Heart valve replacement with the Björk-Shiley and St Jude Medical prostheses: A randomized comparison in 178 patients
In 178 patients, a randomized prospective comparison between the 60° spherical disc Björk-Shiley (BS) and the St Jude Medical (SJM) heart valve prostheses was performed. Four-week perioperative mortality was zero in the BS (n = 84) and 4.3% in the SJM group (n = 94). During a mean ( ± SD) follow-up of 52 ± 20 months or 778 patient-years, late cardiac mortality per year was 2.4% in the BS and 2.2% in the SJM group. The yearly thromboembolic rates were 1.4% in the BS and 2.0% in the SJM group. There was no mechanical valve failure or haemolytic anaemia. Paravalvular leaks and major bleeding complications occurred at low rates in both groups (1.1% and2.2% per year in BS; 0.7% and 1.7% per year in SJM). Functional results were similarly good with 96% of patients with BS valves and 95% of patients with SJM prostheses being in NYHA classes I and II, respectively. We conclude that heart valve replacement with mechanical prostheses can be performed with equally good results using either the Björk-Shiley spherical disc valve or the St Jude Medical bileaflet prosthesi
Solide Raumforderung im vorderen Mediastinum
Please help us populate SUNScholar with the post print version of this article. It can be e-mailed to: [email protected] Geneeskund
ZENTRAS. Entwicklung eines zuverlaessigen Energieuebertragungs-Systems fuer den Duo-Bus Schlussbericht