Dose assessment of digital tomosynthesis in pediatric imaging
YesWe investigated the potential for digital tomosynthesis (DT) to reduce pediatric x-ray dose while maintaining
image quality. We utilized the DT feature (VolumeRadTM) on the GE DefiniumTM 8000 flat panel system installed in the
Winnipeg Children’s Hospital. Facial bones, cervical spine, thoracic spine, and knee of children aged 5, 10, and 15 years
were represented by acrylic phantoms for DT dose measurements. Effective dose was estimated for DT and for
corresponding digital radiography (DR) and computed tomography (CT) patient image sets. Anthropomorphic phantoms
of selected body parts were imaged by DR, DT, and CT. Pediatric radiologists rated visualization of selected anatomic
features in these images. Dose and image quality comparisons between DR, DT, and CT determined the usefulness of
tomosynthesis for pediatric imaging.
CT effective dose was highest; total DR effective dose was not always lowest – depending how many
projections were in the DR image set. For the cervical spine, DT dose was close to and occasionally lower than DR
dose. Expert radiologists rated visibility of the central facial complex in a skull phantom as better than DR and
comparable to CT. Digital tomosynthesis has a significantly lower dose than CT. This study has demonstrated DT
shows promise to replace CT for some facial bones and spinal diagnoses. Other clinical applications will be evaluated in
the future.Medical Physics Division at CancerCare Manitoba and GE Healthcare (Waukesha, WI)
Context sensitive cardiac x-ray imaging: a machine vision approach to x-ray dose control
Modern cardiac x-ray imaging systems regulate their radiation output based on the thickness of the patient to maintain an acceptable signal at the input of the x-ray detector. This approach does not account for the context of the examination or the content of the image displayed. We have developed a machine vision algorithm that detects iodine-filled blood vessels and fits an idealized vessel model with the key parameters of contrast, diameter, and linear attenuation coefficient. The spatio-temporal distribution of the linear attenuation coefficient samples, when appropriately arranged, can be described by a simple linear relationship, despite the complexity of scene information. The algorithm was tested on static anthropomorphic chest phantom images under different radiographic factors and 60 dynamic clinical image sequences. It was found to be robust and sensitive to changes in vessel contrast resulting from variations in system parameters. The machine vision algorithm has the potential of extracting real-time context sensitive information that may be used for augmenting existing dose control strategies
Can Image Enhancement Allow Radiation Dose to Be Reduced Whilst Maintaining the Perceived Diagnostic Image Quality Required for Coronary Angiography?
Objectives: The aim of this research was to quantify the reduction in radiation dose facilitated by image processing alone for percutaneous coronary intervention (PCI) patient angiograms, without reducing the perceived image quality required to confidently make a diagnosis. Methods: Incremental amounts of image noise were added to five PCI angiograms, simulating the angiogram as having been acquired at corresponding lower dose levels (10-89% dose reduction). Sixteen observers with relevant experience scored the image quality of these angiograms in three states - with no image processing and with two different modern image processing algorithms applied. These algorithms are used on state-of-the-art and previous generation cardiac interventional X-ray systems. Ordinal regression allowing for random effects and the delta method were used to quantify the dose reduction possible by the processing algorithms, for equivalent image quality scores. Results: Observers rated the quality of the images processed with the state-of-the-art and previous generation image processing with a 24.9% and 15.6% dose reduction respectively as equivalent in quality to the unenhanced images. The dose reduction facilitated by the state-of-the-art image processing relative to previous generation processing was 10.3%. Conclusions: Results demonstrate that statistically significant dose reduction can be facilitated with no loss in perceived image quality using modern image enhancement; the most recent processing algorithm was more effective in preserving image quality at lower doses. Advances in knowledge: Image enhancement was shown to maintain perceived image quality in coronary angiography at a reduced level of radiation dose using computer software to produce synthetic images from real angiograms simulating a reduction in dose
How much image noise can be added in cardiac x-ray imaging without loss in perceived image quality?
YesCardiologists use x-ray image sequences of the moving heart acquired in real-time to diagnose and
treat cardiac patients. The amount of radiation used is proportional to image quality; however, exposure to radiation
is damaging to patients and personnel. The amount by which radiation dose can be reduced without compromising
patient care was determined. For five patient image sequences, increments of computer-generated
quantum noise (white + colored) were added to the images, frame by frame using pixel-to-pixel addition, to
simulate corresponding increments of dose reduction. The noise adding software was calibrated for settings
used in cardiac procedures, and validated using standard objective and subjective image quality measurements.
