46 research outputs found

    Cone-beam Computed Tomography-guided Stereotactic Liver Punctures: A Phantom Study

    Get PDF
    Purpose: Images from computed tomography (CT), combined with navigation systems, improve the outcomes of local thermal therapies that are dependent on accurate probe placement. Although the usage of CT is desired, its availability for time-consuming radiological interventions is limited. Alternatively, three-dimensional images from C-arm cone-beam CT (CBCT) can be used. The goal of this study was to evaluate the accuracy of navigated CBCT-guided needle punctures, controlled with CT scans. Methods: Five series of five navigated punctures were performed on a nonrigid phantom using a liver specific navigation system and CBCT volumetric dataset for planning and navigation. To mimic targets, five titanium screws were fixed to the phantom. Target positioning accuracy (TPECBCT) was computed from control CT scans and divided into lateral and longitudinal components. Additionally, CBCT-CT guidance accuracy was deducted by performing CBCT-to-CT image coregistration and measuring TPECBCT-CT from fused datasets. Image coregistration was evaluated using fiducial registration error (FRECBCT-CT) and target registration error (TRECBCT-CT). Results: Positioning accuracies in lateral directions pertaining to CBCT (TPECBCT=2.1±1.0mm) were found to be better to those achieved from previous study using CT (TPECT=2.3±1.3mm). Image coregistration error was 0.3±0.1mm, resulting in an average TRE of 2.1±0.7mm (N=5 targets) and average Euclidean TPECBCT-CT of 3.1±1.3mm. Conclusions: Stereotactic needle punctures might be planned and performed on volumetric CBCT images and controlled with multidetector CT with positioning accuracy higher or similar to those performed using CT scanner

    Cone-beam computed tomography-guided stereotactic liver punctures: a phantom study

    Get PDF
    PURPOSE Images from computed tomography (CT), combined with navigation systems, improve the outcomes of local thermal therapies that are dependent on accurate probe placement. Although the usage of CT is desired, its availability for time-consuming radiological interventions is limited. Alternatively, three-dimensional images from C-arm cone-beam CT (CBCT) can be used. The goal of this study was to evaluate the accuracy of navigated CBCT-guided needle punctures, controlled with CT scans. METHODS Five series of five navigated punctures were performed on a nonrigid phantom using a liver specific navigation system and CBCT volumetric dataset for planning and navigation. To mimic targets, five titanium screws were fixed to the phantom. Target positioning accuracy (TPECBCT) was computed from control CT scans and divided into lateral and longitudinal components. Additionally, CBCT-CT guidance accuracy was deducted by performing CBCT-to-CT image coregistration and measuring TPECBCT-CT from fused datasets. Image coregistration was evaluated using fiducial registration error (FRECBCT-CT) and target registration error (TRECBCT-CT). RESULTS Positioning accuracies in lateral directions pertaining to CBCT (TPECBCT = 2.1 ± 1.0 mm) were found to be better to those achieved from previous study using CT (TPECT = 2.3 ± 1.3 mm). Image coregistration error was 0.3 ± 0.1 mm, resulting in an average TRE of 2.1 ± 0.7 mm (N = 5 targets) and average Euclidean TPECBCT-CT of 3.1 ± 1.3 mm. CONCLUSIONS Stereotactic needle punctures might be planned and performed on volumetric CBCT images and controlled with multidetector CT with positioning accuracy higher or similar to those performed using CT scanners

    Thermal ablation of colorectal liver metastases: a position paper by an international panel of ablation experts, the interventional oncology sans frontières meeting 2013

