Performance Evaluation of Gas Chromatography–Atmospheric Pressure Chemical Ionization–Time-of-Flight Mass Spectrometry for Metabolic Fingerprinting and Profiling

Abstract

Gas chromatography–atmospheric-pressure chemical ionization–time-of-flight mass spectrometry (GC–APCI–TOFMS) was compared to GC × GC–electron ionization (EI)–TOFMS, GC–EI–TOFMS, GC–chemical ionization (CI)–quadrupole mass spectrometry (qMS), and GC–EI–qMS in terms of reproducibility, dynamic range, limit of detection, and quantification using a mix of 43 metabolites and 12 stable isotope-labeled standards. Lower limits of quantification for GC–APCI–TOFMS ranged between 0.06 and 7.81 μM, and relative standard deviations for calibration replicates were between 0.4% and 8.7%. For all compounds and techniques, except in four cases, R2 values were above 0.99. Regarding limits of quantification, GC–APCI–TOFMS was inferior to only GC × GC–EI–TOFMS, but outperformed all other techniques tested. GC–APCI–TOFMS was further applied to the metabolic fingerprinting of two Escherichia coli strains. Of 45 features that differed significantly (false discovery rate m ± SD below 5 mDa over m/z 190–722) mass measurements. Starting from the quasimolecular ion, six additional metabolites were identified that had not been found in a previous study using GC × GC–EI–TOFMS and an EI mass spectral library for identification purposes. Silylation adducts formed in the APCI source assisted the identification of unknown compounds, as their formation is structure-dependent and is not observed for compounds lacking a carboxylic group

Similar works

Full text

thumbnail-image

The Francis Crick Institute

redirect
Last time updated on 16/03/2018

This paper was published in The Francis Crick Institute.

Having an issue?

Is data on this page outdated, violates copyrights or anything else? Report the problem now and we will take corresponding actions after reviewing your request.

Licence: CC BY-NC 4.0