    Evaluation of a novel quantitative canine species-specific point-of-care assay for C-reactive protein

    Get PDF
    Background: Species-specific point-of-care tests (POCT) permit a rapid analysis of canine C-reactive protein (CRP), enabling veterinarians to include CRP in clinical decisions. Aim of the study was to evaluate a novel POCT for canine CRP (Point Stripâ TM Canine CRP Assay) run on a small in-house-analyzer (Point Reader TM V) using lithium heparin plasma and to compare assay performance to an already established canine CRP assay (Gentian Canine CRP Immunoassay) run on two different bench top analyzers serving as reference methods (ABX Pentra 400, AU 5800). Linearity was assessed by stepwise dilution of plasma samples with high CRP concentrations. Limit of quantification (LoQ) was determined by repeated measurements of samples with low CRP concentrations. Coefficient of variation (CV) at low (10-50 mg/l), moderate (50-100 mg/l), and high (100-200 mg/l) CRP concentrations was investigated as well as possible interferences. Method comparison study was performed using 45 samples of healthy and diseased dogs. Quality criteria were fulfilled if the total observed error (TEobs=2CV%+bias%) was below the minimal total allowable error of 44.4% (TE min). Additionally, a reference range (n =60 healthy dogs) was established. Results: Linearity was present at CRP concentrations of 10-132 mg/l (&#8793; 361 mg/l CRP with reference method) with a LoQ set at 10 mg/l. At moderate to high CRP concentrations, intra- and inter-assay CVs were< =8% and <=11% respectively, while CVs<=22% and <=28% were present at low concentrations. No interferences were observed at concentrations of 4 g/l hemoglobin, 800 mg/l bilirubin and 8 g/l triglycerides. Method comparison study demonstrated an excellent correlation with both reference methods (r =0.98 for ABX Pentra 400; 0.99 for AU 5800), though revealing a proportional bias of 19.7% (ABX Pentra 400) and 10.7% (AU 5800) respectively. TEobs was 26.7-31.9% and 16.7-21.9% and thus < TEmin. Healthy dogs presented with CRP values <=11.9 mg/l. Conclusions The POCT precisely detects canine CRP at clinically relevant moderate and high CRP concentrations. The assay correlates well with both reference methods. Due to the bias, however, follow-up examinations should be performed with the same assay and analyzer

    Evaluation of a species-specific C-reactive protein assay for the dog on the ABX Pentra 400 clinical chemistry analyzer

    Get PDF
    Abstract Background A canine-specific immunoturbidimetric CRP assay, Gentian Canine CRP Immunoassay) with species-specific controls and calibrators was introduced and recently evaluated on the clinical chemistry analyzer Abbott Architect c4000 as well as on the Olympus AU600. Aims of our study were 1) to independently evaluate the canine-specific CRP assay on the ABX Pentra 400 clinical chemistry analyzer in comparison to the previously validated human-based immunoturbidimetric assay (Randox Canine CRP assay) and 2) to assess the impact of different sample types (serum versus heparinized plasma) on the results. Imprecision, accuracy, interference and the prozone effect were determined using samples from healthy and diseased dogs (n = 278). The Randox Canine CRP assay calibrated with canine specific control calibration material served as a reference method. Additionally, the impact of the sample type (serum and lithium heparin) was evaluated based on samples of healthy and diseased dogs (n = 49) in a second part of the study. Results Linearity was present for CRP concentrations ranging from 4 to 281 mg/l. For clinically relevant CRP concentrations of 7–281 mg/l, recovery ranged between 90 and 105% and intra- and inter-assay CVs ranged between 0.68% - 12.12% and 0.88% - 7.84%, respectively. CV was thus lower than 12.16%, i.e. the desired CV% based on biological variation. Interference was not present up to a concentration of 5 g/l hemoglobin, 800 mg/l bilirubin and 10 g/l triglycerides. No prozone effect occurred up to 676 mg/l CRP. Method comparison study revealed a Spearman’s rank correlation coefficient of rs = 0.98 and a mean constant bias of 5.2%. The sample type had a significant (P = 0.008) but clinically not relevant impact on the results (median CRP of 30.9 mg/l in lithium heparin plasma versus 31.4 mg/l in serum). Conclusions The species-specific Gentian Canine CRP Immunoassay reliably detects canine CRP on the ABX Pentra 400 clinical chemistry analyzer whereby both serum and heparin plasma can be used. The quality criteria reached on the Abbott Architect c4000 and Olympus AU600 could be met

    Evaluation of a species-specific C-reactive protein assay for the dog on the ABX Pentra 400 clinical chemistry analyzer.

    Full text link
    peer reviewed[en] BACKGROUND: A canine-specific immunoturbidimetric CRP assay, Gentian Canine CRP Immunoassay) with species-specific controls and calibrators was introduced and recently evaluated on the clinical chemistry analyzer Abbott Architect c4000 as well as on the Olympus AU600. Aims of our study were 1) to independently evaluate the canine-specific CRP assay on the ABX Pentra 400 clinical chemistry analyzer in comparison to the previously validated human-based immunoturbidimetric assay (Randox Canine CRP assay) and 2) to assess the impact of different sample types (serum versus heparinized plasma) on the results. Imprecision, accuracy, interference and the prozone effect were determined using samples from healthy and diseased dogs (n = 278). The Randox Canine CRP assay calibrated with canine specific control calibration material served as a reference method. Additionally, the impact of the sample type (serum and lithium heparin) was evaluated based on samples of healthy and diseased dogs (n = 49) in a second part of the study. RESULTS: Linearity was present for CRP concentrations ranging from 4 to 281 mg/l. For clinically relevant CRP concentrations of 7-281 mg/l, recovery ranged between 90 and 105% and intra- and inter-assay CVs ranged between 0.68% - 12.12% and 0.88% - 7.84%, respectively. CV was thus lower than 12.16%, i.e. the desired CV% based on biological variation. Interference was not present up to a concentration of 5 g/l hemoglobin, 800 mg/l bilirubin and 10 g/l triglycerides. No prozone effect occurred up to 676 mg/l CRP. Method comparison study revealed a Spearman's rank correlation coefficient of rs = 0.98 and a mean constant bias of 5.2%. The sample type had a significant (P = 0.008) but clinically not relevant impact on the results (median CRP of 30.9 mg/l in lithium heparin plasma versus 31.4 mg/l in serum). CONCLUSIONS: The species-specific Gentian Canine CRP Immunoassay reliably detects canine CRP on the ABX Pentra 400 clinical chemistry analyzer whereby both serum and heparin plasma can be used. The quality criteria reached on the Abbott Architect c4000 and Olympus AU600 could be met
    corecore