    PEEK Primary Crowns with Cobalt-Chromium, Zirconia and Galvanic Secondary Crowns with Different Tapers-A Comparison of Retention Forces

    Get PDF
    In prosthetic dentistry, double crown systems have proved their suitability as retainers for removable partial dentures. However, investigations in this context, regarding polyetheretherketone, are scarce. Therefore, the aim of this study was to test the retention force (RF) between polyetheretherketone (PEEK) primary and cobalt-chromium (CoCr), zirconia (ZrO2) and galvanic (GAL) secondary crowns with three different tapers. Primary PEEK-crowns were milled with the tapers 0 degrees, 1 degrees, and 2 degrees (n = 10/taper, respectively). Afterwards, 90 secondary crowns were fabricated: (i) 30 CoCr-crowns milled from Ceramill Sintron (AmannGirrbach, Koblach, Austria) (n = 10/taper), (ii) 30 ZrO2-crowns milled from Ceramill ZI (AmannGirrbach, Koblach, Austria) (n = 10/taper), and (iii) 30 GAL-crowns made using electroforming (n = 10/taper). RF was measured in a pull-off test (20 pull-offs/specimen) and data were analyzed using 2-/1-way Analysis of Variance (ANOVA) followed by the Tukey-Honestly Significant Difference (HSD) post hoc test and linear regression analyses (p 0.141). At 0 degrees taper, no differences in retention forces between GAL, CrCr, and ZrO2 crowns were found (p = 0.075). However, at 1 degrees and 2 degrees taper, lower RF for GAL-crowns were observed (p < 0.009, p < 0.001, respectively). According to this laboratory study, PEEK might be a suitable material for primary crowns, regardless of the taper and the material of secondary crown. Long-term results, however, are still necessary

    Retention Load Values of Telescopic Crowns Made of Y-TZP and CoCr with Y-TZP Secondary Crowns: Impact of Different Taper Angles

    Get PDF
    This study aimed to examine and compare the retention load values (RL) of different telescopic crown assemblies (Y-TZP and CoCr primary crowns with electroformed and Y-TZP secondary crowns each) with three different taper angles (0 degrees, 1 degrees and 2 degrees). Thirty Y-TZP primary crowns with electroformed gold copings (Z/G group) and Y-TZP secondary crowns (Z/Z group) and 30 CoCr primary crowns with electroformed gold copings (C/G group) and Y-TZP secondary crowns (C/Z group), each with taper angles of 0 degrees, 1 degrees and 2 degrees, were fabricated, respectively. With the exception of the electroformed gold copings, all specimens were Computer-Aided-Design/Computer-Aided-Manufacturing (CAD/CAM)-milled, then sintered and afterwards manually adapted. In order to stabilize the gold copings, they were fixed in a tertiary structure. The secondary crowns were constructed with a hook, which ensured self-alignment with an upper chain. Afterwards, 20 pull-off test cycles were performed in a universal testing machine under artificial saliva and after weighing the secondary crowns with a 5 kg object for 20 s. Data were analyzed by one-way and two-way Analysis of Variance (ANOVA). C/Z with 1 degrees showed higher (p = 0.009) RL than 0 degrees and 2 degrees tapers. C/G at 1 degrees also showed higher (p = 0.001) RL than at tapers of 0 degrees and 2 degrees. Z/G and C/G at 0 degrees showed lower RL than Z/Z and C/Z (p < 0.001). Primary crowns had no impact on the 0 degrees group. Z/G showed lower RL as compared to C/Z within the 1 degrees group (p = 0.007) and Z/Z in the 2 degrees group (p = 0.006). The primary crown material had no influence on RL. Electroformed copings showed lower RL. Further investigations for 1 degrees as well as for the long-term performance after thermomechanical aging are necessary

    Factors associated with self-rated health status in university students: a cross-sectional study in three European countries

    Get PDF
    Mikolajczyk RT, Brzoska P, Maier C, et al. Factors associated with self-rated health status in university students: a cross-sectional study in three European countries. BMC Public Health. 2008;8(1): 215.Background: Self-rated health status (SRHS) is a reliable and valid measure for assessing the subjective and objective health of individuals. Previous studies have either focused predominantly on the elderly or investigated only a narrow range of factors potentially associated with SRHS. In examining student populations, these past studies were limited to single countries. The objectives of this study were to assess which candidate variables were independently associated with SRHS in university students, to compare these variables by country and by gender, and to investigate which of the variables was most important as a rating frame for SRHS. Methods: The data is from the Cross-National Student Health Survey, conducted in 2005 in universities in Germany, Bulgaria, and Poland (n = 2103; mean age = 20.7 years). SRHS was assessed with a single question using a five-point scale ranging from "excellent" to "poor". The study also measured a wide range of variables including: physical and psychological health, studying, social contacts/social support, and socio-demographic status. Results: Psychosomatic complaints (considered an aspect of physical health and, adjusted for psychological health) were the most important indicators in forming a rating frame for students' SRHS. There were few differences in the effects of variables associated with SRHS by gender (well-being: a measure of psychological health) and the variables associated with SRHS by country (well-being and self-efficacy). The remaining variables showed homogenous effects for both genders and for all three countries. Conclusion: The results suggest that SRHS can be reasonably used to compare students' health across countries. SRHS is affected by different physical, psychological and psychosomatic aspects of health; however, its strongest association is with psychosomatic complaints

