Modelling of glucose and insulin kinetics to facilitate the development of a wearable artificial pancreas
A wearable artificial pancreas (AP) has been a research goal for over three decades. The aim of this device is to establish effective closed-loop control of blood glucose in patients with type 1 diabetes mellitus (T1DM) using subcutaneous (SC) glucose measurements and SC insulin delivery. The main difficulties that hamper the successful development of a wearable AP concern the stability and accuracy of SC glucose sensing, the predictability of the absorption kinetics of the injected insulin, and finally the performance of the glucose control algorithm. As clinical tests on humans are costly, time consuming, and demand ethical approval, in silico testing of the AP has become a critical feature to facilitate an accelerated development of the AP and, specifically, the control algorithm.
The primary aim of the work reported in this thesis was to explore the use of compartmental modelling techniques with in-built physiological constraints to facilitate the development of a wearable AP.
In particular, the study aimed to extend and evaluate an existing model of whole-body glucose kinetics on a set of data obtained in a clinical trial designed to test the AP algorithm. The model was extended to represent the input-output relationship between the SC insulin and intravenous glucose concentrations. The extended model was re-evaluated in subjects with T1DM under new conditions with the objective to obtain sets of parameters to represent âvirtualâ subjects with T1DM in the AP simulator. The parameter estimation was completed, but the âvirtualâ subjects for use in the AP simulator could not be generated due to the uncertain validity of the tested model.
Further objectives included the support for in silico testing of an AP through investigating insulin lispro and interstitial glucose kinetics.
To explore the kinetics of SC administered insulin lispro, ten competing models were proposed assuming a number of physiological effects. The principle of parsimony was used to select the model, which best represented our data. The best model included slow and fast absorption channels, and the presence of local insulin degradation at the injection site.
In order to establish the relationship between the interstitial glucose (IG) and plasma glucose (PG), nine models of IG kinetics were postulated. The model which best represented the experimental data was selected using the principle of parsimony. Two mechanisms explaining the temporal variation in the IG-PG ratio were identified, a zero-order removal of glucose from the interstitial fluid (ISF) and the stimulatory effect of insulin on glucose transfer from the plasma to the ISF. This best model found its use in the simulator to represent SC glucose measurements.
In conclusion, valuable insights were obtained into the mechanisms involved in the insulin and interstitial glucose kinetics, as well as the whole-body glucose kinetics
Recommended from our members
Adaptability of Closed-Loop during Labour, Delivery and Postpartum: A secondary analysis of data from two randomized crossover trials in type 1 diabetes pregnancy
Background: Tight glucose control during labour and delivery is recommended for pregnant women with type 1 diabetes. This can be challenging to achieve using the current treatment modalities. The automated nature of closed-loop and its ability to adapt to real-time glucose levels make it well suited for use during labour, delivery and the immediate postpartum period.
Methods: We report observational data of participants from two randomized crossover trials who chose to continue using closed-loop during labour, delivery and postpartum. Labour was defined as the 24 hours prior to delivery and postpartum as the 48 hours after delivery. The glucose target range during pregnancy was 3.5-7.8mmol/L (63-140mg/dL) and 3.9-10mmol/L (70-180mg/dL) after delivery.
Results: Twenty-seven (84.4%) of the potential 32 trial participants used closed-loop through labor, delivery, and postpartum. Use of closed-loop was associated with 82.0% (IQR 49.3,93.0) time-in-target range during labor and delivery and a mean glucose of 6.9±1.4mmol/L (124±25mg/dL). Closed-loop performed well throughout vaginal, elective and emergency caesarean section deliveries. Postpartum, women spent 83.3% (IQR 75.2,94.6) time-in-target range (3.9-10.0mmol/L [70-180mg/dL]), with a mean glucose of 7.2±1.4mmol/L (130±25mg/dL). There was no difference in maternal glucose concentration between mothers of infants with and without neonatal hypoglycemia (6.9±1.6 and 6.8±1.1 mmol/L [124±29 and 122±20mg/dL] respectively; p=0.84).
Conclusions: Automated closed-loop insulin delivery is feasible during hospital admissions for labour, delivery and postpartum. Larger scale studies are needed to evaluate its efficacyThe trials were funded by the National Institute for Health Research (HRM Career Development Fellowship, CDF-2013-06-035), Diabetes UK (BDA 07/0003551), Gates Cambridge Trust PhD fellowship (ZAS), Jean Hailes for Womenâs Health (ZAS); Allen-Carey Scholarship in Womenâs Health (JMY) and a grant from the NIHR Cambridge Biomedical Research Centre (RH). Abbott Diabetes Care supplied discounted CGM devices, sensors, and details of communication protocol to facilitate real-time connectivity. HRM conducts independent research supported by the National Institute for Health Research
Pharmacokinetics of diluted (U20) insulin aspart compared with standard (U100) in children aged 3-6 years with type 1 diabetes during closed-loop insulin delivery: a randomised clinical trial.
