    Measurement delay associated with the Guardian RT continuous glucose monitoring system.

    Get PDF
    AIMS: Using compartment modelling, we assessed the time delay between blood glucose and sensor glucose measured by the Guardian RT continuous glucose monitoring system in young subjects with Type 1 diabetes (T1D). METHODS: Twelve children and adolescents with T1D treated by continuous subcutaneous insulin infusion (male/female 7/5; age 13.1 +/- 4.2 years; body mass index 21.9 +/- 4.3 kg/m(2); mean +/- sd) were studied over 19 h in a Clinical Research Facility. Guardian RT was calibrated every 6 h and sensor glucose measured every 5 min. Reference blood glucose was measured every 15 min using a YSI 2300 STAT Plus Analyser. A population compartment model of sensor glucose-blood glucose kinetics was adopted to estimate the time delay, the calibration scale and the calibration shift. RESULTS: The population median of the time delay was 15.8 (interquartile range 15.2, 16.5) min, which was corroborated by correlation analysis between blood glucose and 15-min delayed sensor glucose. The delay has a relatively low intersubject variability, with 95% of individuals predicted to have delays between 10.4 and 24.3 min. Population medians (interquartile range) for the scale and shift are 0.800 (0.777, 0.823) (unitless) and 1.66 (1.47, 1.84) mmol/l, respectively. CONCLUSIONS: In young subjects with T1D, the total time delay associated with the Guardian RT system was approximately 15 min. This is twice that expected on physiological grounds, suggesting a 5- to 10-min delay because of data processing. Delays above 25 min are rarely to be observed

    Safety, efficacy and glucose turnover of reduced prandial boluses during closed-loop therapy in adolescents with type 1 diabetes: a randomized clinical trial.

    Get PDF
    AIMS: To evaluate safety, efficacy and glucose turnover during closed-loop with meal announcement using reduced prandial insulin boluses in adolescents with type 1 diabetes (T1D). METHODS: We conducted a randomized crossover study comparing closed-loop therapy with standard prandial insulin boluses versus closed-loop therapy with prandial boluses reduced by 25%. Eight adolescents with T1D [3 males; mean (standard deviation) age 15.9 (1.5) years, glycated haemoglobin 74 (17) mmol/mol; median (interquartile range) total daily dose 0.9 (0.7, 1.1) IU/kg/day] were studied on two 36-h-long visits. In random order, subjects received closed-loop therapy with either standard or reduced insulin boluses administered with main meals (50-80 g carbohydrates) but not with snacks (15-30 g carbohydrates). Stable-label tracer dilution methodology measured total glucose appearance (Ra_total) and glucose disposal (Rd). RESULTS: The median (interquartile range) time spent in target (3.9-10 mmol/l) was similar between the two interventions [74 (66, 84)% vs 80 (65, 96)%; p = 0.87] as was time spent above 10 mmol/l [21.8 (16.3, 33.5)% vs 18.0 (4.1, 34.2)%; p = 0.87] and below 3.9 mmol/l [0 (0, 1.5)% vs 0 (0, 1.8)%; p = 0.88]. Mean plasma glucose was identical during the two interventions [8.4 (0.9) mmol/l; p = 0.98]. Hypoglycaemia occurred once 1.5 h post-meal during closed-loop therapy with standard bolus. Overall insulin delivery was lower with reduced prandial boluses [61.9 (55.2, 75.0) vs 72.5 (63.6, 80.3) IU; p = 0.01] and resulted in lower mean plasma insulin concentration [186 (171, 260) vs 252 (198, 336) pmol/l; p = 0.002]. Lower plasma insulin was also documented overnight [160 (136, 192) vs 191 (133, 252) pmol/l; p = 0.01, pooled nights]. Ra_total was similar [26.3 (21.9, 28.0) vs 25.4 (21.0, 29.2) µmol/kg/min; p = 0.19] during the two interventions as was Rd [25.8 (21.0, 26.9) vs 25.2 (21.2, 28.8) µmol/kg/min; p = 0.46]. CONCLUSIONS: A 25% reduction in prandial boluses during closed-loop therapy maintains similar glucose control in adolescents with T1D whilst lowering overall plasma insulin levels. It remains unclear whether closed-loop therapy with a 25% reduction in prandial boluses would prevent postprandial hypoglycaemia.US National Institute of Diabetes and Digestive and Kidney Diseases (1R01DK085621). Support for the Artificial Pancreas research programme by the JDRF, Diabetes UK, NIHR Cambridge Biomedical Research Centre, and Wellcome Trust Strategic Award (100574/Z/12/Z) is acknowledged.This is the final version of the article. It first appeared from Wiley via http://dx.doi.org/10.1111/dom.1254

    Unsupervised home use of an overnight closed-loop system over 3-4 weeks: a pooled analysis of randomized controlled studies in adults and adolescents with type 1 diabetes.