    Get PDF
    Objectives: Previous attempts at meta-analysis and systematic review have not provided clear recommendations for the clinical application of thermal ablation in metastatic colorectal cancer. Many authors believe that the probability of gathering randomised controlled trial (RCT) data is low. Our aim is to provide a consensus document making recommendations on the appropriate application of thermal ablation in patients with colorectal liver metastases. Methods: This consensus paper was discussed by an expert panel at The Interventional Oncology Sans Frontières 2013. A literature review was presented. Tumour characteristics, ablation technique and different clinical applications were considered and the level of consensus was documented. Results: Specific recommendations are made with regard to metastasis size, number, and location and ablation technique. Mean 31 % 5-year survival post-ablation in selected patients has resulted in acceptance of this therapy for those with technically inoperable but limited liver disease and those with limited liver reserve or co-morbidities that render them inoperable. Conclusions: In the absence of RCT data, it is our aim that this consensus document will facilitate judicious selection of the patients most likely to benefit from thermal ablation and provide a unified interventional oncological perspective for the use of this technology. Key Points: • Best results require due consideration of tumour size, number, volume and location. • Ablation technology, imaging guidance and intra-procedural imaging assessment must be optimised. • Accepted applications include inoperable disease due to tumour distribution or inadequate liver reserve. • Other current indications include concurrent co-morbidity, patient choice and the test-of-time approach. • Future applications may include resectable disease, e.g. for small solitary tumours

    A novel but frequent variant in LPA KIV-2 is associated with a pronounced Lp(a) and cardiovascular risk reduction

    Get PDF
    Aims Lp(a) concentrations represent a major cardiovascular risk factor and are almost entirely controlled by one single locus (LPA). However, many genetic factors in LPA governing the enormous variance of Lp(a) levels are still unknown. Since up to 70% of the LPA coding sequence are located in a difficult to access hypervariable copy number variation named KIV-2, we hypothesized that it may contain novel functional variants with pronounced effects on Lp(a) concentrations. We performed a large scale mutation analysis in the KIV-2 using an extreme phenotype approach Methods and results We compiled an discovery set of 123 samples showing discordance between LPA isoform phenotype and Lp(a) concentrations and controls. Using ultra-deep sequencing, we identified a splice site variant (G4925A) in preferential association with the smaller LPA isoforms. Follow-up in a European general population (n = 2892) revealed an exceptionally high carrier frequency of 22.1% in the general population. The variant explains 20.6% of the Lp(a) variance in carriers of low molecular weight (LMW) apo(a) isoforms (P = 5.75e-38) and reduces Lp(a) concentrations by 31.3 mg/dL. Accordingly the odds ratio for cardiovascular disease was reduced from 1.39 [95% confidence interval (CI): 1.17-1.66, P = 1.89e-04] for wildtype LMW individuals to 1.19 [95% CI: 0.92;1.56, P = 0.19] in LMW individuals who were additionally positive for G4925A. Functional studies point towards a reduction of splicing efficiency by this novel variant. Conclusion A highly frequent but until now undetected variant in the LPA KIV-2 region is strongly associated with reduced Lp(a) concentrations and reduced cardiovascular risk in LMW individuals

    A gut bacterial signature in blood and liver tissue characterizes cirrhosis and hepatocellular carcinoma

    Get PDF
    BackgroundHCC is the leading cause of cancer in chronic liver disease. A growing body of experimental mouse models supports the notion that gut-resident and liver-resident microbes control hepatic immune responses and, thereby, crucially contribute to liver tumorigenesis. However, a comprehensive characterization of the intestinal microbiome in fueling the transition from chronic liver disease to HCC in humans is currently missing.MethodsHere, we profiled the fecal, blood, and liver tissue microbiome of patients with HCC by 16S rRNA sequencing and compared profiles to nonmalignant cirrhotic and noncirrhotic NAFLD patients.ResultsWe report a distinct bacterial profile, defined from 16S rRNA gene sequences, with reduced α-and β-diversity in the feces of patients with HCC and cirrhosis compared to NAFLD. Patients with HCC and cirrhosis exhibited an increased proportion of fecal bacterial gene signatures in the blood and liver compared to NAFLD. Differential analysis of the relative abundance of bacterial genera identified an increased abundance of Ruminococcaceae and Bacteroidaceae in blood and liver tissue from both HCC and cirrhosis patients compared to NAFLD. Fecal samples from cirrhosis and HCC patients both showed a reduced abundance for several taxa, including short-chain fatty acid-producing genera, such as Blautia and Agathobacter. Using paired 16S rRNA and transcriptome sequencing, we identified a direct association between gut bacterial genus abundance and host transcriptome response within the liver tissue.ConclusionsOur study indicates perturbations of the intestinal and liver-resident microbiome as a critical determinant of patients with cirrhosis and HCC

    Consensus guidelines for the definition of time-to-event end points in image-guided tumor ablation: results of the SIO and DATECAN initiative