AIMS/HYPOTHESIS: The aim of this study was to compare the pharmacokinetics of two different concentrations of insulin aspart (B28Asp human insulin) in children aged 3-6 years with type 1 diabetes. METHODS: Young children with type 1 diabetes underwent an open-label, randomised, two-period crossover study in a clinical research facility, 2-6 weeks apart. In random order, diluted (1:5 dilution with saline [154 mmol/l NaCl]; 20 U/ml) or standard strength (100 U/ml) insulin aspart was administered via an insulin pump as a meal bolus and then overnight by closed-loop insulin delivery as determined by a model predictive algorithm. Plasma insulin was measured every 30-60 min from 17:00 hours on day 1 to 8:00 hours on day 2. We measured the time-to-peak insulin concentration (tmax), insulin metabolic clearance rate (MCR(I)) and background insulin concentration (ins(c)) using compartmental modelling. RESULTS: Eleven children (six male; age range 3.75-6.96 years, HbA1c 7.6% ± 1.3% [60 ± 14 mmol/mol], BMI standard deviation score 1.0 ± 0.8, duration of diabetes 2.2 ± 1.0 years, total daily dose 12.9 [10.6-16.5] U, fasting C-peptide concentration 5 [5-17.1] pmol/l; mean ± SD or median [interquartile range]) participated in the study. No differences between standard and diluted insulin were observed in terms of t max (59.2 ± 14.4 vs 61.6 ± 8.7) min for standard vs diluted, p = 0.59; MCR I (1.98 à 10(-2) ± 0.99 à 10(-2) vs 1.89 à 10(-2) ± 0.82 à 10(-2) 1/kg/min, p = 0.47), and ins c (34 [1-72] vs 23 [3-65] pmol/l, p = 0.66). However, t max showed less intersubject variability following administration of diluted aspart (SD 14.4 vs 8.7 min, p = 0.047). CONCLUSIONS/INTERPRETATION: Diluting insulin aspart does not change its pharmacokinetics. However, it may result in less variable absorption and could be used in young children with type 1 diabetes undergoing closed-loop insulin delivery. TRIAL REGISTRATION: Clinicaltrials.gov NCT01557634. FUNDING: FUNDING was provided by the JDRF, 7th Framework Programme of the European Union, Wellcome Trust Strategic Award and the National Institute for Health Research Cambridge Biomedical Research Centre.Funding was provided by the JDRF (grant number 22-2011-
668), 7th Framework Programme of the European Union (Spidiman
project; grant agreement number 305343), Wellcome Trust Strategic
Award (100574/Z/12/Z) and the National Institute for Health Research
Cambridge Biomedical Research Centre.This is the final published version. It first appeared at http://link.springer.com/article/10.1007%2Fs00125-014-3483-6
Home Use of Day-and-Night Hybrid Closed-Loop Insulin Delivery in Suboptimally Controlled Adolescents With Type 1 Diabetes: A 3-Week, Free-Living, Randomized Crossover Trial.
