Lymph Node Characterization in Vivo Using Endoscopic Ultrasound Spectrum Analysis With Electronic Array Echo Endoscopes
Our purpose was to demonstrate the use of radiofrequency spectral analysis to distinguish between benign and malignant lymph nodes with data obtained using electronic array echo endoscopes, as we have done previously using mechanical echo endoscopes. In a prospective study, images were obtained from eight patients with benign-appearing lymph nodes and 11 with malignant lymph nodes, as verified by fine-needle aspiration. Midband fit, slope, intercept, correlation coefficient, and root-mean-square (RMS) deviation from a linear regression of the calibrated power spectra were determined and compared between the groups. Significant differences were observable for mean midband fit, intercept, and RMS deviation (t test P \u3c 0.05). For benign (n = 16) vs. malignant (n = 12) lymph nodes, midband fit and RMS deviation provided classification with 89 % accuracy and area under receiver operating characteristic (ROC) curve of 0.95 based on linear discriminant analysis. We concluded that the mean spectral parameters of the backscattered signals from electronic array echo endoscopy can provide a noninvasive method to quantitatively discriminate between benign and malignant lymph nodes
Characterization of the Pancreas in Vivo Using EUS Spectrum Analysis with Electronic Array Echoendoscopes
Background
Spectral analysis of the radiofrequency (RF) signals that underlie grayscale EUS images has been used to provide quantitative, objective information about tissue histology.
Objective
Our purpose was to validate RF spectral analysis as a method to distinguish between chronic pancreatitis (CP) and pancreatic cancer (PC).
Design and Setting
A prospective study of eligible patients was conducted to analyze the RF data obtained by using electronic array echoendoscopes.
Patients
Pancreatic images were obtained by using electronic array echoendoscopes from 41 patients in a prospective study, including 15 patients with PC, 15 with CP, and 11 with a normal pancreas.
Main Outcome Measurements
Midband fit, slope, intercept, correlation coefficient, and root mean square deviation from a linear regression of the calibrated power spectra were determined and compared among the groups.
Results
Statistical analysis showed that significant differences were observable between groups for mean midband fit, intercept, and root mean square deviation (t test, P \u3c .05). Discriminant analysis of these parameters was then performed to classify the data. For CP (n = 15) versus PC (n = 15), the same parameters provided 83% accuracy and an area under the curve of 0.83.
Limitations
Moderate sample size and spatial averaging inherent in the technique.
Conclusions
This study shows that mean spectral parameters of the backscattered signals obtained by using electronic array echoendoscopes can provide a noninvasive method to quantitatively discriminate between CP and PC