73 research outputs found

    Direct cooling of the catheter tip increases safety for CMR-guided electrophysiological procedures

    Get PDF
    <p>Abstract</p> <p>Background</p> <p>One of the safety concerns when performing electrophysiological (EP) procedures under magnetic resonance (MR) guidance is the risk of passive tissue heating due to the EP catheter being exposed to the radiofrequency (RF) field of the RF transmitting body coil. Ablation procedures that use catheters with irrigated tips are well established therapeutic options for the treatment of cardiac arrhythmias and when used in a modified mode might offer an additional system for suppressing passive catheter heating.</p> <p>Methods</p> <p>A two-step approach was chosen. Firstly, tests on passive catheter heating were performed in a 1.5 T Avanto system (Siemens Healthcare Sector, Erlangen, Germany) using a ASTM Phantom in order to determine a possible maximum temperature rise. Secondly, a phantom was designed for simulation of the interface between blood and the vascular wall. The MR-RF induced temperature rise was simulated by catheter tip heating via a standard ablation generator. Power levels from 1 to 6 W were selected. Ablation duration was 120 s with no tip irrigation during the first 60 s and irrigation at rates from 2 ml/min to 35 ml/min for the remaining 60 s (Biotronik Qiona Pump, Berlin, Germany). The temperature was measured with fluoroscopic sensors (Luxtron, Santa Barbara, CA, USA) at a distance of 0 mm, 2 mm, 4 mm, and 6 mm from the catheter tip.</p> <p>Results</p> <p>A maximum temperature rise of 22.4°C at the catheter tip was documented in the MR scanner. This temperature rise is equivalent to the heating effect of an ablator's power output of 6 W at a contact force of the weight of 90 g (0.883 N). The catheter tip irrigation was able to limit the temperature rise to less than 2°C for the majority of examined power levels, and for all examined power levels the residual temperature rise was less than 8°C.</p> <p>Conclusion</p> <p>Up to a maximum of 22.4°C, the temperature rise at the tissue surface can be entirely suppressed by using the catheter's own irrigation system. The irrigated tip system can be used to increase MR safety of EP catheters by suppressing the effects of unwanted passive catheter heating due to RF exposure from the MR scanner.</p

    Cardiac Involvement in Fabry Disease: JACC Review Topic of the Week

    Get PDF
    Fabry disease (FD) is a rare X-linked inherited lysosomal storage disorder caused by deficient α-galactosidase A activity that leads to an accumulation of globotriasylceramide (Gb3) in affected tissues, including the heart. Cardiovascular involvement usually manifests as left ventricular hypertrophy, myocardial fibrosis, heart failure, and arrhythmias, which limit quality of life and represent the most common causes of death. Following the introduction of enzyme replacement therapy, early diagnosis and treatment have become essential to slow disease progression and prevent major cardiac complications. Recent advances in the understanding of FD pathophysiology suggest that in addition to Gb3 accumulation, other mechanisms contribute to the development of Fabry cardiomyopathy. Progress in imaging techniques have improved diagnosis and staging of FD-related cardiac disease, suggesting a central role for myocardial inflammation and setting the stage for further research. In addition, with the recent approval of oral chaperone therapy and new treatment developments, the FD-specific treatment landscape is rapidly evolving

    Use of mechanical circulatory support in patients with non-ischaemic cardiogenic shock

    Get PDF
    Aims Despite its high incidence and mortality risk, there is no evidence-based treatment for non-ischaemic cardiogenic shock (CS). The aim of this study was to evaluate the use of mechanical circulatory support (MCS) for non-ischaemic CS treatment.Methods and results In this multicentre, international, retrospective study, data from 890 patients with non-ischaemic CS, defined as CS due to severe de-novo or acute-on-chronic heart failure with no need for urgent revascularization, treated with or without active MCS, were collected. The association between active MCS use and the primary endpoint of 30-day mortality was assessed in a 1:1 propensity-matched cohort. MCS was used in 386 (43%) patients. Patients treated with MCS presented with more severe CS (37% vs. 23% deteriorating CS, 30% vs. 25% in extremis CS) and had a lower left ventricular ejection fraction at baseline (21% vs. 25%). After matching, 267 patients treated with MCS were compared with 267 patients treated without MCS. In the matched cohort, MCS use was associated with a lower 30-day mortality (hazard ratio 0.76, 95% confidence interval 0.59-0.97). This finding was consistent through all tested subgroups except when CS severity was considered, indicating risk reduction especially in patients with deteriorating CS. However, complications occurred more frequently in patients with MCS; e.g. severe bleeding (16.5% vs. 6.4%) and access-site related ischaemia (6.7% vs. 0%).Conclusion In patients with non-ischaemic CS, MCS use was associated with lower 30-day mortality as compared to medical therapy only, but also with more complications. Randomized trials are needed to validate these findings.[GRAPHICS

    Estimating population size, density and dynamics of Pre-Pottery Neolithic villages in the central and southern Levant: an analysis of Beidha, southern Jordan

    Get PDF
    The Pre-Pottery Neolithic (PPN) of the central and southern Levant played an integral role in the Neolithic Demographic Transition (NDT) from mobile hunter-gatherer to village-based, agro-pastoralist societies. An understanding of population dynamics is essential for reconstructing the trajectories of these early village societies. However, few investigations have produced absolute estimates of population parameters for these villages and those which have base estimates on a limited methodological framework. This research examines the methodological and theoretical basis for existing estimates, and explores a range of methodologies in order to derive more empirically-robust demographic data. Results reveal that commonly utilized methodologies and population density coefficients employed for estimating PPN village populations require re-evaluation. This article presents the application of methodologies to the PPNB site of Beidha in southern Jordan

    Efficacy and Safety of Appropriate Shocks and Antitachycardia Pacing in Transvenous and Subcutaneous Implantable Defibrillators: Analysis of All Appropriate Therapy in the PRAETORIAN Trial.

