XENON10 is an experiment designed to directly detect particle dark matter. It
is a dual phase (liquid/gas) xenon time-projection chamber with 3D position
imaging. Particle interactions generate a primary scintillation signal (S1) and
ionization signal (S2), which are both functions of the deposited recoil energy
and the incident particle type. We present a new precision measurement of the
relative scintillation yield \leff and the absolute ionization yield Q_y, for
nuclear recoils in xenon. A dark matter particle is expected to deposit energy
by scattering from a xenon nucleus. Knowledge of \leff is therefore crucial for
establishing the energy threshold of the experiment; this in turn determines
the sensitivity to particle dark matter. Our \leff measurement is in agreement
with recent theoretical predictions above 15 keV nuclear recoil energy, and the
energy threshold of the measurement is 4 keV. A knowledge of the ionization
yield \Qy is necessary to establish the trigger threshold of the experiment.
The ionization yield \Qy is measured in two ways, both in agreement with
previous measurements and with a factor of 10 lower energy threshold.Comment: 8 pages, 9 figures. To be published in Nucl. Instrum. Methods