41 research outputs found

    Prevalence of bradycardia in 4876 newborns in the first minute after birth and association with positive pressure ventilation: A population-based cross-sectional study

    Get PDF
    Objective To determine the prevalence of bradycardia in the first minute after birth and association with positive pressure ventilation (PPV). Method A population-based cross-sectional study was conducted from June 2019 to December 2021 at Stavanger University Hospital, Norway. Parents consented to participation during pregnancy, and newborns ≥28 weeks’ gestation were included at birth. Heart rate (HR) was captured immediately after birth and continuously for the first minute(s). Time of birth was registered on a tablet. Provision of PPV was captured using video. Results Of 4876 included newborns, 164 (3.4%) did not breathe (two-thirds) or breathed ineffectively (one-third) and received PPV at birth. HR in the first minute had a wide distribution. The prevalence of first measured HR <100 and <60 beats/minute at median 16 s was 16.3% and 0.6%, respectively. HR increased in most cases. At 60 s, 3.7% had HR <100 beats/minute, of which 82% did not require PPV. In total, 25% of newborns had some registered HR <100 beats/minute during the first minute, of which 95% did not require PPV. Among newborns who received PPV, 76% and 62% had HR ≥100 beats/minute at 60 s and at start PPV, respectively. Conclusion Bradycardia with HR <100 bpm in the first minute of life was frequent, but mostly self-resolved. Among the 4% of newborns that remained bradycardic at 60 s, only 20% received PPV. Two-thirds of resuscitated newborns had HR ≥100 beats/minute at start PPV. None of the ventilated newborns were breathing adequately at start PPV.publishedVersio

    Comparison of heart rate feedback from dry-electrode ecg, 3-lead ecg, and pulse oximetry during newborn resuscitation

    Get PDF
    Background: Assessment of heart rate (HR) is essential during newborn resuscitation, and comparison of dry-electrode ECG technology to standard monitoring by 3-lead ECG and Pulse Oximetry (PO) is lacking. Methods: NeoBeat, ECG, and PO were applied to newborns resuscitated at birth. Resuscitations were video recorded, and HR was registered every second. Results: Device placement time from birth was median (quartiles) 6 (4, 18) seconds for NeoBeat versus 138 (97, 181) seconds for ECG and 152 (103, 216) seconds for PO. Time to first HR presentation from birth was 22 (13, 45) seconds for NeoBeat versus 171 (129, 239) seconds for ECG and 270 (185, 357) seconds for PO. Proportion of time with HR feedback from NeoBeat during resuscitation from birth was 85 (69, 93)%, from arrival at the resuscitation table 98 (85, 100)%, and during positive pressure ventilation 100 (95, 100)%. For ECG, these proportions were, 25 (0, 43)%, 28 (0, 56)%, and 33 (0, 66)% and for PO, 0 (0, 16)%, 0 (0, 16)%, and 0 (0, 18)%. All p < 0.0001. Conclusions: NeoBeat was faster to place, presented HR more rapidly, and provided feedback on HR for a larger proportion of time during ongoing resuscitation compared to 3-lead ECG and PO.publishedVersio

    Newborns requiring resuscitation: Two thirds have heart rate ≥100 beats/minute in the first minute after birth

    Get PDF
    Aim The aim was to study the prevalence of bradycardia at birth in newborns requiring positive pressure ventilation (PPV), distribution of first measured heart rate (HR), changes in HR before start of PPV and HR response to PPV. Methods A population-based study including newborns ≥30 weeks' gestation receiving PPV at birth. HR was captured immediately after birth and continuously throughout resuscitation using the dry-electrode ECG device NeoBeat. Time of birth was registered in the Liveborn app. Provision of PPV was captured by video. Results We included 98 newborns receiving PPV at birth. Among newborns with HR measured within 60 s after birth, median (quartiles) first HR was 112 (84, 149) bpm recorded 19 (14, 37) s after birth, of which 33% had first HR <100 and 10% had first HR <60 bpm respectively. First HR was widely distributed. Median HR at start PPV 69 s after birth was 129 bpm. In newborns with an initial low HR, HR typically remained low for 20 s of PPV before increasing rapidly over the next 20–30 s. Conclusions First measured HR was ≥100 bpm in two thirds of newborns receiving PPV. In bradycardic infants, HR did not increase until after 20 s of PPV.publishedVersio

    Heart rate detection properties of dry-electrode ECG compared to conventional 3-lead gel-electrode ECG in newborns

