62 research outputs found

    Small field dosimetry: experimental methods and Monte Carlo simulation in small field radiation therapy dosimetry

    Get PDF
    The goal of the thesis was to investigate, and better define, what the requirements are for accurate small field relative dosimetry. Diode detector selection and experimental techniques were evaluated. EGSnrc Monte Carlo simulations were used to predict diode detector dosimetric parameters and assist in interpreting measured data. An emerging scintillator based detector technology was also tested and methods developed to standardize the reporting of small field dosimetric data. Using careful experimental methods the relative output uncertainty for the smallest square field size of side 0.5 cm was reduced to better than ±1.00% for all detector types. Monte Carlo simulation data revealed that for the same small field size the relative output measured using unshielded and shielded diodes will be 5% and 10% greater than the actual relative output in water. Further simulation work showed that simplified diode detector models are valid for use in small field dosimetry simulations. The diode detector over-response was also shown to be insensitive to variations in the electron energy and spot size incident on the Bremsstrahlung target. Experimental methods were refined to include the definition of an effective field size, which was shown to remove much of the ambiguity in reporting small field relative output data across a population of linear accelerators. Each of the for mentioned areas of investigation have been shown to be requirements for accurate small field relative

    Characterization of ELEKTA SRS cone collimator using high spatial resolution monolithic silicon detector array

    Get PDF
    Purpose: To investigate the accuracy of the dosimetry of radiation fields produced by small ELEKTA cone collimators used for stereotactic radiosurgery treatments (SRS) using commercially available detectors EBT3 Gafchromic TM film, IBA Stereotactic diode (SFD), and the recently developed detector DUO, which is a monolithic silicon orthogonal linear diode array detector. Methods: These three detectors were used for the measurement of beam profiles, output factors, and percentage depth dose for SRS cone collimators with cone sizes ranging from 5 to 50 mm diameter. The measurements were performed at 10 cm depth and 90 cm SSD. Results: The SRS cone beam profiles measured with DUO, EBT3 film, and IBA SFD agreed well, results being in agreement within ±0.5 mm in the FWHM, and ±0.7 mm in the penumbra region. The output factor measured by DUO with 0.5 mm air gap above agrees within ±1% with EBT3. The OF measured by IBA SFD (corrected for the over-response) agreed with both EBT3 and DUO within ±2%. All three detectors agree within ±2% for PDD measurements for all SRS cones. Conclusions: The characteristics of the ELEKTA SRS cone collimator have been evaluated by using a monolithic silicon high spatial resolution detector DUO, EBT3, and IBA SFD diode. The DUO detector is suitable for fast real-time quality assurance dosimetry in small radiation fields typical for SRS/SRT. This has been demonstrated by its good agreement of measured doses with EBT 3 films

    A practical and theoretical definition of very small field size for radiotherapy output factor measurements

    No full text
    Purpose This work introduces the concept of very small field size. Output factor (OPF) measurements at these field sizes require extremely careful experimental methodology including the measurement of dosimetric field size at the same time as each OPF measurement. Two quantifiable scientific definitions of the threshold of very small field size are presented. Methods A practical definition was established by quantifying the effect that a 1 mm error in field size or detector position had on OPFs, and setting acceptable uncertainties on OPF at 1%. Alternatively, for a theoretical definition of very small field size, the OPFs were separated into additional factors to investigate the specific effects of lateral electronic disequilibrium, photon scatter in the phantom and source occlusion. The dominant effect was established and formed the basis of a theoretical definition of very small fields. Each factor was obtained using Monte Carlo simulations of a Varian iX linear accelerator for various square field sizes of side length from 4 mm to 100 mm, using a nominal photon energy of 6 MV. Results According to the practical definition established in this project, field sizes < 15 mm were considered to be very small for 6 MV beams for maximal field size uncertainties of 1 mm. If the acceptable uncertainty in the OPF was increased from 1.0 % to 2.0 %, or field size uncertainties are 0.5 mm, field sizes < 12 mm were considered to be very small. Lateral electronic disequilibrium in the phantom was the dominant cause of change in OPF at very small field sizes. Thus the theoretical definition of very small field size coincided to the field size at which lateral electronic disequilibrium clearly caused a greater change in OPF than any other effects. This was found to occur at field sizes < 12 mm. Source occlusion also caused a large change in OPF for field sizes < 8 mm. Based on the results of this study, field sizes < 12 mm were considered to be theoretically very small for 6 MV beams. Conclusions Extremely careful experimental methodology including the measurement of dosimetric field size at the same time as output factor measurement for each field size setting and also very precise detector alignment is required at field sizes at least < 12 mm and more conservatively < 15 mm for 6 MV beams. These recommendations should be applied in addition to all the usual considerations for small field dosimetry, including careful detector selection

    A methodological approach to reporting corrected small field relative outputs

    Get PDF
    Purpose The goal of this work was to set out a methodology for measuring and reporting small field relative output and to assess the application of published correction factors across a population of linear accelerators. Methods and materials Measurements were made at 6 MV on five Varian iX accelerators using two PTW T60017 unshielded diodes. Relative output readings and profile measurements were made for nominal square field sizes of side 0.5 to 1.0 cm. The actual in-plane (A) and cross-plane (B) field widths were taken to be the FWHM at the 50% isodose level. An effective field size, defined as FSeff=A·B, was calculated and is presented as a field size metric. FSeffFSeff was used to linearly interpolate between published Monte Carlo (MC) calculated kQclin,Qmsrfclin,fmsr values to correct for the diode over-response in small fields. Results The relative output data reported as a function of the nominal field size were different across the accelerator population by up to nearly 10%. However, using the effective field size for reporting showed that the actual output ratios were consistent across the accelerator population to within the experimental uncertainty of ±1.0%. Correcting the measured relative output using kQclin,Qmsrfclin,fmsr at both the nominal and effective field sizes produce output factors that were not identical but differ by much less than the reported experimental and/or MC statistical uncertainties. Conclusions In general, the proposed methodology removes much of the ambiguity in reporting and interpreting small field dosimetric quantities and facilitates a clear dosimetric comparison across a population of linac

