78 research outputs found

    Improved retrieval of nitrogen dioxide (NO2) column densities by means of MKIV Brewer spectrophotometers

    Get PDF
    A new algorithm to retrieve nitrogen dioxide (NO2) column densities using MKIV ("Mark IV") Brewer spectrophotometers is described. The method includes several improvements, such as a more recent spectroscopic data set, the reduction of measurement noise, interference by other atmospheric species and instrumental settings, and a better determination of the zenith sky air mass factor. The technique was tested during an ad hoc calibration campaign at the high-altitude site of Izaña (Tenerife, Spain) and the results of the direct sun and zenith sky geometries were compared to those obtained by two reference instruments from the Network for the Detection of Atmospheric Composition Change (NDACC): a Fourier Transform Infrared Radiometer (FTIR) and an advanced visible spectrograph (RASAS-II) based on the differential optical absorption spectrometry (DOAS) technique

    Antarctic network of lamp-calibrated multichannel radiometers for continuous ozone and uv radiation data

    Get PDF
    International audienceThree NILU-UV multichannel radiometers have been installed in 1999 at the Argentinian sites of Ushuaia (54S), Marambio (64S) and Belgrano-II (77S) in order to continuously monitor UV radiation, photosynthetically active radiation and total ozone. The measurements were established by INM, Spain in collaboration with FMI, Finland, DNA-IAA, Argentina and CADIC, Argentina to observe and characterize the spatial and temporal evolution of ozone and ultraviolet radiation in the Antarctic region. Special attention has been given to the quality control and quality assurance of the measurements under harsh climatological conditions. The ozone and UV time series of 2000?2006 were calibrated using a polynomial fit for lamp measurements performed every second week all year round. The gaps in these data are minimal, with almost no data missing, and the data products are available from http://www.polarvortex.org in near real time. The data products include the erythemally-weighted UV, UVB and UVA radiation, photosynthetically active radiation (PAR), total ozone (O3) and a cloud parameter (CLT). For UV data, dose rates as well as daily doses are available; from these the maximum measured UV indices (UVI), during 2000?2006, were 12.0, 9.7 and 8.1 at Ushuaia, Marambio and Belgrano-II, respectively

    Description of the Baseline Surface Radiation Network (BSRN) station at the Izaña Observatory (2009–2017): measurements and quality control/assurance procedures

    Get PDF
    The Baseline Surface Radiation Network (BSRN) was implemented by the World Climate Research Programme (WCRP) starting observations with nine stations in 1992, under the auspices of the World Meteorological Organization (WMO). Currently, 59 BSRN stations submit their data to the WCRP. One of these stations is the Izaña station (station IZA, no. 61) that enrolled in this network in 2009. This is a high-mountain station located in Tenerife (Canary Islands, Spain, at 28.3∘ N, 16.5∘ W; 2373 m a.s.l.) and is a representative site of the subtropical North Atlantic free troposphere. It contributes with basic-BSRN radiation measurements, such as global shortwave radiation (SWD), direct radiation (DIR), diffuse radiation (DIF) and longwave downward radiation (LWD), and extended-BSRN measurements, including ultraviolet ranges (UV-A and UV-B), shortwave upward radiation (SWU) and longwave upward radiation (LWU), and other ancillary measurements, such as vertical profiles of temperature, humidity and wind obtained from radiosonde profiles (WMO station no. 60018) and total column ozone from the Brewer spectrophotometer. The IZA measurements present high-quality standards since more than 98 % of the data are within the limits recommended by the BSRN. There is an excellent agreement in the comparison between SWD, DIR and DIF (instantaneous and daily) measurements with simulations obtained with the LibRadtran radiative transfer model. The root mean square error (RMSE) for SWD is 2.28 % for instantaneous values and 1.58 % for daily values, while the RMSE for DIR is 2.00 % for instantaneous values and 2.07 % for daily values. IZA is a unique station that provides very accurate solar radiation data in very contrasting scenarios: most of the time under pristine sky conditions and periodically under the effects of the Saharan air layer characterized by a high content of mineral dust. A detailed description of the BSRN program at IZA, including quality control and quality assurance activities, is given in this work.The IZA BSRN program has benefited from results obtained within POLARMOON project funded by the Ministerio de Economía y Competividad from Spain, CTM2015-66742-R

    Description of the Baseline Surface Radiation Network (BSRN) station at the Izaña Observatory (2009–2017): measurements and quality control/assurance procedures [Discussion]

