    Postoperative lead migration in deep brain stimulation surgery: Incidence, risk factors, and clinical impact.

    No full text
    Deep brain stimulation (DBS) is an effective treatment for multiple movement disorders and shows substantial promise for the treatment of some neuropsychiatric and other disorders of brain neurocircuitry. Optimal neuroanatomical lead position is a critical determinant of clinical outcomes in DBS surgery. Lead migration, defined as an unintended post-operative displacement of the DBS lead, has been previously reported. Despite several reports, however, there have been no systematic investigations of this issue. This study aimed to: 1) quantify the incidence of lead migration in a large series of DBS patients, 2) identify potential risk factors contributing to DBS lead migration, and 3) investigate the practical importance of this complication by correlating its occurrence with clinical outcomes.A database of all DBS procedures performed at UF was queried for patients who had undergone multiple post-operative DBS lead localization imaging studies separated by at least two months. Bilateral DBS implantation has commonly been performed as a staged procedure at UF, with an interval of six or more months between sides. To localize the position of each DBS lead, a head CT is acquired ~4 weeks after lead implantation and fused to the pre-operative targeting MRI. The fused targeting images (MR + stereotactic CT) acquired in preparation for the delayed second side lead implantation provide an opportunity to repeat the localization of the first implanted lead. This paradigm offers an ideal patient population for the study of delayed DBS lead migration because it provides a large cohort of patients with localization of the same implanted DBS lead at two time points. The position of the tip of each implanted DBS lead was measured on both the initial post-operative lead localization CT and the delayed CT. Lead tip displacement, intracranial lead length, and ventricular indices were collected and analyzed. Clinical outcomes were characterized with validated rating scales for all cases, and a comparison was made between outcomes of cases with lead migration versus those where migration of the lead did not occur.Data from 138 leads in 132 patients with initial and delayed lead localization CT scans were analyzed. The mean distance between initial and delayed DBS lead tip position was 2.2 mm and the mean change in intracranial lead length was 0.45 mm. Significant delayed migration (>3 mm) was observed in 17 leads in 16 patients (12.3% of leads, 12.1% of patients). Factors associated with lead migration were: technical error, repetitive dystonic head movement, and twiddler's syndrome. Outcomes were worse in dystonia patients with lead migration (p = 0.035). In the PD group, worse clinical outcomes trended in cases with lead migration.Over 10% of DBS leads in this large single center cohort were displaced by greater than 3 mm on delayed measurement, adversely affecting outcomes. Multiple risk factors emerged, including technical error during implantation of the DBS pulse generator and failure of lead fixation at the burr hole site. We hypothesize that a change in surgical technique and a more effective lead fixation device might mitigate this problem

    Active recharge biphasic stimulation for the intraoperative monopolar review in deep brain stimulation

    Get PDF
    IntroductionCharge balancing is used in deep brain stimulation (DBS) to avoid net charge accumulation at the tissue-electrode interface that can result in neural damage. Charge balancing paradigms include passive recharge and active recharge. In passive recharge, each cathodic pulse is accompanied by a waiting period before the next stimulation, whereas active recharge uses energy to deliver symmetric anodic and cathodic stimulation pulses sequentially, producing a net zero charge. We sought to determine differences in stimulation induced side effect thresholds between active vs. passive recharge during the intraoperative monopolar review.MethodsSixty-five consecutive patients undergoing DBS from 2021 to 2022 were retrospectively reviewed. Intraoperative monopolar review was performed with both active recharge and passive recharge for all included patients to determine side effect stimulation thresholds. Sixteen patients with 64 total DBS contacts met inclusion criteria for further analysis. Intraoperative monopolar review results were compared with the monopolar review from the first DBS programming visit.ResultsThe mean intraoperative active recharge stimulation threshold was 4.1 mA, while the mean intraoperative passive recharge stimulation threshold was 3.9 mA, though this difference was not statistically significant on t-test (p = 0.442). Mean stimulation threshold at clinic follow-up was 3.2 mA. In Pearson correlation, intraoperative passive recharge thresholds had stronger correlation with follow-up stimulation thresholds (Pearson r = 0.5281, p < 0.001) than intraoperative active recharge (Pearson r = 0.340, p = 0.018), however the difference between these correlations was not statistically significant on Fisher Z correlation test (p = 0.294). The mean difference between intraoperative passive recharge stimulation threshold and follow-up stimulation threshold was 0.8 mA, while the mean difference between intraoperative active recharge threshold and follow-up threshold was 1.2 mA. This difference was not statistically significant on a t-test (p = 0.134).ConclusionsBoth intraoperative active recharge and passive recharge stimulation were well-correlated with the monopolar review at the first programming visit. No statistically significant differences were observed suggesting that either passive or active recharge may be utilized intraoperatively

    Representative images of dislocated leads.

    No full text
    <p><b>Cases described in the text. Case 1: A unique ventral lead migration due to repetitive dystonic head movements. Cases 3 and 5: Dorsal lead migrations secondary to “twiddler syndrome”</b>. Arrows and numbers indicate the direction and the distance of the movement, respectively.</p
    corecore