Optical calibration hardware for the Sudbury Neutrino Observatory
The optical properties of the Sudbury Neutrino Observatory (SNO) heavy water
Cherenkov neutrino detector are measured in situ using a light diffusing sphere
("laserball"). This diffuser is connected to a pulsed nitrogen/dye laser via
specially developed underwater optical fibre umbilical cables. The umbilical
cables are designed to have a small bending radius, and can be easily adapted
for a variety of calibration sources in SNO. The laserball is remotely
manipulated to many positions in the D2O and H2O volumes, where data at six
different wavelengths are acquired. These data are analysed to determine the
absorption and scattering of light in the heavy water and light water, and the
angular dependence of the response of the detector's photomultiplier tubes.
This paper gives details of the physical properties, construction, and optical
characteristics of the laserball and its associated hardware.Comment: 17 pages, 8 figures, submitted to Nucl. Inst. Meth.
Radon backgrounds in the DEAP-1 liquid-argon-based Dark Matter detector
The DEAP-1 \SI{7}{kg} single phase liquid argon scintillation detector was
operated underground at SNOLAB in order to test the techniques and measure the
backgrounds inherent to single phase detection, in support of the
\mbox{DEAP-3600} Dark Matter detector. Backgrounds in DEAP are controlled
through material selection, construction techniques, pulse shape discrimination
and event reconstruction. This report details the analysis of background events
observed in three iterations of the DEAP-1 detector, and the measures taken to
reduce them.
The Rn decay rate in the liquid argon was measured to be between 16
and \SI{26}{\micro\becquerel\per\kilogram}. We found that the background
spectrum near the region of interest for Dark Matter detection in the DEAP-1
detector can be described considering events from three sources: radon
daughters decaying on the surface of the active volume, the expected rate of
electromagnetic events misidentified as nuclear recoils due to inefficiencies
in the pulse shape discrimination, and leakage of events from outside the
fiducial volume due to imperfect position reconstruction. These backgrounds
statistically account for all observed events, and they will be strongly
reduced in the DEAP-3600 detector due to its higher light yield and simpler
geometry
The calibration of the Sudbury Neutrino Observatory using uniformly distributed radioactive sources
The production and analysis of distributed sources of 24Na and 222Rn in the
Sudbury Neutrino Observatory (SNO) are described. These unique sources provided
accurate calibrations of the response to neutrons, produced through
photodisintegration of the deuterons in the heavy water target, and to low
energy betas and gammas. The application of these sources in determining the
neutron detection efficiency and response of the 3He proportional counter
array, and the characteristics of background Cherenkov light from trace amounts
of natural radioactivity is described.Comment: 24 pages, 13 figure
Improving Photoelectron Counting and Particle Identification in Scintillation Detectors with Bayesian Techniques
Many current and future dark matter and neutrino detectors are designed to
measure scintillation light with a large array of photomultiplier tubes (PMTs).
The energy resolution and particle identification capabilities of these
detectors depend in part on the ability to accurately identify individual
photoelectrons in PMT waveforms despite large variability in pulse amplitudes
and pulse pileup. We describe a Bayesian technique that can identify the times
of individual photoelectrons in a sampled PMT waveform without deconvolution,
even when pileup is present. To demonstrate the technique, we apply it to the
general problem of particle identification in single-phase liquid argon dark
matter detectors. Using the output of the Bayesian photoelectron counting
algorithm described in this paper, we construct several test statistics for
rejection of backgrounds for dark matter searches in argon. Compared to simpler
methods based on either observed charge or peak finding, the photoelectron
counting technique improves both energy resolution and particle identification
of low energy events in calibration data from the DEAP-1 detector and
simulation of the larger MiniCLEAN dark matter detector.Comment: 16 pages, 16 figure
Measurement of the scintillation time spectra and pulse-shape discrimination of low-energy beta and nuclear recoils in liquid argon with DEAP-1
The DEAP-1 low-background liquid argon detector was used to measure
scintillation pulse shapes of electron and nuclear recoil events and to
demonstrate the feasibility of pulse-shape discrimination (PSD) down to an
electron-equivalent energy of 20 keV.
In the surface dataset using a triple-coincidence tag we found the fraction
of beta events that are misidentified as nuclear recoils to be (90% C.L.) for energies between 43-86 keVee and for a nuclear recoil
acceptance of at least 90%, with 4% systematic uncertainty on the absolute
energy scale. The discrimination measurement on surface was limited by nuclear
recoils induced by cosmic-ray generated neutrons. This was improved by moving
the detector to the SNOLAB underground laboratory, where the reduced background
rate allowed the same measurement with only a double-coincidence tag.
The combined data set contains events. One of those, in the
underground data set, is in the nuclear-recoil region of interest. Taking into
account the expected background of 0.48 events coming from random pileup, the
resulting upper limit on the electronic recoil contamination is
(90% C.L.) between 44-89 keVee and for a nuclear recoil
acceptance of at least 90%, with 6% systematic uncertainty on the absolute
energy scale.
We developed a general mathematical framework to describe PSD parameter
distributions and used it to build an analytical model of the distributions
observed in DEAP-1. Using this model, we project a misidentification fraction
of approx. for an electron-equivalent energy threshold of 15 keV for
a detector with 8 PE/keVee light yield. This reduction enables a search for
spin-independent scattering of WIMPs from 1000 kg of liquid argon with a
WIMP-nucleon cross-section sensitivity of cm, assuming
negligible contribution from nuclear recoil backgrounds.Comment: Accepted for publication in Astroparticle Physic
In-situ characterization of the Hamamatsu R5912-HQE photomultiplier tubes used in the DEAP-3600 experiment
The Hamamatsu R5912-HQE photomultiplier-tube (PMT) is a novel high-quantum
efficiency PMT. It is currently used in the DEAP-3600 dark matter detector and
is of significant interest for future dark matter and neutrino experiments
where high signal yields are needed.
We report on the methods developed for in-situ characterization and
monitoring of DEAP's 255 R5912-HQE PMTs. This includes a detailed discussion of
typical measured single-photoelectron charge distributions, correlated noise
(afterpulsing), dark noise, double, and late pulsing characteristics. The
characterization is performed during the detector commissioning phase using
laser light injected through a light diffusing sphere and during normal
detector operation using LED light injected through optical fibres
The calibration system for the photomultiplier array of the SNO+ experiment
A light injection system using LEDs and optical fibres was designed for the calibration and monitoring of the photomultiplier array of the SNO+ experiment at SNOLAB. Large volume, non-segmented, low-background detectors for rare event physics, such as the multi-purpose SNO+ experiment, need a calibration system that allow an accurate and regular measurement of the performance parameters of their photomultiplier arrays, while minimising the risk of radioactivity ingress. The design implemented for SNO+ uses a set of optical fibres to inject light pulses from external LEDs into the detector. The design, fabrication and installation of this light injection system, as well as the first commissioning tests, are described in this paper. Monte Carlo simulations were compared with the commissioning test results, confirming that the system meets the performance requirements
- …