OBJECTIVE: This study evaluated the feasibility, safety, and efficacy of day-and-night hybrid closed-loop insulin delivery in adolescents with type 1 diabetes under free-living conditions. RESEARCH DESIGN AND METHODS: In an open-label randomized crossover study, 12 suboptimally controlled adolescents on insulin pump therapy (mean ± SD age 14.6 ± 3.1 years; HbA1c 69 ± 8 mmol/mol [8.5 ± 0.7%]; duration of diabetes 7.8 ± 3.5 years) underwent two 21-day periods in which hybrid closed-loop insulin delivery was compared with sensor-augmented insulin pump therapy in random order. During the closed-loop intervention, a model predictive algorithm automatically directed insulin delivery between meals and overnight. Participants used a bolus calculator to administer prandial boluses. RESULTS: The proportion of time that sensor glucose was in the target range (3.9-10 mmol/L; primary end point) was increased during the closed-loop intervention compared with sensor-augmented insulin pump therapy by 18.8 ± 9.8 percentage points (mean ± SD; P < 0.001), the mean sensor glucose level was reduced by 1.8 ± 1.3 mmol/L (P = 0.001), and the time spent above target was reduced by 19.3 ± 11.3 percentage points (P < 0.001). The time spent with sensor glucose levels below 3.9 mmol/L was low and comparable between interventions (median difference 0.4 [interquartile range -2.2 to 1.3] percentage points; P = 0.33). Improved glucose control during closed-loop was associated with increased variability of basal insulin delivery (P < 0.001) and an increase in the total daily insulin dose (53.5 [39.5-72.1] vs. 51.5 [37.6-64.3] units/day; P = 0.006). Participants expressed positive attitudes and experience with the closed-loop system. CONCLUSIONS: Free-living home use of day-and-night closed-loop in suboptimally controlled adolescents with type 1 diabetes is safe, feasible, and improves glucose control without increasing the risk of hypoglycemia. Larger and longer studies are warranted.National Institute of Diabetes and Digestive and Kidney Diseases (Grant ID: 1R01DK085621-01), JDRF, National Institute for Health Research Cambridge Biomedical Research Centre, Wellcome Trust (Strategic Award: 100574/Z/12/Z)This is the author accepted manuscript. The final version is available from American Diabetes Association via http://dx.doi.org/10.2337/dc16-109
Sensor Life and Overnight Closed Loop: A Randomized Clinical Trial.
BACKGROUND: Closed-loop (CL) systems direct insulin delivery based on continuous glucose monitor (CGM) sensor values. CGM accuracy varies with sensor life, being least accurate on day 1 of sensor insertion. We evaluated the effect of sensor life (enhanced Enlite, Medtronic MiniMed, Northridge, CA) on overnight CL. METHODS: In an open-label, randomized, 2-period, inpatient crossover pilot study, 12 adolescents on insulin pump (age 16.7 ± 1.9 years; HbA1c 66 ± 10 mmol/mol) attended a clinical research facility on 2 overnight occasions. In random order, participants received CL on day 1 or on day 3-4 after sensor insertion. During both periods, glucose was automatically controlled by a model predictive control algorithm informed by sensor glucose. Plasma glucose was measured every 30 to 60 min. RESULTS: During overnight CL (22:30 to 07:30), the proportion of time with plasma glucose readings in the target range (3.9-8.0 mmol/l, primary endpoint) when initiated on day 1 of sensor insertion vs day 3-4 were comparable (58 ± 32% day 1 vs 56 ± 36% day 3-4; P = .34), and there were no significant differences between interventions in terms of mean plasma glucose ( P = .26), percentage time above 8.0 mmol/l ( P = .49), and time spent below 3.9 mmol/l ( P = .93). Sensor accuracy varied with sensor life (mean absolute relative difference 19.8 ± 15.0% on day 1 and 13.7 ± 10.2% on day 3 to 4). Sensor glucose tended to under-read plasma glucose inflating benefits of CL on glucose control. CONCLUSIONS: In spite of differences in sensor accuracy, overnight CL glucose control informed by sensor glucose on day 1 or day 3-4 after sensor insertion was comparable. The model predictive controller appears to mitigate against sensor inaccuracies.This work was funded by the JDRF (#22-2011-668). Additional support for the Artificial Pancreas work by National Institute for Health Research Cambridge Biomedical Research Centre and Wellcome Strategic Award (100574/Z/12/Z). Medtronic supplied study pump, translator device, sensor transmitter, Amber user interface, and supported regulatory approval
Closed-loop insulin delivery in inpatients with type 2 diabetes: a randomised, parallel-group trial.