    Single-cell transcriptomic atlas-guided development of CAR-T cells for the treatment of acute myeloid leukemia

    Get PDF
    A single-cell screening approach identifies targets for CAR-T cells in acute myeloid leukemia. Chimeric antigen receptor T cells (CAR-T cells) have emerged as a powerful treatment option for individuals with B cell malignancies but have yet to achieve success in treating acute myeloid leukemia (AML) due to a lack of safe targets. Here we leveraged an atlas of publicly available RNA-sequencing data of over 500,000 single cells from 15 individuals with AML and tissue from 9 healthy individuals for prediction of target antigens that are expressed on malignant cells but lacking on healthy cells, including T cells. Aided by this high-resolution, single-cell expression approach, we computationally identify colony-stimulating factor 1 receptor and cluster of differentiation 86 as targets for CAR-T cell therapy in AML. Functional validation of these established CAR-T cells shows robust in vitro and in vivo efficacy in cell line- and human-derived AML models with minimal off-target toxicity toward relevant healthy human tissues. This provides a strong rationale for further clinical development

    Retention Load of Telescopic Crowns with Different Taper Angles between Cobalt-Chromium and Polyetheretherketone Made with Three Different Manufacturing Processes Examined by Pull-Off Test

    Full text link
    PURPOSE To investigate the retention loads of differently fabricated secondary telescopic polyetheretherketone (PEEK) crowns on cobalt-chromium primary crowns with different tapers. MATERIALS AND METHODS Cobalt-chromium primary crowns with 0°, 1°, and 2° tapers were constructed, milled, and sintered. Corresponding secondary crowns were fabricated by milling, pressing from pellets, and pressing from granules. For these nine test groups, the pull-off tests of each crown combination were performed 20 times, and the retention loads were measured (Zwick 1445, 50 mm/min). Data were analyzed using linear regression, covariance analysis, mixed models, Kruskal-Wallis, and Mann-Whitney U-test, together with the Benferroni-Holm correction. RESULTS The mixed models covariance analysis reinforced stable retention load values (p = 0.162) for each single test sequence. There was no interaction between the groups and the separation cycles (p = 0.179). Milled secondary crowns with 0° showed the lowest mean retention load values compared to all tested groups (p = 0.003) followed by those pressed form pellets with 1°. Regarding the different tapers, no effect of manufacturing method on the results was observed within 1° and 2° groups (p = 0.540; p = 0.052); however, among the 0° groups, the milled ones showed significantly the lowest retention load values (p = 0.002). Among the manufacturing methods, both pressed groups showed no impact of taper on the retention load values (p > 0.324 and p > 0.123, respectively), whereas among the milled secondary crowns, the 0° taper showed significantly lower retention load values than the 1° and 2° taper (p < 0.002). CONCLUSION Based on these results, telescopic crowns made of PEEK seem to show stable retention load values for each test sequence; however, data with thermo-mechanical aging are still required. In addition, further developments in CAD/CAM manufacturing of PEEK materials for telescopic crowns are warranted, especially for 0°

    Über Rassismus sprechen. Auf dem Weg zu einer rassismuskritischen Professionalität

    No full text
    Kourabas V, Mecheril P. Über Rassismus sprechen. Auf dem Weg zu einer rassismuskritischen Professionalität. In: Stock M, Hodaie N, Immerfall S, Menz M, eds. Arbeitstitel: Migrationsgesellschaft. Pädagogik – Profession – Praktik. MiGS: Migration - Gesellschaft - Schule. Wiesbaden: Springer Fachmedien Wiesbaden; 2022: 13-33.Rassismus stellt eine kontingente Unterscheidungspraxis dar, die strukturell, organisational und interaktiv wirksam ist und grundlegend auf Handlungs-, Empfindungs- und Deutungsweisen einwirkt. Über eine rassismuskritische Perspektive können diese gewöhnlichen Unterscheidungspraktiken, die auch in pädagogischen Kontexten wirksam sind, sichtbar gemacht und kritisch auf ihre Angemessenheit betrachtet werden. Der Beitrag argumentiert für ein rassismuskritisch orientiertes Verständnis von (pädagogischer) Professionalität und entwickelt Schritte einer solchen Denk- und Handlungsperspektive

