128 research outputs found

    Acute human defibrillation performance of a subcutaneous implantable cardioverter-defibrillator with an additional coil electrode

    Get PDF
    Background: The subcutaneous implantable cardioverter-defibrillator (S-ICD) delivers 80 J shocks from an 8 cm left-parasternal coil to a 59 cm3 left lateral pulse generator (PG). A system that defibrillates with lower energy could significantly reduce PG size. Computer modeling and animal studies suggested that a second shock coil either parallel to the left-parasternal coil or transverse from the xiphoid to the PG pocket would significantly reduce the defibrillation threshold. Objective: The purpose of this study was to acutely assess the defibrillation efficacy of parallel and transverse configurations in patients receiving an S-ICD. Methods: Testing was performed in patients receiving a conventional S-ICD system. Success at 65 J was required before investigational testing. A second electrode was temporarily inserted from the xiphoid incision connected to the PG with an investigational Y-adapter. Phase 1 (n = 11) tested the parallel configuration. Phase 2 (n = 21) tested both parallel and transverse configurations in random order.Results: This study enrolled 35 patients (28 males (80%); mean age 51 ± 17 years; left ventricular ejection fraction 40% ± 15%; body mass index 26 ± 4 kg/m2; prior myocardial infarction 46%; congestive heart failure 49%; cardiomyopathy 63%). Compared to the conventional S-ICD system, mean shock impedance decreased for both parallel (69 ± 15 Ω vs 86 ± 20 Ω; n = 33; P &lt; .001) and transverse (56 ± 14 Ω vs 81 ± 21 Ω; n = 20; P &lt; .001) configurations. Shock success rates at 20, 30, and 40 J were 55%, 79%, 97%, and 25%, 70%, 90% for parallel and transverse configurations, respectively. Defibrillation threshold testing was well tolerated with no serious adverse events. Conclusion: Adding a second shock coil, particularly in the parallel configuration, significantly reduced the impedance and had a high likelihood of defibrillation success at energies ≤40 J. This may enable the development of a smaller S-ICD.</p

    Prospective blinded evaluation of a novel sensing methodology designed to reduce inappropriate shocks by the subcutaneous implantable cardioverter-defibrillator

    Get PDF
    Background: Most inappropriate shocks from the subcutaneous implantable cardioverter-defibrillator (S-ICD) are caused by cardiac oversensing. A novel sensing methodology, SMART Pass (SP; Boston Scientific Corporation, Natick, MA), aims to reduce cardiac oversensing. Objective: The purpose of this study was to evaluate the effect of SP on shocks in ambulatory patients with S-ICD. Methods: Patients implanted in 2015–2016 and enrolled in a remote patient monitoring system were included and followed for 1 year. Shocks were adjudicated by 3 independent blinded reviewers as appropriate or inappropriate. Shock incidence was calculated for patients with SP programmed enabled or disabled at implantation, censoring patients when SP programming changed or at the last transmission. The SP setting (enabled vs disabled) was modeled as a time-dependent Cox regression variable. Results: The cohort consisted of 1984 patients, and a total of 880 shocks were adjudicated. At implantation, SP was enabled in 655 patients (33%) and disabled in 1329 patients (67%). SP reduced the risk for the first inappropriate shock by 50% (P <.001) and the risk for all inappropriate shocks by 68% (P <.001) in multivariate analysis adjusted for age and device programming. The incidence of inappropriate shocks was 4.3% in the SP enabled arm vs 9.7% in the SP disabled arm. The incidence of appropriate shocks was similar (5.2

    Subcutaneous implantable cardioverter-defibrillators: long-term results of the EFFORTLESS study