BACKGROUND: We assessed whether fully closed-loop insulin delivery (the so-called artificial pancreas) is safe and effective compared with standard subcutaneous insulin therapy in patients with type 2 diabetes in the general ward. METHODS: For this single-centre, open-label, parallel-group, randomised controlled trial, we enrolled patients aged 18 years or older with type 2 diabetes who were receiving insulin therapy. Patients were recruited from general wards at Addenbrooke's Hospital, Cambridge, UK. Participants were randomly assigned (1:1) by a computer-generated minimisation method to receive closed-loop insulin delivery (using a model-predictive control algorithm to direct subcutaneous delivery of rapid-acting insulin analogue without meal-time insulin boluses) or conventional subcutaneous insulin delivery according to local clinical guidelines. The primary outcome was time spent in the target glucose concentration range of 5·6-10·0 mmol/L during the 72 h study period. Analyses were by intention to treat. This trial is registered with ClinicalTrials.gov, number NCT01774565. FINDINGS: Between Feb 20, 2015, and March 24, 2016, we enrolled 40 participants, of whom 20 were randomly assigned to the closed-loop intervention group and 20 to the control group. The proportion of time spent in the target glucose range was 59·8% (SD 18·7) in the closed-loop group and 38·1% (16·7) in the control group (difference 21·8% [95% CI 10·4-33·1]; p=0·0004). No episodes of severe hypoglycaemia or hyperglycaemia with ketonaemia occurred in either group. One adverse event unrelated to study devices occurred during the study (gastrointestinal bleed). INTERPRETATION: Closed-loop insulin delivery without meal-time boluses is effective and safe in insulin-treated adults with type 2 diabetes in the general ward. FUNDING: Diabetes UK; European Foundation for the Study of Diabetes; JDRF; National Institute for Health Research Cambridge Biomedical Research Centre; Wellcome Trust.This study was supported by Diabetes UK (#14/0004878) and the European Foundation for the Study of Diabetes. Additional support for research on the artificial pancreas was received from the Juvenile Diabetes Research Foundation, National Institute for Health Research Cambridge Biomedical Research Centre, and Wellcome Strategic Award (100574/Z/12/Z).This is the final version of the article. It first appeared from Elsevier via https://doi.org/10.1016/S2213-8587(16)30280-
CamAPS FX hybrid closed-loop with ultra-rapid lispro compared with standard lispro in adults with type 1 diabetes: a double-blind, randomized, crossover study.
INTRODUCTION
To evaluate hybrid closed-loop with ultra-rapid insulin lispro (Lyumjev) compared with hybrid closed-loop with standard insulin lispro in adults with type 1 diabetes.
MATERIALS AND METHODS
In a single-center, double-blind, randomized, crossover study, 28 adults with type 1 diabetes (mean±SD: age 44.5±10.7, HbA1c 7.1±0.9% [54±10mmol/mol]) underwent two 8-week periods comparing hybrid closed-loop with ultra-rapid insulin lispro and hybrid closed-loop with standard insulin lispro in random order. CamAPS FX closed-loop system was used in both periods.
RESULTS
In an intention-to-treat analysis, the proportion of time sensor glucose was in target range (3.9 to 10mmol/L; primary endpoint) was greater with ultra-rapid lispro compared with standard insulin lispro (mean±SD: 78.7±9.8% vs. 76.2±9.6%; mean difference 2.5 percentage points [95%CI 0.8 to 4.2]; p=0.005). Mean sensor glucose was lower with ultra-rapid lispro compared with standard insulin lispro (7.9±0.8mmol/L vs. 8.1±0.9mmol/L; p=0.048). The proportion of time with sensor glucose <3.9mmol/L was similar between interventions (median [IQR] ultra-rapid lispro 2.3% [1.3-2.7%] vs. standard insulin lispro 2.1% [1.4-3.3%]; p=0.33). No severe hypoglycemia or ketoacidosis occurred.
CONCLUSIONS
The use of ultra-rapid lispro with CamAPS FX hybrid closed-loop increases time in range and reduces mean glucose with no difference in hypoglycemia compared with standard insulin lispro in adults with type 1 diabetes
Shortâterm fully closedâloop insulin delivery using faster insulin aspart compared to standard insulin aspart in type 2 diabetes
We evaluated the efficacy and safety of shortâterm fully closedâloop insulin delivery using faster versus standard insulin aspart in type 2 diabetes. Fifteen adults with insulinâtreated type 2 diabetes underwent 22 hours of closedâloop insulin delivery with either faster or standard insulin aspart in a doubleâblind randomised crossover design. Basalâbolus regimen was replaced by model predictive control algorithmâdirected insulin delivery based on sensor glucose levels. The primary outcome was time with plasma glucose in target range (5.6â10.0mmol/l) and did not differ between treatments (mean difference [95%CI] â3.3% [8.2;1.7], p=0.17). Mean glucose and glucose variability were comparable, as was time spent below and above target range. Hypoglycaemia (<3.5mmol/l) occurred once with faster insulin aspart and twice with standard insulin aspart. Mean total insulin dose was higher with faster insulin aspart (mean difference [95%CI] 3.7U [0.7;6.8], p=0.021). No episodes of severe hypoglycaemia or other serious adverse events occurred. In conclusion, shortâterm fully closedâloop in type 2 diabetes may require higher dose of faster insulin aspart compared to standard insulin aspart to achieve comparable glucose control.Swiss National Science Foundation (P1BEP3_165297), UDEM Scientific Fund, Cambridge Biomedical Research Centre - NIHR
Safety, efficacy and glucose turnover of reduced prandial boluses during closed-loop therapy in adolescents with type 1 diabetes: a randomized clinical trial.
