1 research outputs found

    Measurement of cortisol in saliva: a comparison of measurement error within and between international academic-research laboratories

    Get PDF
    Objective: Hundreds of scientific publications are produced annually that involve the measurement of cortisol in saliva. Intra- and inter-laboratory variation in salivary cortisol results has the potential to contribute to cross- study inconsistencies in findings, and the perception that salivary cortisol results are unreliable. This study rigor- ously estimates sources of measurement variability in the assay of salivary cortisol within and between established international academic-based laboratories that specialize in saliva analyses. One hundred young adults (Mean age: 23.10 years; 62 females) donated 2 mL of whole saliva by passive drool. Each sample was split into multiple- 100 µL aliquots and immediately frozen. One aliquot of each of the 100 participants’ saliva was transported to academic laboratories (N = 9) in the United States, Canada, UK, and Germany and assayed for cortisol by the same commercially available immunoassay. Results: 1.76% of the variance in salivary cortisol levels was attributable to differences between duplicate assays of the same sample within laboratories, 7.93% of the variance was associated with differences between laboratories, and 90.31% to differences between samples. In established-qualified laboratories, measurement error of salivary cortisol is minimal, and inter-laboratory differences in measurement are unlikely to have a major influence on the determined values