    Get PDF
    International audienceThere is currently no consensus regarding preferred clinical outcome measures following image-guided tumor ablation or clear definitions of oncologic end points. This consensus document proposes standardized definitions for a broad range of oncologic outcome measures with recommendations on how to uniformly document, analyze, and report outcomes. The initiative was coordinated by the Society of Interventional Oncology in collaboration with the Definition for the Assessment of Time-to-Event End Points in Cancer Trials, or DATECAN, group. According to predefined criteria, based on experience with clinical trials, an international panel of 62 experts convened. Recommendations were developed using the validated three-step modified Delphi consensus method. Consensus was reached on when to assess outcomes per patient, per session, or per tumor; on starting and ending time and survival time definitions; and on time-to-event end points. Although no consensus was reached on the preferred classification system to report complications, quality of life, and health economics issues, the panel did agree on using the most recent version of a validated patient-reported outcome questionnaire. This article provides a framework of key opinion leader recommendations with the intent to facilitate a clear interpretation of results and standardize worldwide communication. Widespread adoption will improve reproducibility, allow for accurate comparisons, and avoid misinterpretations in the field of interventional oncology research. Published under a CC BY 4.0 license. Online supplemental material is available for this article. See also the editorial by Liddell in this issue

    Combined Method for Segmentation of Multiple Needles in 3D Liver CT

    No full text
    Radiofrequency ablation is a minimally invasive technique for treating liver or kidney tumors, where precisely inserted ablation probes destroy the tumor tissue with induced heat around their tips. Automatic segmentation of needles from the abdominal computed tomography (CT) scans enables large-scale procedure evaluation and creates a baseline ground truth for automated planning approaches. Our method for needle segmentation enables the detection of multiple needles from unlabeled CT scans by sequential combination of two approaches, a classical method and a neural network. The former method detects intensity-prominent needles in 3D. The latter focuses on the detection of needles with lower intensity, that remained undetected by the classical method. A well-known encoder-decoder approach on axial slices of the CT scan gives a segmentation mask for the needles prominent in slices. To train the network, we create a synthetic dataset, adding needles to clean abdominal CT slices. Our network trained on a small synthetic dataset is able to generalize to clinical CT scans. The performance of the combined method is evaluated on ten anonymous patients from the Medical University of Innsbruck each with up to 18 needles inserted in the liver without any template

    Tumour ablation: technical aspects

    No full text
    Image-guided percutaneous radiofrequency ablation (RFA) is a minimally invasive, relatively low-risk procedure for tumour treatment. Local recurrence and survival rates depend on the rate of complete ablation of the entire tumour including a sufficient margin of surrounding healthy tissue. Currently a variety of different RFA devices are available. The interventionalist must be able to predict the configuration and extent of the resulting ablation necrosis. Accurate planning and execution of RFA according to the size and geometry of the tumour is essential. In order to minimize complications, individualized treatment strategies may be necessary for tumours close to vital structures. This review examines the state-of-the art of different device technologies, approaches, and treatment strategies for percutaneous RFA of liver tumours

    Accurate physics-based registration for the outcome validation of minimal invasive interventions and open liver surgeries

    No full text
    The purpose of this paper is to present an outcome validation tool for tumor radiofrequency (RF) ablation and resection. Methods: Intervention assessment tools require an accurate registration of both pre- and postoperative computed tomographies able to handle big deformations. Therefore, a physics-based method is proposed with that purpose. To increase the accuracy both automatically detected internal and surface physical landmarks are incorporated in the registration process. Results: The algorithm has been evaluated in 25 clinical datasets containing RF ablations, resections, and patients with recurrent tumors. The achieved accuracy is 1.2 mm measured as mean internal distance between vessel landmarks and a positive predictive value of 0.95. The quantitative and qualitative results of the outcome validation tool show that in 50% of the cases tumors were only partially covered by the treatment. Conclusion: The use of internal and surface landmarks combined with a physics-based registration method increases the accuracy of the results compared to the accuracy of state of the art methods. An accurate outcome validation tool is important in order to certify that the tumor and its safety margin were fully covered by the treatment. Significance: An accurate outcome validation tool can result in a decrease of the tumor recurrence rate
    corecore