AIMS: To evaluate safety, efficacy and glucose turnover during closed-loop with meal announcement using reduced prandial insulin boluses in adolescents with type 1 diabetes (T1D). METHODS: We conducted a randomized crossover study comparing closed-loop therapy with standard prandial insulin boluses versus closed-loop therapy with prandial boluses reduced by 25%. Eight adolescents with T1D [3 males; mean (standard deviation) age 15.9 (1.5) years, glycated haemoglobin 74 (17) mmol/mol; median (interquartile range) total daily dose 0.9 (0.7, 1.1) IU/kg/day] were studied on two 36-h-long visits. In random order, subjects received closed-loop therapy with either standard or reduced insulin boluses administered with main meals (50-80 g carbohydrates) but not with snacks (15-30 g carbohydrates). Stable-label tracer dilution methodology measured total glucose appearance (Ra_total) and glucose disposal (Rd). RESULTS: The median (interquartile range) time spent in target (3.9-10 mmol/l) was similar between the two interventions [74 (66, 84)% vs 80 (65, 96)%; p = 0.87] as was time spent above 10 mmol/l [21.8 (16.3, 33.5)% vs 18.0 (4.1, 34.2)%; p = 0.87] and below 3.9 mmol/l [0 (0, 1.5)% vs 0 (0, 1.8)%; p = 0.88]. Mean plasma glucose was identical during the two interventions [8.4 (0.9) mmol/l; p = 0.98]. Hypoglycaemia occurred once 1.5 h post-meal during closed-loop therapy with standard bolus. Overall insulin delivery was lower with reduced prandial boluses [61.9 (55.2, 75.0) vs 72.5 (63.6, 80.3) IU; p = 0.01] and resulted in lower mean plasma insulin concentration [186 (171, 260) vs 252 (198, 336) pmol/l; p = 0.002]. Lower plasma insulin was also documented overnight [160 (136, 192) vs 191 (133, 252) pmol/l; p = 0.01, pooled nights]. Ra_total was similar [26.3 (21.9, 28.0) vs 25.4 (21.0, 29.2) ”mol/kg/min; p = 0.19] during the two interventions as was Rd [25.8 (21.0, 26.9) vs 25.2 (21.2, 28.8) ”mol/kg/min; p = 0.46]. CONCLUSIONS: A 25% reduction in prandial boluses during closed-loop therapy maintains similar glucose control in adolescents with T1D whilst lowering overall plasma insulin levels. It remains unclear whether closed-loop therapy with a 25% reduction in prandial boluses would prevent postprandial hypoglycaemia.US National Institute of Diabetes and Digestive and Kidney Diseases (1R01DK085621). Support for the Artificial Pancreas research programme by the JDRF, Diabetes UK, NIHR Cambridge Biomedical Research Centre, and Wellcome Trust Strategic Award (100574/Z/12/Z) is acknowledged.This is the final version of the article. It first appeared from Wiley via http://dx.doi.org/10.1111/dom.1254
Closed-loop insulin delivery during pregnancy complicated by type 1 diabetes.
OBJECTIVE: This study evaluated closed-loop insulin delivery with a model predictive control (MPC) algorithm during early (12-16 weeks) and late gestation (28-32 weeks) in pregnant women with type 1 diabetes. RESEARCH DESIGN AND METHODS: Ten women with type 1 diabetes (age 31 years, diabetes duration 19 years, BMI 24.1 kg/m(2), booking A1C 6.9%) were studied over 24 h during early (14.8 weeks) and late pregnancy (28.0 weeks). A nurse adjusted the basal insulin infusion rate from continuous glucose measurements (CGM), fed into the MPC algorithm every 15 min. Mean glucose and time spent in target (63-140 mg/dL), hyperglycemic (>140 to â„ 180 mg/dL), and hypoglycemic (140 mg/dL) was 7% (0-40%) in early and 0% (0-6%) in late pregnancy (P = 0.25) and hypoglycemic (<63 mg/dL) was 0% (0-3%) and 0% (0-0%), respectively (P = 0.18). Postprandial glucose control, glucose variability, insulin infusion rates, and CGM sensor accuracy were no different in early or late pregnancy. CONCLUSIONS: MPC algorithm performance was maintained throughout pregnancy, suggesting that overnight closed-loop insulin delivery could be used safely during pregnancy. More work is needed to achieve optimal postprandial glucose control.Diabetes UK (07/0003551), TCC (PDF/01/036), MRC (G0600717