    Get PDF
    Objective To compare the accuracy of heart rate detection properties of a novel, wireless, dry-electrode electrocardiogram (ECG) device, NeoBeat®, to that of a conventional 3-lead gel-electrode ECG monitor (PropaqM®) in newborns. Results The study population had a mean gestational age of 39 weeks and 2 days (1.5 weeks) and birth weight 3528 g (668 g). There were 950 heart rate notations from each device, but heart rate was absent from the reference monitor in 14 of these data points, leaving 936 data pairs to compare. The mean (SD) difference when comparing NeoBeat to the reference monitor was -0.25 (9.91) beats per minute (bpm) (p = 0.44). There was a deviation of more than 10 bpm in 7.4% of the data pairs, which primarily (78%) was attributed to ECG signal disturbance, and secondly (22%) due to algorithm differences between the devices. Excluding these outliers, the correlation was equally consistent (r2 = 0.96) in the full range of heart rate captured measurements with a mean difference of − 0.16 (3.09) bpm. The mean difference was less than 1 bpm regardless of whether outliers were included or not.publishedVersio

    Delivery Room ST Segment Analysis to Predict Short Term Outcomes in Near-Term and Term Newborns

    Get PDF
    Background: ST-segment changes to the fetal electrocardiogram (ECG) may indicate fetal acidosis. No large-scale characterization of ECG morphology immediately after birth has been performed, but ECG is used for heart rate (HR) assessment. We aimed to investigate ECG morphology immediately after birth in asphyxiated infants, using one-lead dry-electrode ECG developed for HR measurement. Methods: Observational study in Tanzania, between 2013-2018. Near-term and term infants that received bag-mask ventilation (BMV), and healthy controls, were monitored with one-lead dry-electrode ECG with a non-diagnostic bandwidth. ECGs were classified as normal, with ST-elevations or other ST-segment abnormalities including a biphasic ST-segment. We analyzed ECG morphology in relation to perinatal variables or short-term outcomes. Results: A total of 494 resuscitated and 25 healthy infants were included. ST-elevations were commonly seen both in healthy infants (7/25; 28%) and resuscitated (320/494; 65%) infants. The apparent ST-elevations were not associated with perinatal variables or short-term outcomes. Among the 32 (6.4%) resuscitated infants with other ST-segment abnormalities , duration of BMV was longer, 1-min Apgar score lower and normal outcomes less frequent than in the resuscitated infants with normal ECG or ST-elevations. Conclusions: ST-segment elevation was commonly seen and not associated with negative outcomes when using one-lead dry-electrode ECG. Other ST-segment abnormalities were associated with prolonged BMV and worse outcome. ECG with appropriate bandwidth and automated analysis may potentially in the future aid in the identification of severely asphyxiated infants

    Delivery Room ST Segment Analysis to Predict Short Term Outcomes in Near-Term and Term Newborns

    Get PDF
    Background: ST-segment changes to the fetal electrocardiogram (ECG) may indicate fetal acidosis. No large-scale characterization of ECG morphology immediately after birth has been performed, but ECG is used for heart rate (HR) assessment. We aimed to investigate ECG morphology immediately after birth in asphyxiated infants, using one-lead dry-electrode ECG developed for HR measurement. Methods: Observational study in Tanzania, between 2013–2018. Near-term and term infants that received bag-mask ventilation (BMV), and healthy controls, were monitored with one-lead dry-electrode ECG with a non-diagnostic bandwidth. ECGs were classified as normal, with ST-elevations or other ST-segment abnormalities including a biphasic ST-segment. We analyzed ECG morphology in relation to perinatal variables or short-term outcomes. Results: A total of 494 resuscitated and 25 healthy infants were included. ST-elevations were commonly seen both in healthy infants (7/25; 28%) and resuscitated (320/494; 65%) infants. The apparent ST-elevations were not associated with perinatal variables or short-term outcomes. Among the 32 (6.4%) resuscitated infants with “other ST-segment abnormalities”, duration of BMV was longer, 1-min Apgar score lower and normal outcomes less frequent than in the resuscitated infants with normal ECG or ST-elevations. Conclusions: ST-segment elevation was commonly seen and not associated with negative outcomes when using one-lead dry-electrode ECG. Other ST-segment abnormalities were associated with prolonged BMV and worse outcome. ECG with appropriate bandwidth and automated analysis may potentially in the future aid in the identification of severely asphyxiated infants

    Incomplete Exhalation during Resuscitation—Theoretical Review and Examples from Ventilation of Newborn Term Infants

    Get PDF
    Background: Newborn resuscitation guidelines recommend positive pressure ventilation (PPV) for newborns who do not establish effective spontaneous breathing after birth. T-piece resuscitator systems are commonly used in high-resource settings and can additionally provide positive end-expiratory pressure (PEEP). Short expiratory time, high resistance, rapid dynamic changes in lung compliance and large tidal volumes increase the possibility of incomplete exhalation. Previous publications indicate that this may occur during newborn resuscitation. Our aim was to study examples of incomplete exhalations in term newborn resuscitation and discuss these against the theoretical background. Methods: Examples of flow and pressure data from respiratory function monitors (RFM) were selected from 129 term newborns who received PPV using a T-piece resuscitator. RFM data were not presented to the user during resuscitation. Results: Examples of incomplete exhalation with higher-than-set PEEP-levels were present in the recordings with visual correlation to factors affecting time needed to complete exhalation. Conclusions: Incomplete exhalation and the relationship to expiratory time constants have been well described theoretically. We documented examples of incomplete exhalations with increased PEEP-levels during resuscitation of term newborns. We conclude that RFM data from resuscitations can be reviewed for this purpose and that incomplete exhalations should be further explored, as the clinical benefit or risk of harm are not known.publishedVersio

