11 research outputs found
The post-hemodialysis rebound: Predicting and quantifying its effect on Kt/V
The post-hemodialysis rebound: Predicting and quantifying its effect on Kt/V. Immediately after hemodialysis, the urea concentration rebounds upwards as urea continues to be transferred into the arterial circulation from peripheral body compartments. This rebound takes at least 30 minutes to complete. Hemodialysis is quantified as the Kt/V, calculated prom pre- and post-dialysis urea samples. Unless the post-dialysis sample is taken at least 30 minutes after dialysis, the Kt/V will be overestimated. This overestimation will be relatively greater in short high-efficiency dialyses, which have greater post-dialysis rebounds. We propose a method of correction that uses only the conventional pre- and immediate post-dialysis samples and is based on the physiologically-appropriate patient clearance time (tp). This is the time needed to clear all body compartments when the dialyzer clearance is infinite. The tp can be calculated from the pre-, immediate post- and 30-minute post-dialysis urea concentrations and was 35 minutes (SD 16) in 29 patients undergoing short (149 min) hemodiafiltration and standard (243 min) hemodialysis the following week. There was no significant difference between tp values calculated during the two treatments. Standard Kt/V can be corrected by multiplying by t/(t + tp) and dialysis time should be increased by tp × Kt/V minutes to compensate for the rebound. Despite individual variations in tp, a value of tp = 35 was sufficient to correct Kt/V in all patients. Kt/V corrected in this way agreed with Kt/V calculated using a 60-minute post-dialysis sample (r = 0.856, P < 0.001). The method predicted the 60-minute post-rebound concentration (SE 0.5mM, r = 0.983, P < 0.001) and the addition of 35 minutes to the treatment time corrected for the rebound in both conventional and short treatments. Similar simple equations corrected the error in V caused by rebound effects
Validating the use of bioimpedance spectroscopy for assessment of fluid status in children.
BACKGROUND
Bioimpedance spectroscopy (BIS) with a whole-body model to distinguish excess fluid from major body tissue hydration can provide objective assessment of fluid status. BIS is integrated into the Body Composition Monitor (BCM) and is validated in adults, but not children. This study aimed to (1) assess agreement between BCM-measured total body water (TBW) and a gold standard technique in healthy children, (2) compare TBW_BCM with TBW from Urea Kinetic Modelling (UKM) in haemodialysis children and (3) investigate systematic deviation from zero in measured excess fluid in healthy children across paediatric age range.
METHODS
TBW_BCM and excess fluid was determined from standard wrist-to-ankle BCM measurement. TBW_D2O was determined from deuterium concentration decline in serial urine samples over 5Â days in healthy children. UKM was used to measure body water in children receiving haemodialysis. Agreement between methods was analysed using paired t test and Bland-Altman method comparison.
RESULTS
In 61 healthy children (6-14 years, 32 male), mean TBW_BCM and TBW_D2O were 21.1 ± 5.6 and 20.5 ± 5.8 L respectively. There was good agreement between TBW_BCM and TBW_D2O (R = 0.97). In six haemodialysis children (4-13 years, 4 male), 45 concomitant measurements over 8 months showed good TBW_BCM and TBW_UKM agreement (mean difference - 0.4 L, 2SD = ± 3.0 L). In 634 healthy children (2-17 years, 300 male), BCM-measured overhydration was - 0.1 ± 0.7 L (10-90th percentile - 0.8 to + 0.6 L). There was no correlation between age and OH (p = 0.28).
CONCLUSIONS
These results suggest BCM can be used in children as young as 2Â years to measure normally hydrated weight and assess fluid status