    Radiation exposure to fetus from extremity CBCT examinations

    Get PDF
    Purpose: To evaluate fetal doses from extremity CBCT examinations at different stages of pregnancy and to investigate different methods of fetal dose optimization.Method: Fetal doses were measured in an anthropomorphic phantom for two CBCT examination protocols - knee and elbow. The measurements were made at three different heights representing the three trimesters during pregnancy and three different depths in the phantom. The effect of soft tissue layer, tube voltage, add-on device shield and body angulation on fetal dose were investigated.Results: The fetal doses in clinical examination protocols were in the range of 3.4 to 6.0 mu Gy during knee ex-aminations and 2.9 to 7.7 mu Gy during elbow examinations depending on the depth of the fetus and the stage of pregnancy. A soft tissue layer representing variative body composition above abdomen region decreased the fetal dose up to 19 % in knee and up to 21 % in elbow examinations. Using lower tube voltage decreased the fetal doses up to 45 % (knee) and 51 % (elbow). An add-on device shield decreased the fetal doses up to 91 % (knee) and up to 75 % (elbow). Turning the body away from the device bore reduced the fetal doses up to 62 %. The conversion factor to convert an entrance surface dose to the fetal dose ranged from 0.4 to 0.6.Conclusions: The fetal doses from CBCT examinations of extremities are low and do not produce a concern about radiation detriment to the fetus. The most efficient way found to reduce the fetal dose was to use the add-on device shielding.Peer reviewe