    Get PDF
    AIMS: To compare overnight closed-loop and sensor-augmented pump therapy in patients with type 1 diabetes by combining data collected during free-living unsupervised randomized crossover home studies. METHODS: A total of 40 participants with type 1 diabetes, of whom 24 were adults [mean ± standard deviation (s.d.) age 43 ± 12 years and glycated haemoglobin (HbA1c) 8.0 ± 0.9%] and 16 were adolescents (mean ± s.d. age 15.6 ± 3.6 years and HbA1c 8.1 ± 0.8%), underwent two periods of sensor-augmented pump therapy in the home setting, in combination with or without an overnight closed-loop insulin delivery system that uses a model predictive control algorithm to direct insulin delivery. The order of the two interventions was random; each period lasted 4 weeks in adults and 3 weeks in adolescents. The primary outcome was time during which sensor glucose readings were in the target range of 3.9-8.0 mmol/l. RESULTS: The proportion of time when sensor glucose was in the target range (3.9-8.0 mmol/l) overnight (between 24:00 and 08:00 hours) was 18.5% greater during closed-loop insulin delivery than during sensor-augmented therapy (p < 0.001). Closed-loop therapy significantly reduced mean overnight glucose levels by 0.9 mmol/l (p < 0.001), with no difference in glycaemic variability, as measured by the standard deviation of sensor glucose. Time spent above the target range was reduced (p = 0.001), as was time spent in hypoglycaemia (<3.9 mmol/l; p = 0.014) during closed-loop therapy. Lower mean overnight glucose levels during closed-loop therapy were brought about by increased overnight insulin delivery (p < 0.001) without changes to the total daily delivery (p = 0.84). CONCLUSION: Overnight closed-loop insulin therapy at home in adults and adolescents with type 1 diabetes is feasible, showing improvements in glucose control and reducing the risk of nocturnal hypoglycaemia.Juvenile Diabetes Research Foundation (#22-2009-802) and Diabetes UK (BDA07/0003549) with additional support for the Artificial Pancreas work by National Institute of Diabetes and Digestive and Kidney Diseases (1R01DK085621), and National Institute for Health Research Cambridge Biomedical Research Centre. Abbott Diabetes Care supplied continuous glucose delivery devices and sensors and modified devices to facilitate real-time connectivity.This if the final version of the article. It was originally published by Wiley in Diabetes, Obesity and Metabolism at http://onlinelibrary.wiley.com/doi/10.1111/dom.12427/abstrac

    Safety and efficacy of 24-h closed-loop insulin delivery in well-controlled pregnant women with type 1 diabetes: a randomized crossover case series

    Get PDF
    OBJECTIVE: To evaluate the safety and efficacy of closed-loop insulin delivery in well-controlled pregnant women with type 1 diabetes treated with continuous subcutaneous insulin infusion (CSII). RESEARCH DESIGN AND METHODS: A total of 12 women with type 1 diabetes (aged 32.9 years, diabetes duration 17.6 years, BMI 27.1 kg/m(2), and HbA(1c) 6.4%) were randomly allocated to closed-loop or conventional CSII. They performed normal daily activities (standardized meals, snacks, and exercise) for 24 h on two occasions at 19 and 23 weeks' gestation. Plasma glucose time in target (63-140 mg/dL) and time spent hypoglycemic were calculated. RESULTS: Plasma glucose time in target was comparable for closed-loop and conventional CSII (median [interquartile range]: 81 [59-87] vs. 81% [54-90]; P = 0.75). Less time was spent hypoglycemic (<45 mg/dL [0.0 vs. 0.3%]; P = 0.04), with a lower low blood glucose index (2.4 [0.9-3.5] vs. 3.3 [1.9-5.1]; P = 0.03), during closed-loop insulin delivery. CONCLUSIONS: Closed-loop insulin delivery was as effective as conventional CSII, with less time spent in extreme hypoglycemia.This work was funded by a Diabetes UK project grant (BDA 07/003551). H.R.M. was funded by a National Institute for Health Research (NIHR) research fellowship (PDF/08/01/036). The research was conducted with support from the Juvenile Diabetes Research Foundation, Abbott Diabetes Care (FreeStyle Navigator continuous glucose monitors and sensors), the investigator-initiated study program of the Animas Corporation (Animas 2020 insulin pump), the Medical Research Council Center for Obesity and Related Metabolic Diseases, the NIHR Cambridge Biomedical Research Center, and Addenbrooke's Clinical Research Facility (Cambridge, U.K.). No funder had any role in the study design; data collection, analysis, and interpretation; or manuscript preparation

