35 research outputs found

    Electrochemical Enzyme Immunoassay Using Sequential Saturation Technique in A 20-μl Capillary: Digoxin as A Model Analyte

    Get PDF
    Capillary enzyme immunoassay with flow-injection analysis for digoxin using the sequential saturation technique has been developed. Glass capillary tubes (10 cm × 0.53 mm i.d.) with immobilized digoxin antibody were used as the immunoassay reactor. The product of enzymatic reaction. 4-aminophenol, was detected amperometrically. The digoxin and the labeled digoxin binding reaction with the immobilized digoxin antibody were completed in 2 and 10 min, respectively. Digoxin was determined in a 20-μl sample with a detection limit of 10 pg ml−1 (200 fg or 260 attomoles) and a 3 orders of magnitude range

    Determination of Favorite Wine from Comparison of Wine Aroma Attributes

    No full text

    Electrochemical Enzyme Immunoassay Using Sequential Saturation Technique in A 20-μl Capillary: Digoxin as A Model Analyte

    No full text
    Capillary enzyme immunoassay with flow-injection analysis for digoxin using the sequential saturation technique has been developed. Glass capillary tubes (10 cm × 0.53 mm i.d.) with immobilized digoxin antibody were used as the immunoassay reactor. The product of enzymatic reaction. 4-aminophenol, was detected amperometrically. The digoxin and the labeled digoxin binding reaction with the immobilized digoxin antibody were completed in 2 and 10 min, respectively. Digoxin was determined in a 20-μl sample with a detection limit of 10 pg ml−1 (200 fg or 260 attomoles) and a 3 orders of magnitude range
    corecore