    Expired carbon dioxide during newborn resuscitation as predictor of outcome

    Get PDF
    Aim: To explore and compare expired CO2 (ECO2) and heart rate (HR), during newborn resuscitation with bag-mask ventilation, as predictors of 24-h outcome. Methods: Observational study from March 2013 to June 2017 in a rural Tanzanian hospital. Side-stream measures of ECO2, ventilation parameters, HR, clinical information, and 24-h outcome were recorded in live born bag-mask ventilated newborns with initial HR \u3c 120 bpm. We analysed the data using logistic regression models and compared areas under the receiver operating curves (AUC) for ECO2 and HR within three selected time intervals after onset of ventilation (0-30 s, 30.1-60 s and 60.1-300 s). Results: Among 434 included newborns (median birth weight 3100 g), 378 were alive at 24 h, 56 had died. Both ECO2 and HR were independently significant predictors of 24-h outcome, with no differences in AUCs. In the first 60 s of ventilation, ECO2 added extra predictive information compared to HR alone. After 60 s, ECO2 lost significance when adjusted for HR. In 70% of newborns with initial ECO2 Conclusions: Higher levels and a faster rise in ECO2 and HR during newborn resuscitation were independently associated with improved survival compared to persisting low values. ECO2 increased before HR and may serve as an earlier predictor of survival

    Newborn resuscitation simulation training and changes in clinical performance and perinatal outcomes: a clinical observational study of 10,481 births

    Get PDF
    Annually, 1.5 million intrapartum-related deaths occur; fresh stillbirths and early newborn deaths. Most of these deaths are preventable with skilled ventilation starting within the first minute of life. Helping Babies Breathe is an educational program shown to improve simulated skills in newborn resuscitation. However, translation into clinical practice remains a challenge. The aim was to describe changes in clinical resuscitation and perinatal outcomes (i.e., fresh stillbirths and 24-h newborn deaths) after introducing a novel simulator (phase 1) and then local champions (phase 2) to facilitate ongoing Helping Babies Breathe skill and scenario simulation training. Methods This is a 3-year prospective before/after (2 phases) clinical observational study in Tanzania. Research assistants observed all deliveries from September 2015 through August 2018 and recorded labor/newborn information and perinatal outcomes. A novel simulator with automatic feedback to stimulate self-guided skill training was introduced in September 2016. Local champions were introduced in October 2017 to motivate midwives for weekly training, also team simulations. Results The study included 10,481 births. Midwives had practiced self-guided skill training during the last week prior to a real newborn resuscitation in 34% of cases during baseline, 30% in phase 1, and 71% in phase 2. Most real resuscitations were provided by midwives, increasing from 66% in the baseline, to 77% in phase 1, and further to 83% in phase 2. The median time from birth to first ventilation decreased between baseline and phase 2 from 118 (85–165) to 101 (72–150) s, and time pauses during ventilation decreased from 28 to 16%. Ventilations initiated within the first minute did not change significantly (13–16%). The proportion of high-risk deliveries increased during the study period, while perinatal mortality remained unchanged. Conclusions This study reports a gradual improvement in real newborn resuscitation skills after introducing a novel simulator and then local champions. The frequency of trainings increased first after the introduction of motivating champions. Time from birth to first ventilation decreased; still, merely 16% of newborns received ventilation within the first minute as recommended. This is a remaining challenge that may require more targeted team-scenario training and quality improvement efforts to improve.publishedVersio

    Video Analysis of Newborn Resuscitations After Simulation-Based Helping Babies Breathe Training

    Get PDF
    Background: Simulation-based Helping Babies Breathe (HBB) training is currently rolled-out in around 80 low-income countries with various results. Method: Workflow was analyzed in 76 video-recorded newborn resuscitations performed by regularly HBB-trained nurse-midwives over 3 years in rural Tanzania. Results: Actual newborn resuscitation practice deviated from HBB intention/guideline: most newborns underwent prolonged suction and stimulation before ventilation; ventilation was delayed and frequently interrupted. Nurse-midwives often worked together. Conclusions:There is a gap between training intention and clinical practice. HBB trainings should focus more on urgency, ventilation skills, and team training. Combining clinical debriefing with HBB simulations could facilitate continuous learning and applicatio
    corecore