11 research outputs found

    Evaluation of TruCount Absolute-Count Tubes for Determining CD4 and CD8 Cell Numbers in Human Immunodeficiency Virus-Positive Adults

    Get PDF
    A single-platform technology that uses an internal bead standard and three-color flow cytometry to determine CD4 and CD8 absolute counts was evaluated for reproducibility and agreement. Values obtained using TruCount absolute-count tubes were compared to those obtained using a two-color predicate methodology. Sixty specimens from human immunodeficiency virus type 1-infected donors were shipped to five laboratories. Each site also analyzed replicates of 14 human immunodeficiency virus type 1-infected local specimens at 6 h and again at 24 h. The interlaboratory variability was significantly less with TruCount (median difference in percent coefficient of variation [%CV] between the two methods was −8% and −3% for CD4 and CD8, respectively) than with the predicate method. Intralaboratory variability was smaller, with a median difference in %CV of −1% for both CD4 and CD8 with 6-h samples and −2% and −3% for CD4 and CD8, respectively, with 24-h samples. Use of TruCount for shipped samples resulted in a median CD4 count change of 7 cells (50th estimated percentile) when all laboratories and CD4 strata were combined. For on-site samples, the median CD4 count change was 10 CD4 cells for 6-h samples and 2 CD4 cells for 24-h samples. Individual site biases occurred in both directions and cancelled each other when the data were combined for all laboratories. Thus, the combined data showed a smaller change in median CD4 count than what may have occurred at an individual site. In summary, the use of TruCount decreased both the inter- and intralaboratory variability in determining absolute CD4 and CD8 counts

    Modeling flow cytometry data for cancer vaccine immune monitoring

    Get PDF
    Flow cytometry (FCM) is widely used in cancer research for diagnosis, detection of minimal residual disease, as well as immune monitoring and profiling following immunotherapy. In all these applications, the challenge is to detect extremely rare cell subsets while avoiding spurious positive events. To achieve this objective, it helps to be able to analyze FCM data using multiple markers simultaneously, since the additional information provided often helps to minimize the number of false positive and false negative events, hence increasing both sensitivity and specificity. However, with manual gating, at most two markers can be examined in a single dot plot, and a sequential strategy is often used. As the sequential strategy discards events that fall outside preceding gates at each stage, the effectiveness of the strategy is difficult to evaluate without laborious and painstaking back-gating. Model-based analysis is a promising computational technique that works using information from all marker dimensions simultaneously, and offers an alternative approach to flow analysis that can usefully complement manual gating in the design of optimal gating strategies. Results from model-based analysis will be illustrated with examples from FCM assays commonly used in cancer immunotherapy laboratories

    Multisite Comparison of CD4 and CD8 T-Lymphocyte Counting by Single- versus Multiple-Platform Methodologies: Evaluation of Beckman Coulter Flow-Count Fluorospheres and the tetraONE System

    Get PDF
    New analytic methods that permit absolute CD4 and CD8 T-cell determinations to be performed entirely on the flow cytometer have the potential for improving assay precision and accuracy. In a multisite trial, we compared two different single-platform assay methods with a predicate two-color assay in which the absolute lymphocyte count was derived by conventional hematology. A two-color method employing lymphocyte light scatter gating and Beckman Coulter Flow-Count fluorospheres for absolute counting produced within-laboratory precision equivalent to that of the two-color predicate method, as measured by coefficient of variation of replicate measurements. The fully automated Beckman Coulter tetraONE System four-color assay employing CD45 lymphocyte gating, automated analysis, and absolute counting by fluorospheres resulted in a small but significant improvement in the within-laboratory precision of CD4 and CD8 cell counts and percentages suggesting that the CD45 lymphocyte gating and automated analysis might have contributed to the improved performance. Both the two-color method employing Flow-Count fluorospheres and the four-color tetraONE System provided significant and substantial improvements in between-laboratory precision of absolute counts. In some laboratories, absolute counts obtained by the single-platform methods showed small but consistent differences relative to the predicate method. Comparison of each laboratory's absolute counts with the five-laboratory median value suggested that these differences resulted from a bias in the absolute lymphocyte count obtained from the hematology instrument in some laboratories. These results demonstrate the potential for single-platform assay methods to improve within-laboratory and between-laboratory precision of CD4 and CD8 T-cell determinations compared with conventional assay methods

    Pulmonary immunoglobulin responses to Streptococcus pneumoniae are altered but not reduced in human immunodeficiency virus-infected Malawian adults

    No full text
    We tested the hypothesis that human immunodeficiency virus (HIV)-infected adults have a specific defect in anti-pneumococcal capsular polysaccharide (Pn-specific) immunoglobulin (Ig) in fluid obtained from the lower respiratory tract. Higher levels of total IgG and IgM were present in bronchoalveolar lavage samples from HIV-infected subjects than in those from HIV-uninfected subjects. Pn-specific IgG and IgM in bronchoalveolar lavage samples were not significantly different between HIV-infected and -uninfected subjects. After pneumococcal infection, HIV-infected patients had higher bronchoalveolar lavage levels of Pn-specific IgG than HIV-infected patients without recent infection (geometric means, 387 vs. 30 ng/mL, P=.001).</p

    CD4 enumeration technologies: a systematic review of test performance for determining eligibility for antiretroviral therapy.

    Get PDF
    BACKGROUND: Measurement of CD4+ T-lymphocytes (CD4) is a crucial parameter in the management of HIV patients, particularly in determining eligibility to initiate antiretroviral treatment (ART). A number of technologies exist for CD4 enumeration, with considerable variation in cost, complexity, and operational requirements. We conducted a systematic review of the performance of technologies for CD4 enumeration. METHODS AND FINDINGS: Studies were identified by searching electronic databases MEDLINE and EMBASE using a pre-defined search strategy. Data on test accuracy and precision included bias and limits of agreement with a reference standard, and misclassification probabilities around CD4 thresholds of 200 and 350 cells/μl over a clinically relevant range. The secondary outcome measure was test imprecision, expressed as % coefficient of variation. Thirty-two studies evaluating 15 CD4 technologies were included, of which less than half presented data on bias and misclassification compared to the same reference technology. At CD4 counts 350 cells/μl, bias ranged from -70.7 to +47 cells/μl, compared to the BD FACSCount as a reference technology. Misclassification around the threshold of 350 cells/μl ranged from 1-29% for upward classification, resulting in under-treatment, and 7-68% for downward classification resulting in overtreatment. Less than half of these studies reported within laboratory precision or reproducibility of the CD4 values obtained. CONCLUSIONS: A wide range of bias and percent misclassification around treatment thresholds were reported on the CD4 enumeration technologies included in this review, with few studies reporting assay precision. The lack of standardised methodology on test evaluation, including the use of different reference standards, is a barrier to assessing relative assay performance and could hinder the introduction of new point-of-care assays in countries where they are most needed