    Get PDF
    AIMS: To report 5-year outcomes of EFFORTLESS registry patients with early generation subcutaneous implantable cardioverter-defibrillator (S-ICD) devices. METHODS AND RESULTS: Kaplan-Meier, trend and multivariable analyses were performed for mortality and late (years 2-5) complications, appropriate shock (AS) and inappropriate shock (IAS) rates. Nine hundred and eighty-four of 994 enrolled patients with diverse diagnoses (28% female, 48 ± 17 years, body mass index 27 ± 6 kg/m2, ejection fraction 43 ± 18%) underwent S-ICD implantation. Median follow-up was 5.1 years (interquartile range 4.7-5.5 years). All-cause mortality was 9.3% (95% confidence interval 7.2-11.3%) at 5 years; 703 patients remained in follow-up on study completion, 171 withdrew including 87 (8.8%) with device explanted, and 65 (6.6%) lost to follow-up. Of the explants, only 20 (2.0%) patients needed a transvenous device for pacing indications. First and final shock efficacy for discrete ventricular arrhythmias was consistent at 90% and 98%, respectively, with storm episode final shock efficacy at 95.2%. Time to therapy remained unaltered. Overall 1- and 5-year complication rates were 8.9% and 15.2%, respectively. Early complications did not predict later complications. There were no structural lead failures. Inappropriate shock rates at 1 and 5 years were 8.7% and 16.9%, respectively. Self-terminating inappropriately sensed episodes predicted late IAS. Predictors of late AS included self-terminating appropriately sensed episodes and earlier AS. CONCLUSION: In this diverse S-ICD registry population, spontaneous shock efficacy was consistently high over 5 years. Very few patients underwent S-ICD replacement with a transvenous device for pacing indications. Treated and self-terminating arrhythmic episodes predict future shock events, which should encourage more personalized device optimization. KEY QUESTION: Is subcutaneous implantable cardioverter-defibrillator (S-ICD) shock efficacy maintained over time? KEY FINDING: Subcutaneous implantable cardioverter-defibrillator shock efficacy remains high for discrete and storm episodes. For discrete episodes first and final shock efficacy do not change over time or by rhythm type. TAKE-HOME MESSAGE: The EFFORTLESS study 5-year results provide the longest follow up of a large patient cohort implanted with the S-ICD. For 984 patients with a median follow-up of 5.1 years, the S-ICD maintains a high level of shock efficacy of 98%

    Five-year safety and efficacy of leadless pacemakers in a Dutch cohort

    Get PDF
    BACKGROUND: Adequate real-world safety and efficacy of leadless pacemakers (LPs) have been demonstrated up to 3 years after implantation. Longer-term data are warranted to assess the net clinical benefit of leadless pacing.OBJECTIVE: The purpose of this study was to evaluate the long-term safety and efficacy of LP therapy in a real-world cohort.METHODS: In this retrospective cohort study, all consecutive patients with a first LP implantation from December 21, 2012, to December 13, 2016, in 6 Dutch high-volume centers were included. The primary safety endpoint was the rate of major procedure- or device-related complications (ie, requiring surgery) at 5-year follow-up. Analyses were performed with and without Nanostim battery advisory-related complications. The primary efficacy endpoint was the percentage of patients with a pacing capture threshold ≤2.0 V at implantation and without ≥1.5-V increase at the last follow-up visit.RESULTS: A total of 179 patients were included (mean age 79 ± 9 years), 93 (52%) with a Nanostim and 86 (48%) with a Micra VR LP. Mean follow-up duration was 44 ± 26 months. Forty-one major complications occurred, of which 7 were not advisory related. The 5-year major complication rate was 4% without advisory-related complications and 27% including advisory-related complications. No advisory-related major complications occurred a median 10 days (range 0-88 days) postimplantation. The pacing capture threshold was low in 163 of 167 patients (98%) and stable in 157 of 160 (98%).CONCLUSION: The long-term major complication rate without advisory-related complications was low with LPs. No complications occurred after the acute phase and no infections occurred, which may be a specific benefit of LPs. The performance was adequate with a stable pacing capture threshold.</p

    The learning curve associated with the introduction of the subcutaneous implantable defibrillator