The degraded images were viewed next to corresponding original (not degraded) images in a two-alternativeforced-
choice staircase psychophysics experiment. Seven cardiologists and five radiographers selected their
preferred image based on visualization of the coronary arteries. The point of subjective equality, i.e., level
of degradation where the observer could not perceive a difference between the original and degraded images,
was calculated; for all patients the median was 33% 15% dose reduction. This demonstrates that a 33%
15% increase in image noise is feasible without being perceived, indicating potential for 33% 15% dose reduction
without compromising patient care.Funded in part by Philips Healthcare, the Netherlands. Part of this work has been performed in the project PANORAMA, co-funded by grants from Belgium, Italy, France, the Netherlands, and the United Kingdom, and the ENIAC Joint Undertaking
Methods for the analysis of ordinal response data in medical image quality assessment.
The assessment of image quality in medical imaging often requires observers to rate images for some metric or detectability task. These subjective results are used in optimisation, radiation dose reduction or system comparison studies and may be compared to objective measures from a computer vision algorithm performing the same task. One popular scoring approach is to use a Likert scale, then assign consecutive numbers to the categories. The mean of these response values is then taken and used for comparison with the objective or second subjective response. Agreement is often assessed using correlation coefficients. We highlight a number of weaknesses in this common approach, including inappropriate analyses of ordinal data, and the inability to properly account for correlations caused by repeated images or observers. We suggest alternative data collection and analysis techniques such as amendments to the scale and multilevel proportional odds models. We detail the suitability of each approach depending upon the data structure and demonstrate each method using a medical imaging example. Whilst others have raised some of these issues, we evaluated the entire study from data collection to analysis, suggested sources for software and further reading, and provided a checklist plus flowchart, for use with any ordinal data. We hope that raised awareness of the limitations of the current approaches will encourage greater method consideration and the utilisation of a more appropriate analysis. More accurate comparisons between measures in medical imaging will lead to a more robust contribution to the imaging literature and ultimately improved patient care
Recommended from our members
X-ray beam optimisation for paediatric interventional cardiac imaging: paesiatric-specific concerns for radiation dose
N
Recommended from our members
Dose optimization in pediatric cardiac x-ray imaging
NoPurpose: The aim of this research was to explore x-ray beam parameters with intent to optimize
pediatric x-ray settings in the cardiac catheterization laboratory. This study examined the effects of
peak x-ray tube voltage kVp and of copper Cu x-ray beam filtration independently on the image
quality to dose balance for pediatric patient sizes. The impact of antiscatter grid removal on the
image quality to dose balance was also investigated.
Methods: Image sequences of polymethyl methacrylate phantoms approximating chest sizes typical
of pediatric patients were captured using a modern flat-panel receptor based x-ray imaging system.
Tin was used to simulate iodine-based contrast medium used in clinical procedures. Measurements
of tin detail contrast and flat field image noise provided the contrast to noise ratio. Entrance surface
dose ESD and effective dose E measurements were obtained to calculate the figure of merit
FOM , CNR2 / dose, which evaluated the dose efficiency of the x-ray parameters investigated. The
kVp, tube current mA , and pulse duration were set manually by overriding the system’s automatic
dose control mechanisms. Images were captured with 0, 0.1, 0.25, 0.4, and 0.9 mm added Cu
filtration, for 50, 55, 60, 65, and 70 kVp with the antiscatter grid in place, and then with it removed.
Results: For a given phantom thickness, as the Cu filter thickness was increased, lower kVp was
favored. Examining kVp alone, lower values were generally favored, more so for thinner phantoms.
Considering ESD, the 8.5 cm phantom had the highest FOM at 50 kVp using 0.4 mm of Cu
filtration. The 12 cm phantom had the highest FOM at 55 kVp using 0.9 mm Cu, and the 16 cm
phantom had highest FOM at 55 kVp using 0.4 mm Cu. With regard to E, the 8.5 and 12 cm
phantoms had the highest FOM at 50 kVp using 0.4 mm of Cu filtration, and the 16 cm phantom
had the highest FOM at 50 kVp using 0.25 mm Cu. Antiscatter grid removal improved the FOM for
a given set of x-ray conditions. Under aforesaid optimal settings, the 8.5 cm phantom FOM improved
by 24% and 33% for ESD and E, respectively. Corresponding improvements were 26% and
24% for the 12 cm phantom and 6% and 15% for the 16 cm phantom.
Conclusions: For pediatric patients, using 0.25–0.9 mm Cu filtration in the x-ray beam while
maintaining 50–55 kVp, depending on patient size, provided optimal x-ray image quality to dose
ratios. These settings, adjusted for x-ray tube loading limits and clinically acceptable image quality,
should provide a useful strategy for optimizing iodine contrast agent based cardiac x-ray imaging.
Removing the antiscatter grid improved the FOM for the 8.5 and 12 cm phantoms, therefore grid
removal is recommended for younger children. Improvement for the 16 cm phantom declined into
the estimated margin of error for the FOM; the need for grid removal for older children would
depend on practical feasibility in the clinical environment.Philips Healthcar