2,455 research outputs found
Lorentz Force Electrical Impedance Tomography
This article describes a method called Lorentz Force Electrical Impedance
Tomography. The electrical conductivity of biological tissues can be measured
through their sonication in a magnetic field: the vibration of the tissues
inside the field induces an electrical current by Lorentz force. This current,
detected by electrodes placed around the sample, is proportional to the
ultrasonic pressure, to the strength of the magnetic field and to the
electrical conductivity gradient along the acoustic axis. By focusing at
different places inside the sample, a map of the electrical conductivity
gradient can be established. In this study experiments were conducted on a
gelatin phantom and on a beef sample, successively placed in a 300 mT magnetic
field and sonicated with an ultrasonic transducer focused at 21 cm emitting 500
kHz bursts. Although all interfaces are not visible, in this exploratory study
a good correlation is observed between the electrical conductivity image and
the ultrasonic image. This method offers an alternative to detecting
pathologies invisible to standard ultrasonography
Les raisons de l’adoption de nouveaux outils de gestion dans les organisations, ou de l’importance des émotions.
Cette recherche vise à comprendre pourquoi une organisation adopte un nouvel outil de gestion en se centrant sur les raisons cognitives et émotionnelles à l’origine de ce phénomène. Pour ce faire, un dispositif qualitatif en deux étapes est mis en place : une étude de cas multi- site dans une logique abductive suivie d’une étude de cas confirmatoire dans une logique de réfutation poppérienne. Les résultats de cette recherche soulignent le rôle clé des émotions dans l’adoption d’un outil de gestion. Nous montrons en effet que l’élément déclencheur de toute adoption d’outil est de nature émotionnelle. Une organisation adopte un nouvel outil de gestion car il existe un décalage entre le contexte externe auquel elle est confrontée et des éléments de son contexte interne. Mais très concrètement le décisionnaire à l’origine de l’adoption déclenche celle-ci lorsqu’il perçoit des émotions de ses collaborateurs qu’il interprète comme étant défavorables à l’action organisationnelle. En adoptant un nouvel outil de gestion, le décisionnaire souhaite voir transformer ces émotions des acteurs, qu’il juge défavorables, en émotions favorables pour la conduite de la mission de l’organisation. Ces émotions ne sont d’ailleurs pas nécessairement considérées comme positives par les acteurs concernés. Nous pouvons par conséquent parler d’une volonté d’instrumentalisation des émotions par le décisionnaire à l’origine de l’adoption d’un nouvel outil de gestion, qui permettra de cristalliser des émotions défavorables et de générer des émotions favorables : nous parlons de « cristallisation émotionnelle ». En outre, ce découplage entre émotions favorables pour l’action organisationnelle et émotions éprouvées par les acteurs organisationnels nous conduit à mettre en évidence le concept d’émotion organisationnelle. Ces résultats sont ensuite discutés à la lumière des travaux portant sur les fonctions symboliques des outils de gestion mais également de ceux émanant du courant néo-institutionnel. Nous examinons en conclusion les apports théoriques et managériaux de cette recherche.Cognition; cristallisation émotionnelle; émotion organisationnelle; Outil de gestion; cas confirmatoire;
Imaging of Shear Waves Induced by Lorentz Force in Soft Tissues
This study presents the first observation of elastic shear waves generated in
soft solids using a dynamic electromagnetic field. The first and second
experiments of this 5 study showed that Lorentz force can induce a displacement
in a soft phantom and that this displacement was detectable by an ultrasound
scanner using speckle-tracking algorithms. For a 100 mT magnetic field and a 10
ms, 100 mA peak-to-peak electrical burst, the displacement reached a magnitude
of 1 um. In the third experiment, we showed that Lorentz force can induce shear
waves in a phantom. A physical model 10 using electromagnetic and elasticity
equations was proposed. Computer simulations were in good agreement with
experimental results. The shear waves induced by Lorentz force were used in the
last experiment to estimate the elasticity of a swine liver sample
Acousto-electrical speckle pattern in Lorentz force electrical impedance tomography
Ultrasound speckle is a granular texture pattern appearing in ultrasound
imaging. It can be used to distinguish tissues and identify pathologies.
Lorentz force electrical impedance tomography is an ultrasound-based medical
imaging technique of the tissue electrical conductivity. It is based on the
application of an ultrasound wave in a medium placed in a magnetic field and on
the measurement of the induced electric current due to Lorentz force. Similarly
to ultrasound imaging, we hypothesized that a speckle could be observed with
Lorentz force electrical impedance tomography imaging. In this study, we first
assessed the theoretical similarity between the measured signals in Lorentz
force electrical impedance tomography and in ultrasound imaging modalities. We
then compared experimentally the signal measured in both methods using an
acoustic and electrical impedance interface. Finally, a bovine muscle sample
was imaged using the two methods. Similar speckle patterns were observed. This
indicates the existence of an "acousto-electrical speckle" in the Lorentz force
electrical impedance tomography with spatial characteristics driven by the
acoustic parameters but due to electrical impedance inhomogeneities instead of
acoustic ones as is the case of ultrasound imaging
- …