5 research outputs found

    Pitfalls and premature failure of the Freedom SOLO stentless valve

    Get PDF
    OBJECTIVES This study reports a series of pitfalls, premature failures and explantations of the third-generation Freedom SOLO (FS) bovine pericardial stentless valve. METHODS A total of 149 patients underwent aortic valve replacement using the FS. Follow-up was 100% complete with an average observation time of 5.5 ± 2.3 years (maximum 8.7 years) and a total of 825 patient-years. Following intraoperative documentation, all explanted valve prostheses underwent histological examination. RESULTS Freedom from structural valve deterioration (SVD) at 5, 6, 7, 8 and 9 years was 92, 88, 80, 70 and 62%, respectively. Fourteen prostheses required explantation due to valve-independent dysfunction (n = 5; i.e. thrombus formation, oversizing, aortic dilatation, endocarditis and suture dehiscence) or valve-dependent failure (acute leaflet tears, n = 4 and severe stenosis, n = 5). Thus, freedom from explantation at 5, 6, 7, 8 and 9 years was 95, 94, 91, 81 and 72%, respectively. An acute vertical tear along the non-coronary/right coronary commissure to the base occurred at a mean of 6.0 years (range 4.3-7.3 years) and affected size 25 and 27 prostheses exclusively. Four FS required explantation after a mean of 7.5 years (range 7.0-8.3 years) due to severe functional stenosis and gross calcification that included the entire aortic root. CONCLUSIONS The FS stentless valve is safe to implant and shows satisfying mid-term results in our single institution experience. Freedom from SVD and explantation decreased markedly after only 6-7 years, so that patients with FS require close observation and follow-up. Exact sizing, symmetric positioning and observing patient limitations are crucial for optimal outcom

    Association of bilateral acute anterior uveitis with a capsaicin patch

    No full text
    PURPOSE To report a case of bilateral acute anterior uveitis in association with the application of an analgesic transdermal capsaicin patch. METHODS Case report and review of literature. RESULTS A 38-year-old woman suffered from bilateral acute anterior uveitis, manifesting 12-24 h after application of an analgesic capsaicin patch (Isola Capsicum N Plus) that served to alleviate muscular neck pain. Systemic immune-mediated and infectious diseases were excluded by medical history and laboratory testing. Control of inflammation was achieved with topical corticosteroid treatment within 1 week. Over the course of a 1-year follow-up, no further recurrence of uveitis was observed. CONCLUSIONS The case suggests a possible association of capsaicin-containing transdermal patches and acute anterior uveitis

    Pitfalls and premature failure of the Freedom SOLO stentless valve.

    Get PDF
    OBJECTIVES This study reports a series of pitfalls, premature failures and explantations of the third-generation Freedom SOLO (FS) bovine pericardial stentless valve. METHODS A total of 149 patients underwent aortic valve replacement using the FS. Follow-up was 100% complete with an average observation time of 5.5 ± 2.3 years (maximum 8.7 years) and a total of 825 patient-years. Following intraoperative documentation, all explanted valve prostheses underwent histological examination. RESULTS Freedom from structural valve deterioration (SVD) at 5, 6, 7, 8 and 9 years was 92, 88, 80, 70 and 62%, respectively. Fourteen prostheses required explantation due to valve-independent dysfunction (n = 5; i.e. thrombus formation, oversizing, aortic dilatation, endocarditis and suture dehiscence) or valve-dependent failure (acute leaflet tears, n = 4 and severe stenosis, n = 5). Thus, freedom from explantation at 5, 6, 7, 8 and 9 years was 95, 94, 91, 81 and 72%, respectively. An acute vertical tear along the non-coronary/right coronary commissure to the base occurred at a mean of 6.0 years (range 4.3-7.3 years) and affected size 25 and 27 prostheses exclusively. Four FS required explantation after a mean of 7.5 years (range 7.0-8.3 years) due to severe functional stenosis and gross calcification that included the entire aortic root. CONCLUSIONS The FS stentless valve is safe to implant and shows satisfying mid-term results in our single institution experience. Freedom from SVD and explantation decreased markedly after only 6-7 years, so that patients with FS require close observation and follow-up. Exact sizing, symmetric positioning and observing patient limitations are crucial for optimal outcome

    The Freedom SOLO bovine pericardial stentless valve: Single-center experience, outcome and long-term durability

    Get PDF
    Abstract Objectives We report our institutional experience and long-term results with the Sorin Freedom SOLO bovine pericardial stentless bioprosthesis. Methods Between January 2005 and November 2009, 149 patients (mean age 73.6±8.7 years, 68 [45.6%] female) underwent isolated (n=75) or combined (n=74) aortic valve replacement (AVR) using the SOLO in our institution. Follow-up was 100% complete with an average follow-up time of 5.9±2.6 years (maximum 9.6 years) and a total of 885.3 patient years. Results Operative (30-day) mortality was 2.7% (1.3% for isolated AVR [n=1] and 4.0% for combined procedures [n=3]). All causes of death were not valve-related. Preoperative peak (mean) gradients of 74.2±23.0 mmHg (48.6 ± 16.3 mmHg) decreased to 15.6±5.4 (8.8±3.0) after AVR, and remained low for up to 9 years. The postoperative effective orifice area (EOA) was 1.6 ±0.57 cm2, 1.90±0.45 cm2, 2.12±0.48 cm2 and 2.20±0.66 cm2 for the valve sizes 21, 23, 25 and 27, respectively; with absence of severe prosthesis-patient-mismatch (PPM) and 0.7% (n=1) moderate PPM. During follow-up, Twenty-six patients experienced structural valve deterioration (SVD) and 14 patients underwent explantation. Kaplan-Meier estimates for freedom from death, explantation and SVD at 9 years averaged 0.57 [0.47‒0.66], 0.82 [0.69‒0.90] and 0.70 [0.57‒0.79], respectively. Conclusions The Freedom SOLO stentless aortic valve is safe to implant and shows excellent early and mid-term hemodynamic performance. However, SVD was observed in a substantial number of patients after only 5 ̶ 6 years and the need for explantation increased markedly, suggesting low durability
    corecore