22 research outputs found
Task-based quantification of image quality using a model observer in abdominal CT: a multicentre study.
We investigated the variability in diagnostic information inherent in computed tomography (CT) images acquired at 68 different CT units, with the selected acquisition protocols aiming to answer the same clinical question.
An anthropomorphic abdominal phantom with two optional rings was scanned on 68 CT systems from 62 centres using the local clinical acquisition parameters of the portal venous phase for the detection of focal liver lesions. Low-contrast detectability (LCD) was assessed objectively with channelised Hotelling observer (CHO) using the receiver operating characteristic (ROC) paradigm. For each lesion size, the area under the ROC curve (AUC) was calculated and considered as a figure of merit. The volume computed tomography dose index (CTDI <sub>vol</sub> ) was used to indicate radiation dose exposure.
The median CTDI <sub>vol</sub> used was 5.8 mGy, 10.5 mGy and 16.3 mGy for the small, medium and large phantoms, respectively. The median AUC obtained from clinical CT protocols was 0.96, 0.90 and 0.83 for the small, medium and large phantoms, respectively.
Our study used a model observer to highlight the difference in image quality levels when dealing with the same clinical question. This difference was important and increased with growing phantom size, which generated large variations in patient exposure. In the end, a standardisation initiative may be launched to ensure comparable diagnostic information for well-defined clinical questions. The image quality requirements, related to the clinical question to be answered, should be the starting point of patient dose optimisation.
• Model observers enable to assess image quality objectively based on clinical tasks. • Objective image quality assessment should always include several patient sizes. • Clinical diagnostic image quality should be the starting point for patient dose optimisation. • Dose optimisation by applying DRLs only is insufficient for ensuring clinical requirements
EXPOSURE OF THE SWISS POPULATION BY RADIODIAGNOSTICS: 2013 REVIEW.
In 2013, a nationwide investigation was conducted in Switzerland to establish the population's exposure from medical X rays. A hybrid approach was used combining the Raddose database accessible on-line by the participating practices and the Swiss medical tariffication system for hospitals. This study revealed that the average annual number of examinations is 1.2 per inhabitant, and the associated annual effective dose is 1.4 mSv. It also showed that computed tomography is the most irradiating modality and that it delivers 70 % of the total dose. The annual effective dose per inhabitant registered a 17 % increase in 5 y and is comparable with what was recently reported in neighbouring countries
Contemporary snapshot of tumor regression grade (TRG) distribution in locally advanced rectal cancer: a cross sectional multicentric experience.
Pre-operative chemoradiotherapy (CRT) followed by surgical resection is still the standard treatment for locally advanced low rectal cancer. Nowadays new strategies are emerging to treat patients with a complete response to pre-operative treatment, rendering the optimal management still controversial and under debate. The primary aim of this study was to obtain a snapshot of tumor regression grade (TRG) distribution after standard CRT. Second, we aimed to identify a correlation between clinical tumor stage (cT) and TRG, and to define the accuracy of magnetic resonance imaging (MRI) in the restaging setting. Between January 2017 and June 2019, a cross sectional multicentric study was performed in 22 referral centers of colon-rectal surgery including all patients with cT3-4Nx/cTxN1-2 rectal cancer who underwent pre-operative CRT. Shapiro-Wilk test was used for continuous data. Categorical variables were compared with Chi-squared test or Fisher's exact test, where appropriate. Accuracy of restaging MRI in the identification of pathologic complete response (pCR) was determined evaluating the correspondence with the histopathological examination of surgical specimens.In the present study, 689 patients were enrolled. Complete tumor regression rate was 16.9%. The "watch and wait" strategy was applied in 4.3% of TRG4 patients. A clinical correlation between more advanced tumors and moderate to absent tumor regression was found (p = 0.03). Post-neoadjuvant MRI had low sensibility (55%) and high specificity (83%) with accuracy of 82.8% in identifying TRG4 and pCR.Our data provided a contemporary description of the effects of pre-operative CRT on a large pool of locally advanced low rectal cancer patients treated in different colon-rectal surgical centers
BENCHMARKING OF CT FOR PATIENT EXPOSURE OPTIMISATION.
Patient dose optimisation in computed tomography (CT) should be done using clinically relevant tasks when dealing with image quality assessments. In the present work, low-contrast detectability for an average patient morphology was assessed on 56 CT units, using a model observer applied on images acquired with two specific protocols of an anthropomorphic phantom containing spheres. Images were assessed using the channelised Hotelling observer (CHO) with dense difference of Gaussian channels. The results were computed by performing receiver operating characteristics analysis (ROC) and using the area under the ROC curve (AUC) as a figure of merit. The results showed a small disparity at a volume computed tomography dose index (CTDIvol) of 15 mGy depending on the CT units for the chosen image quality criterion. For 8-mm targets, AUCs were 0.999 ± 0.018 at 20 Hounsfield units (HU) and 0.927 ± 0.054 at 10 HU. For 5-mm targets, AUCs were 0.947 ± 0.059 and 0.702 ± 0.068 at 20 and 10 HU, respectively. The robustness of the CHO opens the way for CT protocol benchmarking and optimisation processes
InterCardioRisk: a novel online tool for estimating doses of ionising radiation to occupationally-exposed medical staff and their associated health risks.
Those working in interventional cardiology and related medical procedures are potentially subject to considerable exposure to x-rays. Two types of tissue of particular concern that may receive considerable doses during such procedures are the lens of the eye and the brain. Ocular radiation exposure results in lens changes that, with time, may progress to partial or total lens opacification (cataracts). In the early stages, such opacities do not result in visual disability; the severity of such changes tends to increase progressively with dose and time until vision is impaired and cataract surgery is required. Scattered radiation doses to the eye lens of an interventional cardiologist in typical working conditions can exceed 34 μGy min(-1) in high-dose fluoroscopy modes and 3 μGy per image during image acquisition (instantaneous rate values) when radiation protection tools are not used. A causal relation between exposure to ionising radiation and increased risk of brain and central nervous system tumours has been shown in a number of studies. Although absorbed doses to the brain in interventional cardiology procedures are lower than those to the eye lens by a factor between 3.40 and 8.08 according to our simulations, doses to both tissues are among the highest occupational radiation doses documented for medical staff whose work involves exposures to x-rays. We present InterCardioRisk, a tool featuring an easy-to-use web interface that provides a general estimation of both cumulated absorbed doses experienced by medical staff exposed in the interventional cardiology setting and their estimated associated health risks. The tool is available at http://intercardiorisk.creal.cat