    Cryogenic Q-factor measurement of optical substrates for optimization of gravitational wave detectors

    Get PDF
    Future generations of gravitational wave interferometers are likely to be operated at cryogenic temperatures because one of the sensitivity limiting factors of the present generation is the thermal noise of end mirrors and beam splitters that occurs in the optical substrates as well as in the dielectric coatings. A possible method for minimizing thermal noise is cooling to cryogenic temperatures, maximizing the mechanical quality factor Q, and maximizing the eigenfrequencies of the substrate. We present experimental details of a new cryogenic apparatus that is suitable for the measurement of the temperature-dependent Q-factor of reflective, transmissive as well as nano-structured grating optics down to 5 K. In particular, the SQUID-based and the optical interferometric approaches to the measurement of the amplitude of vibrating test bodies are compared and the method of ring-down recording is described

    High-sensitivity tool for studying phonon related mechanical losses in low loss materials

    Full text link
    Fundamental mechanical loss mechanisms exist even in very pure materials, for instance, due to the interactions of excited acoustic waves with thermal phonons. A reduction of these losses in a certain frequency range is desired in high precision instruments like gravitational wave detectors. Systematic analyses of the mechanical losses in those low loss materials are essential for this aim, performed in a highly sensitive experimental set-up. Our novel method of mechanical spectroscopy, cryogenic resonant acoustic spectroscopy of bulk materials (CRA spectroscopy), is well suited to systematically determine losses at the resonant frequencies of the samples of less than 10^(-9) in the wide temperature range from 5 to 300 K. A high precision set-up in a specially built cryostat allows contactless excitation and readout of the oscillations of the sample. The experimental set-up and measuring procedure are described. Limitations to our experiment due to external loss mechanisms are analysed. The influence of the suspension system as well as the sample preparation is explained.Comment: 4 pages, 3 figures, proceedings of PHONONS07, submitted to Journal of Physics: Conference Serie

    Mechanical losses in low loss materials studied by Cryogenic Resonant Acoustic spectroscopy of bulk materials (CRA spectroscopy)

    Full text link
    Mechanical losses of crystalline silicon and calcium fluoride have been analyzed in the temperature range from 5 to 300 K by our novel mechanical spectroscopy method, cryogenic resonant acoustic spectroscopy of bulk materials (CRA spectrocopy). The focus lies on the interpretation of the measured data according to phonon-phonon interactions and defect induced losses in consideration of the excited mode shape.Comment: 4 pages, 4 figures, proceedings of the PHONONS 2007, submitted to Journal of Physics: Conference Serie

    Mechanical Q-factor measurements on a test mass with a structured surface

    Get PDF
    We present mechanical Q-factors (quality factors) of a crystalline quartz test mass with a nano-structured surface, measured in the temperature regime from 5 to 300 K. The nano-structure was a grating with a period of 2 mu m and a depth of about 0.1 mu m. Comparative measurements were performed on the plain substrate and on the structured test mass with different numbers of SiO2/Ta2O5 coating layers. The measurements at different stages of the test mass fabrication process show that the surface distortion induced by the nanostructure does not severely lower the mechanical Q-factor of the substrate. Damping due to a multi-layer coating stack was found to be orders of magnitude higher. The results provide vital information concerning the potential usage of low-thermal noise nano-structured test masses in future generations of high-precision laser interferometers and in current attempts to measure quantum effects of macroscopic mirror oscillators.DFG/SFB/Transregio

    On the mechanical quality factors of cryogenic test masses from fused silica and crystalline quartz

