the addition of 'low dose' theophylline to inhaled corticosteroids has clinical and cost-effective benefits in COPD. METHOD/DESIGN: TWICS is a randomised double-blind placebo-controlled trial conducted in primary and secondary care sites in the UK. The inclusion criteria are the following: an established predominant respiratory diagnosis of COPD (post-bronchodilator forced expiratory volume in first second/forced vital capacity [FEV1/FVC] of less than 0.7), age of at least 40 years, smoking history of at least 10 pack-years, current inhaled corticosteroid use, and history of at least two exacerbations requiring treatment with antibiotics or oral corticosteroids in the previous year. A computerised randomisation system will stratify 1424 participants by region and recruitment setting (primary and secondary) and then randomly assign with equal probability to intervention or control arms. Participants will receive either 'low dose' theophylline (Uniphyllin MR 200 mg tablets) or placebo for 52 weeks. Dosing is based on pharmacokinetic modelling to achieve a steady-state serum theophylline of 1-5 mg/l. A dose of theophylline MR 200 mg once daily (or placebo once daily) will be taken by participants who do not smoke or participants who smoke but have an ideal body weight (IBW) of not more than 60 kg. A dose of theophylline MR 200 mg twice daily (or placebo twice daily) will be taken by participants who smoke and have an IBW of more than 60 kg. Participants will be reviewed at recruitment and after 6 and 12 months. The primary outcome is the total number of participant-reported COPD exacerbations requiring oral corticosteroids or antibiotics during the 52-week treatment period. DISCUSSION: The demonstration that 'low dose' theophylline increases the efficacy of inhaled corticosteroids in COPD by reducing the incidence of exacerbations is relevant not only to patients and clinicians but also to health-care providers, both in the UK and globally. TRIAL REGISTRATION: Current Controlled Trials ISRCTN27066620 was registered on Sept. 19, 2013, and the first subject was randomly assigned on Feb. 6, 2014
Characteristics of patients making serious inhaler errors with a dry powder inhaler and association with asthma-related events in a primary care setting
Acknowledgements The iHARP database was funded by unrestricted grants from Mundipharma International Ltd and Research in Real-Life Ltd; these analyses were funded by an unrestricted grant from Teva Pharmaceuticals. Mundipharma and Teva played no role in study conduct or analysis and did not modify or approve the manuscript. The authors wish to direct a special appreciation to all the participants of the iHARP group who contributed data to this study and to Mundipharma, sponsors of the iHARP group. In addition, we thank Julie von Ziegenweidt for assistance with data extraction and Anna Gilchrist and Valerie L. Ashton, PhD, for editorial assistance. Elizabeth V. Hillyer, DVM, provided editorial and writing support, funded by Research in Real-Life, Ltd.Peer reviewedPublisher PD
- …