    An experimental approach to air gap optimisation for a 'correction-less' small field diode [Conference Abstract]

    Get PDF
    Aim A recent Monte Carlo based study has shown that it is possible to design a diode that measures small field output factors equivalent to that in water. This is accomplished by placing an appropriate sized air gap above the silicon chip (1) with experimental results subsequently confirming that a particular Monte Carlo design was accurate (2). The aim of this work was to test if a new correction-less diode could be designed using an entirely experimental methodology. Method: All measurements were performed on a Varian iX at a depth of 5 cm, SSD of 95 cm and field sizes of 5, 6, 8, 10, 20 and 30 mm. Firstly, the experimental transfer of kq,clin,kq,msr from a commonly used diode detector (IBA, stereotactic field diode (SFD)) to another diode detector (Sun Nuclear, unshielded diode, (EDGEe)) was tested. These results were compared to Monte Carlo calculated values of the EDGEe. Secondly, the air gap above the EDGEe silicon chip was optimised empirically. Nine different air gap “tops” were placed above the EDGEe (air depth = 0.3, 0.6, 0.9 mm; air width = 3.06, 4.59, 6.13 mm). The sensitivity of the EDGEe was plotted as a function of air gap thickness for the field sizes measured. Results: The transfer of kq,clin,kq,msr from the SFD to the EDGEe was correct to within the simulation and measurement uncertainties. The EDGEe detector can be made “correction-less” for field sizes of 5 and 6 mm, but was ∌2% from being “correction-less” at field sizes of 8 and 10 mm. Conclusion Different materials will perturb small fields in different ways. A detector is only “correction-less” if all these perturbations happen to cancel out. Designing a “correction-less” diode is a complicated process, thus it is reasonable to expect that Monte Carlo simulations should play an important role

    Prostate radiotherapy treatment plan quality: effects of hip prostheses [Conference Abstract]

    No full text
    Introduction This study aimed to examine the geometric and dosimetric results when radiotherapy treatment plans are designed for prostate cancer patients with hip prostheses. Methods Ten EBRT treatment plans for localised prostate cancer, in the presence of hip prostheses, were analysed and compared with a reference set of 196 treatment plans for localised prostate cancer in patients without prostheses. Crowe et al.’s TADA code [1] was used to extract treatment plan parameters and evaluate doses to target volumes and critical structures against recommended goals [2] and constraints [3, 4]. Results The need to avoid transmitting the radiation beam through the hip prostheses limited the range of gantry angles available for use in both the rotational (VMAT) and the non-rotational (3DCRT and IMRT) radiotherapy treatments. This geometric limitation (exemplified in the VMAT data shown in Fig. 1) reduced the overall quality of the treatment plans for patients with prostheses compared to the reference plans. All plans with prostheses failed the PTV dose homogeneity requirement [2], whereas only 4 % of the plans without prostheses failed this test. Several treatment plans for patients with hip prostheses also failed the QUANTEC requirements that less than 50 % of the rectum receive 50 Gy and less than 35 % of the rectum receive 60 Gy to keep the grade 3 toxicity rate below 10 % [3], or the Hansen and Roach requirement that less than 25 % of the bladder receive 75 Gy [4]. Discussion and conclusions The results of this study exemplify the difficulty of designing prostate radiotherapy treatment plans, where beams provide adequate doses to targeted tissues while avoiding nearby organs at risk, when the presence of hip prostheses limits the available treatment geometries. This work provides qualitative evidence of the compromised dose distributions that can result, in such cases

    Dosimetry of cone-defined stereotactic radiosurgery fields with a commercial synthetic diamond detector

    Get PDF
    Purpose Small field x-ray beam dosimetry is difficult due to a lack of lateral electronic equilibrium, source occlusion, high dose gradients and detector volume averaging. Currently there is no single definitive detector recommended for small field dosimetry. The objective of this work was to evaluate the performance of a new commercial synthetic diamond detector, namely the PTW 60019 microDiamond, for the dosimetry of small x-ray fields as used in stereotactic radiosurgery (SRS). Methods Small ïŹeld sizes were deïŹned by BrainLAB circular cones (4 – 30 mm diameter) on a Novalis Trilogy linear accelerator and using the 6 MV SRS x-ray beam mode for all measurements. Percentage depth doses were measured and compared to an IBA SFD and a PTW 60012 E diode. Cross profiles were measured and compared to an IBA SFD diode. Field factors, ℩_(Q_clin,Q_msr)^(f_clin,f_msr ), were calculated by Monte Carlo methods using BEAMnrc and correction factors, k_(Q_clin,Q_msr)^(f_clin,f_msr ), were derived for the PTW 60019 microDiamond detector. Results For the small fields of 4 to 30 mm diameter, there were dose differences in the PDDs of up to 1.5% when compared to an IBA SFD and PTW 60012 E diode detector. For the cross profile measurements the penumbra values varied, depending upon the orientation of the detector. The field factors, ℩_(Q_clin,Q_msr)^(f_clin,f_msr ), were calculated for these field diameters at a depth of 1.4 cm in water and they were within 2.7% of published values for a similar linear accelerator. The corrections factors, k_(Q_clin,Q_msr)^(f_clin,f_msr ), were derived for the PTW 60019 microDiamond detector. Conclusions We conclude that the new PTW 60019 microDiamond detector is generally suitable for relative dosimetry in small 6 MV SRS beams for a Novalis Trilogy linear equipped with circular cones
    • 