    Get PDF
    Current interferometric gravitational wave detectors (IGWDs) are operated at room temperature with test masses made from fused silica. Fused silica shows very low absorption at the laser wavelength of 1064 nm. It is also well suited to realize low thermal noise floors in the detector signal band since it offers low mechanical loss, i. e. high quality factors (Q factors) at room temperature. However, for a further reduction of thermal noise, cooling the test masses to cryogenic temperatures may prove an interesting technique. Here we compare the results of Q factor measurements at cryogenic temperatures of acoustic eigenmodes of test masses from fused silica and its crystalline counterpart. Our results show that the mechanical loss of fused silica increases with lower temperature and reaches a maximum at 30 K for frequencies of slightly above 10 kHz. The losses of crystalline quartz generally show lower values and even fall below the room temperature values of fused silica below 10 K. Our results show that in comparison to fused silica, crystalline quartz has a considerably narrower and lower dissipation peak on cooling and thus has more promise as a test mass material for IGDWs operated at cryogenic temperatures. The origin of the different Q factor versus temperature behavior of the two materials is discussed

    Non-perturbative measurement of low-intensity charged particle beams

    Get PDF
    Non-perturbative measurements of low-intensity charged particle beams are particularly challenging to beam diagnostics due to the low amplitude of the induced electromagnetic fields. In the low-energy antiproton decelerator (AD) and the future extra low energy antiproton rings at CERN, an absolute measurement of the beam intensity is essential to monitor the operation efficiency. Superconducting quantum interference device (SQUID) based cryogenic current comparators (CCC) have been used for measuring slow charged beams in the nA range, showing a very good current resolution. But these were unable to measure fast bunched beams, due to the slew-rate limitation of SQUID devices and presented a strong susceptibility to external perturbations. Here, we present a CCC system developed for the AD machine, which was optimised in terms of its current resolution, system stability, ability to cope with short bunched beams, and immunity to mechanical vibrations. This paper presents the monitor design and the first results from measurements with a low energy antiproton beam obtained in the AD in 2015. These are the first CCC beam current measurements ever performed in a synchrotron machine with both coasting and short bunched beams. It is shown that the system is able to stably measure the AD beam throughout the entire cycle, with a current resolution of 30nA30\,\mathrm{nA}

    Measurements of a low temperature mechanical dissipation peak in a single layer of Ta2O5 doped with TiO2

    Get PDF
    Thermal noise arising from mechanical dissipation in oxide coatings is a major limitation to many precision measurement systems, including optical frequency standards, high resolution optical spectroscopy and interferometric gravity wave detectors. Presented here are measurements of dissipation as a function of temperature between 7 K and 290 K in ion-beam sputtered Ta2O5 doped with TiO2, showing a loss peak at 20 K. Analysis of the peak provides the first evidence of the source of dissipation in doped Ta2O5 coatings, leading to possibilities for the reduction of thermal noise effects

    Untersuchungen zu hochempfindlichen SQUID-Positionsdetektoren fuer die Mission STEP und Durchfuehrung experimenteller Tests am Fallturm Bremen Abschlussbericht

    No full text
    SIGLEAvailable from TIB Hannover: F02B210 / FIZ - Fachinformationszzentrum Karlsruhe / TIB - Technische InformationsbibliothekBundesministerium fuer Bildung und Forschung, Berlin (Germany)DEGerman

    Highly Sensitive Measurements of the Dark Current of Superconducting Cavities for TESLA Using a SQUID Based Cryogenic Current Comparator

    No full text
    This contribution presents a Cryogenic Current Comparator (CCC) as an excellent tool for detecting dark currents generated, e.g. by superconducting cavities for the upcoming TESLA project (X-FEL) at DESY. To achieve the maximum possible energy the gradient of the superconducting RF cavities should be pushed close to the physical limit of 50 MV/m. The undesired field emission of electrons (so-called dark current) of the superconducting RF cavities at strong fields may limit the maximum gradient. The absolute measurement of the dark current in correlation with the gradient will give a proper value to compare and classify the cavities. The main component of the CCC is a highly sensitive LTS-DC SQUID system which is able to measure extremely low magnetic fields, e.g. caused by the dark current. For this reason the input coil of the SQUID is connected across a special designed toroidal niobium pick-up coil for the passing electron beam. A noise limited current resolution of nearly 2 pA/√(Hz) with a measurement bandwidth of up to 70 kHz was achieved in the laboratory. Design issues of the CCC and the application in the CHECHIA cavity test stand at DESY as well as experimental results will be discussed