    Get PDF
    The Baseline Surface Radiation Network (BSRN) was implemented by the World Climate Research Programme (WRCP) starting observations with 9 stations in 1992, under the auspices of the World Meteorological Organization (WMO). Currently, 59 BSRN stations submit their data to the WRCP. One of these stations is the Izaña station (Station: IZA, #61) that enrolled in this network in 2009. This is a high-mountain station located in Tenerife (Canary Islands, Spain; at 28.3°N, 16.5°W, 2373ma.s.l.) and is a representative site of the subtropical North Atlantic free troposphere. It contributes with basic-BSRN radiation measurements, such as, global shortwave radiation (SWD), direct radiation (DIR), diffuse radiation (DIF) and longwave downward radiation (LWD) and extended-BSRN measurements, including ultraviolet ranges (UV-A and UV-B), shortwave upward radiation (SWU) and longwave upward radiation (LWU) and other ancillary measurements, such as vertical profiles of temperature, humidity and wind obtained from radiosonde (WMO, station #60018) and total column ozone from Brewer spectrophotometer. The IZA measurements present high quality standards since more than 98% of the data are within the limits recommended by the BSRN. There is an excellent agreement in the comparison between SWD, DIR and DIF (instantaneous and daily) measurements with simulations obtained with the LibRadtran radiative transfer model. The root mean square error (RMSE) for SWD is 2.28% for instantaneous values and 1.58% for daily values, while the RMSE for DIR is 2.00% for instantaneous values and 2.07% for daily values. IZA is a unique station that provides very accurate solar radiation data in very contrasting scenarios: most of the time under pristine sky conditions, and periodically under the effects of the Saharan Air Layer characterized by a high content of mineral dust. A detailed description of the BSRN program at IZA, including quality control and quality assurance activities, is given in this work.The IZA BSRN program has benefited from results obtained within POLARMOON project funded by the Ministerio de Economía y Competividad from Spain, CTM2015-66742-R

    Quality assurance of the solar UV network in the Antarctic

    Get PDF
    Measuring ultraviolet radiation in the Antarctic region, where weather conditions are extremely challenging, is a demanding task. Proper quality control of the measurements and quality assurance of the data, which are the basis of all scientific use of data, has to be especially well planned and executed. In this paper we show the importance of proper quality assurance and describe the methods used to successfully operate the NILU-UV multichannel radiometers of the Antarctic network stations at Ushuaia, 54S, and Marambio, 64S. According to our experience, even though multichannel instruments are supposed to be rather stable as a function of time, severe drifts can occur in the sensitivity of the channels under these harsh conditions. During 2000–2003 the biggest drifts were 35%, both at Ushuaia and Marambio, with the sensitivity of the channels dropping at different rates. Without proper corrections in the data, this would have seriously affected the calculated UV dose rates. As part of the quality assurance of the network a traveling reference NILU-UV, which was found to be stable, was used to transfer the desired irradiance scale to the site NILU-UV data. Relative lamp tests were used to monitor the stability of the instruments. Each site NILU-UV was scaled channel by channel to the traveling reference by performing solar comparisons. The method of scaling each channel separately was found to be successful, even though the differences between the raw data of the site NILU-UV and the reference instruments were, before the data correction, as much as 40%. After the correction, the mean ratios of erythemally weighted UV dose rates measured during the solar comparisons in 2000–2003 between the reference NILU-UV and the site NILU-UV were 1.007 ± 0.011 and 1.012 ± 0.012 for Ushuaia and Marambio, respectively, when the solar zenith angle varied up to 80. These results make possible the scientific use of NILU-UV data measured simultaneously at quite different locations, e.g., the Antarctic and Arctic, and the method presented is also practicable for other multichannel radiometer networks.S, and Marambio, 64S. According to our experience, even though multichannel instruments are supposed to be rather stable as a function of time, severe drifts can occur in the sensitivity of the channels under these harsh conditions. During 2000–2003 the biggest drifts were 35%, both at Ushuaia and Marambio, with the sensitivity of the channels dropping at different rates. Without proper corrections in the data, this would have seriously affected the calculated UV dose rates. As part of the quality assurance of the network a traveling reference NILU-UV, which was found to be stable, was used to transfer the desired irradiance scale to the site NILU-UV data. Relative lamp tests were used to monitor the stability of the instruments. Each site NILU-UV was scaled channel by channel to the traveling reference by performing solar comparisons. The method of scaling each channel separately was found to be successful, even though the differences between the raw data of the site NILU-UV and the reference instruments were, before the data correction, as much as 40%. After the correction, the mean ratios of erythemally weighted UV dose rates measured during the solar comparisons in 2000–2003 between the reference NILU-UV and the site NILU-UV were 1.007 ± 0.011 and 1.012 ± 0.012 for Ushuaia and Marambio, respectively, when the solar zenith angle varied up to 80. These results make possible the scientific use of NILU-UV data measured simultaneously at quite different locations, e.g., the Antarctic and Arctic, and the method presented is also practicable for other multichannel radiometer networks.S. According to our experience, even though multichannel instruments are supposed to be rather stable as a function of time, severe drifts can occur in the sensitivity of the channels under these harsh conditions. During 2000–2003 the biggest drifts were 35%, both at Ushuaia and Marambio, with the sensitivity of the channels dropping at different rates. Without proper corrections in the data, this would have seriously affected the calculated UV dose rates. As part of the quality assurance of the network a traveling reference NILU-UV, which was found to be stable, was used to transfer the desired irradiance scale to the site NILU-UV data. Relative lamp tests were used to monitor the stability of the instruments. Each site NILU-UV was scaled channel by channel to the traveling reference by performing solar comparisons. The method of scaling each channel separately was found to be successful, even though the differences between the raw data of the site NILU-UV and the reference instruments were, before the data correction, as much as 40%. After the correction, the mean ratios of erythemally weighted UV dose rates measured during the solar comparisons in 2000–2003 between the reference NILU-UV and the site NILU-UV were 1.007 ± 0.011 and 1.012 ± 0.012 for Ushuaia and Marambio, respectively, when the solar zenith angle varied up to 80. These results make possible the scientific use of NILU-UV data measured simultaneously at quite different locations, e.g., the Antarctic and Arctic, and the method presented is also practicable for other multichannel radiometer networks.. These results make possible the scientific use of NILU-UV data measured simultaneously at quite different locations, e.g., the Antarctic and Arctic, and the method presented is also practicable for other multichannel radiometer networks.Fil: Lakkala, K.. Finnish Meteorological Institute; FinlandiaFil: Redondas, A.. Instituto Nacional de Meteorología; EspañaFil: Meinander, O.. Finnish Meteorological Institute; FinlandiaFil: Torres ,Carlos. Instituto Nacional de Meteorología; EspañaFil: Koskela, T.. Finnish Meteorological Institute; FinlandiaFil: Cuevas, Eduardo. Instituto Nacional de Meteorología; EspañaFil: Taalas, P.. Finnish Meteorological Institute; FinlandiaFil: Dahlback, A.. University of Oslo; NoruegaFil: Deferrari, Guillermo Alejandro. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Austral de Investigaciones Científicas; ArgentinaFil: Edvardsen, K.. Instituto Noruego de Investigación del Aire; NoruegaFil: Ochoa, H.. Ministerio de Relaciones Exteriores, Comercio Interno y Culto. Dirección Nacional del Antártico. Instituto Antártico Argentino; Argentin