    Retention force of differently fabricated telescopic PEEK crowns with different tapers

    Full text link
    To assess the retention force between primary and secondary PEEK crowns made by different fabrication methods. Primary crowns with different tapers (0°, 1°, and 2°) were fabricated and secondary crowns that were either milled from breCam BioHPP blanks, pressed from pellets (BioHPP Pellet) or granules (BioHPP Granulat) were produced. Each specimen was measured 20 times in a pulloff-test and results were analyzed using 2-/1-way ANOVA and linear regression analyses (p<0.05). Within 0° tapered crowns milled secondary crowns showed lower retention forces compared to pressed pellet crowns. Crowns with a 1° taper, however, showed no impact of the fabrication method on retention force. At a 2° taper, granular pressed crowns displayed lower values than their milled counterparts. Within the milled group, a 0° taper showed lower retention values than the higher tapers, whereas in the pressed groups, no impact of taper angle on retention force was found

    PEEK Primary Crowns with Cobalt-Chromium, Zirconia and Galvanic Secondary Crowns with Different Tapers—A Comparison of Retention Forces

    Get PDF
    In prosthetic dentistry, double crown systems have proved their suitability as retainers for removable partial dentures. However, investigations in this context, regarding polyetheretherketone, are scarce. Therefore, the aim of this study was to test the retention force (RF) between polyetheretherketone (PEEK) primary and cobalt-chromium (CoCr), zirconia (ZrO2) and galvanic (GAL) secondary crowns with three different tapers. Primary PEEK-crowns were milled with the tapers 0°, 1°, and 2° (n = 10/taper, respectively). Afterwards, 90 secondary crowns were fabricated: (i) 30 CoCr-crowns milled from Ceramill Sintron (AmannGirrbach, Koblach, Austria) (n = 10/taper), (ii) 30 ZrO2-crowns milled from Ceramill ZI (AmannGirrbach, Koblach, Austria) (n = 10/taper), and (iii) 30 GAL-crowns made using electroforming (n = 10/taper). RF was measured in a pull-off test (20 pull-offs/specimen) and data were analyzed using 2-/1-way Analysis of Variance (ANOVA) followed by the Tukey-Honestly Significant Difference (HSD) post hoc test and linear regression analyses (p &lt; 0.05). The measured mean RF values ranged between 9.6 and 38.2 N. With regard to the 0°, 1°, and 2° tapered crowns, no statistically significant differences between CoCr and ZrO2 were observed (p &gt; 0.141). At 0° taper, no differences in retention forces between GAL, CrCr, and ZrO2 crowns were found (p = 0.075). However, at 1° and 2° taper, lower RF for GAL-crowns were observed (p &lt; 0.009, p &lt; 0.001, respectively). According to this laboratory study, PEEK might be a suitable material for primary crowns, regardless of the taper and the material of secondary crown. Long-term results, however, are still necessary

    Retention Load Values of Telescopic Crowns Made of Y-TZP and CoCr with Y-TZP Secondary Crowns: Impact of Different Taper Angles

    Get PDF
    This study aimed to examine and compare the retention load values (RL) of different telescopic crown assemblies (Y-TZP and CoCr primary crowns with electroformed and Y-TZP secondary crowns each) with three different taper angles (0°, 1° and 2°). Thirty Y-TZP primary crowns with electroformed gold copings (Z/G group) and Y-TZP secondary crowns (Z/Z group) and 30 CoCr primary crowns with electroformed gold copings (C/G group) and Y-TZP secondary crowns (C/Z group), each with taper angles of 0°, 1° and 2°, were fabricated, respectively. With the exception of the electroformed gold copings, all specimens were Computer-Aided-Design/Computer-Aided-Manufacturing (CAD/CAM)-milled, then sintered and afterwards manually adapted. In order to stabilize the gold copings, they were fixed in a tertiary structure. The secondary crowns were constructed with a hook, which ensured self-alignment with an upper chain. Afterwards, 20 pull-off test cycles were performed in a universal testing machine under artificial saliva and after weighing the secondary crowns with a 5 kg object for 20 s. Data were analyzed by one-way and two-way Analysis of Variance (ANOVA). C/Z with 1° showed higher (p = 0.009) RL than 0° and 2° tapers. C/G at 1° also showed higher (p = 0.001) RL than at tapers of 0° and 2°. Z/G and C/G at 0° showed lower RL than Z/Z and C/Z (p &lt; 0.001). Primary crowns had no impact on the 0° group. Z/G showed lower RL as compared to C/Z within the 1° group (p = 0.007) and Z/Z in the 2° group (p = 0.006). The primary crown material had no influence on RL. Electroformed copings showed lower RL. Further investigations for 1° as well as for the long-term performance after thermomechanical aging are necessary