    Data-Driven Robust Control for Type 1 Diabetes Under Meal and Exercise Uncertainties

    Get PDF
    We present a fully closed-loop design for an artificial pancreas (AP) which regulates the delivery of insulin for the control of Type I diabetes. Our AP controller operates in a fully automated fashion, without requiring any manual interaction (e.g. in the form of meal announcements) with the patient. A major obstacle to achieving closed-loop insulin control is the uncertainty in those aspects of a patient's daily behavior that significantly affect blood glucose, especially in relation to meals and physical activity. To handle such uncertainties, we develop a data-driven robust model-predictive control framework, where we capture a wide range of individual meal and exercise patterns using uncertainty sets learned from historical data. These sets are then used in the controller and state estimator to achieve automated, precise, and personalized insulin therapy. We provide an extensive in silico evaluation of our robust AP design, demonstrating the potential of this approach, without explicit meal announcements, to support high carbohydrate disturbances and to regulate glucose levels in large clusters of virtual patients learned from population-wide survey data.Comment: Extended version of paper accepted at the 15th International Conference on Computational Methods in Systems Biolog

    Bolusing frequency and amount impacts glucose control during hybrid closed-loop.

    Get PDF
    AIM: To compare bolus insulin delivery patterns during closed-loop home studies in adults with suboptimally [HbA1c 58-86 mmol/mol (7.5%-10%)] and well-controlled [58 mmol/mol (< 7.5%)] Type 1 diabetes. METHODS: Retrospective analysis of daytime and night-time insulin delivery during home use of closed-loop over 4 weeks. Daytime and night-time controller effort, defined as amount of insulin delivered by closed-loop relative to usual basal insulin delivery, and daytime bolus effort, defined as total bolus insulin delivery relative to total daytime insulin delivery were compared between both cohorts. Correlation analysis was performed between individual bolus behaviour (bolus effort and frequency) and daytime controller efforts, and proportion of time spent within and below sensor glucose target range. RESULTS: Individuals with suboptimally controlled Type 1 diabetes had significantly lower bolus effort (P = 0.038) and daily bolus frequency (P < 0.001) compared with those with well-controlled diabetes. Controller effort during both daytime (P = 0.007) and night-time (P = 0.005) were significantly higher for those with suboptimally controlled Type 1 diabetes. Time when glucose was within the target range (3.9-10.0 mmol/L) during daytime correlated positively with bolus effort (r = 0.37, P = 0.016) and bolus frequency (r = 0.33, P = 0.037). Time when glucose was below the target range during daytime was comparable in both groups (P = 0.36), and did not correlate significantly with bolus effort (r = 0.28, P = 0.066) or bolus frequency (r = -0.21, P = 0.19). CONCLUSION: More frequent bolusing and higher proportion of insulin delivered as bolus during hybrid closed-loop use correlated positively with time glucose was in target range. This emphasises the need for user input and educational support to benefit from this novel therapeutic modality.Seventh Framework Programme of the European Union (ICT FP7- 247138). Additional support for the Artificial Pancreas work by JDRF, National Institute for Health Research Cambridge Biomedical Research Centre, Wellcome Strategic Award (100574/Z/12/Z), EC Horizon 2020 (H2020-SC1-731560), NIDDK (DP3DK112176 and 1UC4DK108520-01), Efficacy and Mechanism Evaluation Programme of National Institute for Health Research (14/23/09) and Helmsley Trust (2016PG-T1D045 and #2016PG-T1D046)

    Assessing the effectiveness of a 3-month day-and-night home closed-loop control combined with pump suspend feature compared with sensor-augmented pump therapy in youths and adults with suboptimally controlled type 1 diabetes: a randomised parallel study protocol