    New observations of upper tropospheric NO2 from TROPOMI

    Get PDF
    Nitrogen oxides (NOx ≡ NO + NO2) in the NOx-limited upper troposphere (UT) are long-lived and so have a large influence on the oxidizing capacity of the troposphere and formation of the greenhouse gas ozone. Models misrepresent NOx in the UT and observations to address deficiencies in models are sparse. Here we obtain a year of near-global seasonal mean mixing ratios of NO2 in the UT (450–180 hPa) at 1 ° x 1° by applying cloud-slicing to partial columns of NO2 from TROPOMI. This follows refinement of the cloud-slicing algorithm with synthetic partial columns from the GEOS-Chem chemical transport model. We find that synthetic cloud-sliced UT NO2 are spatially consistent (R = 0.64) with UT NO2 calculated across the same cloud pressure range and scenes as are cloud-sliced (“true” UT NO2), but the cloud-sliced UT NO2 is 11–22 % more than the "true" all-sky seasonal mean. The largest contributors to differences between synthetic cloud-sliced and “true” UT NO2 are target resolution of the cloud-sliced product and uniformity of overlying stratospheric NO2. TROPOMI, prior to cloud-slicing, is corrected for a 13 % underestimate in stratospheric NO2 variance and a 50 % overestimate in free tropospheric NO2 determined by comparison to Pandora total columns at high-altitude sites in Mauna Loa, Izaña and Altzomoni, and MAX-DOAS and Pandora tropospheric columns at Izaña. Two cloud-sliced seasonal mean UT NO2 products for June 2019 to May 2020 are retrieved from corrected TROPOMI total columns using distinct TROPOMI cloud products that assume clouds are reflective boundaries (FRESCO-S) or water droplet layers (ROCINN-CAL). TROPOMI UT NO2 typically ranges from 20-30 pptv over remote oceans to > 80 pptv over locations with intense seasonal lightning. Spatial coverage is mostly in the tropics and subtropics with FRESCO-S and extends to the midlatitudes and polar regions with ROCINN-CAL, due to its greater abundance of optically thick clouds and wider cloud top altitude range. TROPOMI UT NO2 seasonal means are spatially consistent (R = 0.6–0.8) with an existing coarser spatial resolution (5° latitude x 8° longitude) UT NO2 product from the Ozone Monitoring Instrument (OMI). UT NO2 from TROPOMI is 12–26 pptv more than that from OMI due to increase in NO2 with altitude from the OMI pressure ceiling (280 hPa) to that for TROPOMI (180 hPa), but possibly also systematic altitude differences between the TROPOMI and OMI cloud products. The TROPOMI UT NO2 product offers potential to evaluate and improve representation of UT NOx in models and supplement aircraft observations that are sporadic and susceptible to large biases in the UT

    Spectral UV measurements within the EUropean BREWer NETwork: COST Action ES1207 (2013-2017)

    Get PDF
    Presentación realizada en: European Conference on Solar UV Monitoring-ECUVM, celebrada en Viena del 12 al 14 de septiembre de 2018
    corecore