    Get PDF
    BACKGROUND: The PRAETORIAN trial (A Prospective, Randomized Comparison of Subcutaneous and Transvenous Implantable Cardioverter Defibrillator Therapy) showed noninferiority of subcutaneous implantable cardioverter defibrillator (S-ICD) compared with transvenous implantable cardioverter defibrillator (TV-ICD) with regard to inappropriate shocks and complications. In contrast to TV-ICD, S-ICD cannot provide antitachycardia pacing for monomorphic ventricular tachycardia. This prespecified secondary analysis evaluates appropriate therapy and whether antitachycardia pacing reduces the number of appropriate shocks. METHODS: The PRAETORIAN trial was an international, investigator-initiated randomized trial that included patients with an indication for implantable cardioverter defibrillator (ICD) therapy. Patients with previous ventricular tachycardia <170 bpm or refractory recurrent monomorphic ventricular tachycardia were excluded. In 39 centers, 849 patients were randomized to receive an S-ICD (n=426) or TV-ICD (n=423) and were followed for a median of 49.1 months. ICD programming was mandated by protocol. Appropriate ICD therapy was defined as therapy for ventricular arrhythmias. Arrhythmias were classified as discrete episodes and storm episodes (≥3 episodes within 24 hours). Analyses were performed in the modified intention-to-treat population. RESULTS: In the S-ICD group, 86 of 426 patients received appropriate therapy, versus 78 of 423 patients in the TV-ICD group, during a median follow-up of 52 months (48-month Kaplan-Meier estimates 19.4% and 17.5%; P=0.45). In the S-ICD group, 83 patients received at least 1 shock, versus 57 patients in the TV-ICD group (48-month Kaplan-Meier estimates 19.2% and 11.5%; P=0.02). Patients in the S-ICD group had a total of 254 shocks, compared with 228 shocks in the TV-ICD group (P=0.68). First shock efficacy was 93.8% in the S-ICD group and 91.6% in the TV-ICD group (P=0.40). The first antitachycardia pacing attempt successfully terminated 46% of all monomorphic ventricular tachycardias, but accelerated the arrhythmia in 9.4%. Ten patients with S-ICD experienced 13 electrical storms, versus 18 patients with TV-ICD with 19 electrical storms. Patients with appropriate therapy had an almost 2-fold increased relative risk of electrical storms in the TV-ICD group compared with the S-ICD group (P=0.05). CONCLUSIONS: In this trial, no difference was observed in shock efficacy of S-ICD compared with TV-ICD. Although patients in the S-ICD group were more likely to receive an ICD shock, the total number of appropriate shocks was not different between the 2 groups. Registration: URL: https://www.clinicaltrials.gov; Unique identifier: NCT01296022

    Device-related complications in the subcutaneous and transvenous ICD: a secondary analysis of the PRAETORIAN trial.

    Get PDF
    BACKGROUND: The subcutaneous ICD (S-ICD) is developed to overcome lead-related complications and systemic infections, inherent to transvenous ICD (TV-ICD) therapy. The PRAETORIAN trial demonstrated that the S-ICD is non-inferior to the TV-ICD with regard to the combined primary endpoint of inappropriate shocks and complications. This prespecified secondary analysis evaluates all complications in the PRAETORIAN trial. METHODS: The PRAETORIAN trial is an international, multicenter, randomised trial in which 849 patients with an indication for ICD therapy were randomised to receive an SICD (N = 426) or TV-ICD (N = 423) and followed for a median of 49 months. Endpoints were device-related complications, lead-related complications, systemic infections and the need for invasive interventions. RESULTS: Thirty-six device-related complications occurred in 31 patients in the S-ICD group of which bleedings were the most frequent. In the TV-ICD group 49 complications occurred in 44 patients of which lead-dysfunction was most frequent (HR 0.69; P =0.11). In both groups half of all complications were within 30 days after implantation. Lead-related complications and systemic infections occurred significantly less in the S-ICD group compared to the TV-ICD group (P <0.001, P =0.03 respectively). Significantly more complications required invasive interventions in the TV-ICD group compared to the S-ICD group (8.3% vs. 4.3%, HR 0.59; P =0.047). CONCLUSIONS: This secondary analysis shows that, lead-related complications and systemic infections are more prevalent in the TV-ICD group compared to the S-ICD group. In addition, complications in the TV-ICD group were more severe as they required significantly more invasive interventions. This data contributes to shared decision making in clinical practice

    Canonical bases for subalgebras of factor algebras

    No full text
    We introduce canonical bases for subalgebras of quotients of the commutative and non-commutative polynomial ring. The usual theory for Gröbner bases and its counterpart for subalgebras of polynomial rings, also called SAGBI bases, are combined to obtain a tool for computation in subalgebras of factor algebras
    • …