    Get PDF
    Introduction:\textbf{Introduction:} Despite therapeutic advances, many individuals with type 1 diabetes are unable to achieve tight glycaemic target without increasing the risk of hypoglycaemia. The objective of this study is to determine the effectiveness of a 3-month day-and-night home closed-loop glucose control combined with a pump suspend feature, compared with sensor-augmented insulin pump therapy in youths and adults with suboptimally controlled type 1 diabetes. Methods and analysis:\textbf{Methods and analysis:} The study adopts an open-label, multi-centre, multi-national (UK and USA), randomised, single-period, parallel design and aims for 84 randomised patients. Participants are youths (6-21 years) or adults (>21 years) with type 1 diabetes treated with insulin pump therapy and suboptimal glycaemic control (glycated haemoglobin (HbA1c) ≥7.5% (58 mmol/mol) and ≤10% (86 mmol/mol)). Following a 4-week run-in period, eligible participants will be randomised to a 3-month use of automated closed-loop insulin delivery combined with pump suspend feature or to sensor-augmented insulin pump therapy. Analyses will be conducted on an intention-to-treat basis. The primary outcome is the time spent in the target glucose range from 3.9 to 10.0 mmol/L based on continuous glucose monitoring levels during the 3-month free-living phase. Secondary outcomes include HbA1c at 3 months, mean glucose, time spent below and above target; time with glucose levels 16.7 mmol/L, glucose variability; total, basal and bolus insulin dose and change in body weight. Participants' and their families' perception in terms of lifestyle change, daily diabetes management and fear of hypoglycaemia will be evaluated. Ethics and dissemination:\textbf{Ethics and dissemination:} Ethics/institutional review board approval has been obtained. Before screening, all participants/guardians will be provided with oral and written information about the trial. The study will be disseminated by peer-reviewed publications and conference presentations. Trial registration number:\textbf{Trial registration number:} NCT02523131; Pre-results.JDRF, National Institute for Health Research Cambridge Biomedical Research Centre, Wellcome Strategic Award (100574/Z/12/Z)

    Home Use of an Artificial Beta Cell in Type 1 Diabetes.

    Get PDF
    BACKGROUND: The feasibility, safety, and efficacy of prolonged use of an artificial beta cell (closed-loop insulin-delivery system) in the home setting have not been established. METHODS: In two multicenter, crossover, randomized, controlled studies conducted under free-living home conditions, we compared closed-loop insulin delivery with sensor-augmented pump therapy in 58 patients with type 1 diabetes. The closed-loop system was used day and night by 33 adults and overnight by 25 children and adolescents. Participants used the closed-loop system for a 12-week period and sensor-augmented pump therapy (control) for a similar period. The primary end point was the proportion of time that the glucose level was between 70 mg and 180 mg per deciliter for adults and between 70 mg and 145 mg per deciliter for children and adolescents. RESULTS: Among adults, the proportion of time that the glucose level was in the target range was 11.0 percentage points (95% confidence interval [CI], 8.1 to 13.8) greater with the use of the closed-loop system day and night than with control therapy (P<0.001). The mean glucose level was lower during the closed-loop phase than during the control phase (difference, -11 mg per deciliter; 95% CI, -17 to -6; P<0.001), as were the area under the curve for the period when the glucose level was less than 63 mg per deciliter (39% lower; 95% CI, 24 to 51; P<0.001) and the mean glycated hemoglobin level (difference, -0.3%; 95% CI, -0.5 to -0.1; P=0.002). Among children and adolescents, the proportion of time with the nighttime glucose level in the target range was higher during the closed-loop phase than during the control phase (by 24.7 percentage points; 95% CI, 20.6 to 28.7; P<0.001), and the mean nighttime glucose level was lower (difference, -29 mg per deciliter; 95% CI, -39 to -20; P<0.001). The area under the curve for the period in which the day-and-night glucose levels were less than 63 mg per deciliter was lower by 42% (95% CI, 4 to 65; P=0.03). Three severe hypoglycemic episodes occurred during the closed-loop phase when the closed-loop system was not in use. CONCLUSIONS: Among patients with type 1 diabetes, 12-week use of a closed-loop system, as compared with sensor-augmented pump therapy, improved glucose control, reduced hypoglycemia, and, in adults, resulted in a lower glycated hemoglobin level. (Funded by the JDRF and others; AP@home04 and APCam08 ClinicalTrials.gov numbers, NCT01961622 and NCT01778348.).JDRF (22-2011-668) and Seventh Framework Programme of the European Union (ICT FP7- 247138). Additional support for the Artificial Pancreas work by National Institute for Health Research Cambridge Biomedical Research Centre and Wellcome Strategic Award (100574/Z/12/Z). Abbott Diabetes Care supplied discounted continuous glucose monitoring devices, sensors, and details of communication protocol to facilitate real-time connectivity. Diasend provided discounted platform for data upload.This is the final version of the article. It was first available from Massachusetts Medical Society via http://dx.doi.org/10.1056/NEJMoa150935