    Get PDF
    Aims: The subcutaneous implantable cardioverter defibrillator (S-ICD) was introduced to overcome complications related to transvenous leads. Adoption of the S-ICD requires implanters to learn a new implantation technique. The aim of this study was to assess the learning curve for S-ICD implanters with respect to implant-related complications, procedure time, and inappropriate shocks (IASs). Methods and results: In a pooled cohort from two clinical S-ICD databases, the IDE Trial and the EFFORTLESS Registry, complications, IASs at 180 days follow-up and implant procedure duration were assessed. Patients were grouped in quartiles based on experience of the implanter and Kaplan-Meier estimates of complication and IAS rates were calculated. A total of 882 patients implanted in 61 centres by 107 implanters with a median of 4 implants (IQR 1,8) were analysed. There were a total of 59 patients with complications and 48 patients with IAS. The complication rate decreased significantly from 9.8% in Quartile 1 (least experience) to 5.4% in Quartile 4 (most experience) (P = 0.02) and non-significantly for IAS from 7.9 to 4.8% (P = 0.10). Multivariable analysis demonstrated a hazard ratio of 0.78 (P = 0.045) for complications and 1.01 (P = 0.958) for IAS. Dual-zone programming increased with experience of the individual implanter (P 13 implants). Conclusion: There is a short and significant learning curve associated with physicians adopting the S-ICD. Performance stab

    Subcutaneous or Transvenous Defibrillator Therapy.

    Get PDF
    BACKGROUND: The subcutaneous implantable cardioverter-defibrillator (ICD) was designed to avoid complications related to the transvenous ICD lead by using an entirely extrathoracic placement. Evidence comparing these systems has been based primarily on observational studies. METHODS: We conducted a noninferiority trial in which patients with an indication for an ICD but no indication for pacing were assigned to receive a subcutaneous ICD or transvenous ICD. The primary end point was the composite of device-related complications and inappropriate shocks; the noninferiority margin for the upper boundary of the 95% confidence interval for the hazard ratio (subcutaneous ICD vs. transvenous ICD) was 1.45. A superiority analysis was prespecified if noninferiority was established. Secondary end points included death and appropriate shocks. RESULTS: A total of 849 patients (426 in the subcutaneous ICD group and 423 in the transvenous ICD group) were included in the analyses. At a median follow-up of 49.1 months, a primary end-point event occurred in 68 patients in the subcutaneous ICD group and in 68 patients in the transvenous ICD group (48-month Kaplan-Meier estimated cumulative incidence, 15.1% and 15.7%, respectively; hazard ratio, 0.99; 95% confidence interval [CI], 0.71 to 1.39; P = 0.01 for noninferiority; P = 0.95 for superiority). Device-related complications occurred in 31 patients in the subcutaneous ICD group and in 44 in the transvenous ICD group (hazard ratio, 0.69; 95% CI, 0.44 to 1.09); inappropriate shocks occurred in 41 and 29 patients, respectively (hazard ratio, 1.43; 95% CI, 0.89 to 2.30). Death occurred in 83 patients in the subcutaneous ICD group and in 68 in the transvenous ICD group (hazard ratio, 1.23; 95% CI, 0.89 to 1.70); appropriate shocks occurred in 83 and 57 patients, respectively (hazard ratio, 1.52; 95% CI, 1.08 to 2.12). CONCLUSIONS: In patients with an indication for an ICD but no indication for pacing, the subcutaneous ICD was noninferior to the transvenous ICD with respect to device-related complications and inappropriate shocks. (Funded by Boston Scientific; PRAETORIAN ClinicalTrials.gov number, NCT01296022.)

    A new era for cardiac rhythm management devices: Solutions for transvenous lead complications

    Get PDF
    Both pacemakers and cardiac defibrillators (ICD) have been designed to treat cardiac arrhythmias. These arrhythmias often lead to life-threatening conditions. Numerous studies have shown the benefits on survival and quality of life of these cardiac rhythm management (CRM) devices. These devices rely on a transvenous intracardiac lead for sensing and to deliver therapy. Unfortunately the lead is the Achilles-heal of this therapy. Most complications of CRM therapy are related to implantation, malfunction or infection of these leads. Some complications even have fatal consequences. This is obviously in contradiction with one of the most important rules in medicine: First Do No Harm! This thesis describes the introduction of three new technologies designed to overcome these complications. In part I the initial experience in The Netherlands is presented with the wearable cardiac defibrillator. In part II the introduction and further widespread use of the subcutaneous ICD is discussed together with chapters on new implantation techniques and on management of inappropriate therapy. In part III the clinical experience with two different leadless pacemakers is being analyzed and in part IV successful animal experiments with a system consisting of an S-ICD communicating with a leadless pacemaker open the door to a new era for cardiac rhythm management. In this new era leads will probably not be necessary anymore to treat cardiac arrhythmias which will lead to less complications and better patient outcome
    • …