5 research outputs found

    1Design of the Primary Prevention Parameters Evaluation (PREPARE) trial of implantablecardioverter defibrillators to reduce patient morbidity [NCT00279279]

    Get PDF
    BACKGROUND: Implantable Cardioverter Defibrillator (ICD) therapy has been proven to be beneficial and efficacious for the treatment of serious ventricular tachyarrhythmias in primary prevention patients. However, primary prevention patients appear to have a lower incidence of ventricular arrhythmias in comparison to secondary prevention patients and consequently likely experience a higher proportion of detections due to supraventricular arrhythmias. Recent trials have demonstrated that strategic and specific programming choices reduce the number of inappropriate shocks and that anti-tachycardia pacing (ATP) is an effective alternative to shock therapy for many sustained ventricular arrhythmias. METHODS: The Primary Prevention Parameters Evaluation (PREPARE) study is a multi-center cohort study, evaluating the efficacy of a pre-specified strategic profile of VT/VF detection and therapy settings in 700 primary prevention patients in an effort to safely reduce the number of shock therapies delivered. The patients, both with and without cardiac resynchronization therapy, are compared to a well-qualified set (n = 691) of historical controls derived from the MIRACLE ICD and EMPIRIC trials. This manuscript describes the design of the PREPARE study. The study results, to be presented separately, will characterize the efficacy of this programming set (PREPARE) compared with physician-tailored programming (MIRACLE ICD and EMPIRIC)

    Strategic programming of detection and therapy parameters in implantable cardioverter-defibrillators reduces shocks in primary prevention patients - Results from the PREPARE (Primary Prevention Parameters Evaluation) study

    No full text
    Objectives Our purpose was to demonstrate that strategically chosen implantable cardioverter-defibrillator (ICD) ventricular tachycardia (VT) or ventricular fibrillation (VF) detection and therapy parameters can reduce the combined incidence of device-delivered shocks, arrhythmic syncope, and untreated sustained symptomatic VT/VF (morbidity index). Background Strategically chosen ICD VT/VF detection and therapy parameters have been shown in previous studies to reduce the number of shocked episodes. In the PREPARE (Primary Prevention Parameters Evaluation) study, these prior strategies were combined with additional strategies specific to primary prevention patients. Methods The PREPARE study was a prospective, cohort-controlled study that analyzed 700 patients (biventricular [Bi-V] ICD and non-Bi-V ICD) with primary prevention indications for an ICD from 38 centers followed for 1 year. VT/VF was detected for rates >= 182 beats/min that were maintained for at least 30 of 40 beats. Antitachycardia pacing was programmed as the first therapy for regular rhythms with rates of 182 to 250 beats/min, and supraventricular tachycardia discriminators were used for rhythms Results The PREPARE programming significantly reduced the morbidity index incidence density (0.26 events/patient-year for PREPARE study patients vs. 0.69 control cohort, p = 0.003). The PREPARE study patients were less likely to receive a shock in the first year compared with control patients (9% vs. 17%, p <0.01). The incidence of untreated VT and arrhythmic syncope was similar between the PREPARE study patients and the control cohort. Conclusions Strategically chosen VT/VF detection and therapy parameters can safely reduce shocks and other morbidities associated with ICD therapy in patients receiving an ICD for primary prevention indications. (PREPARE-Primary Prevention Parameters Evaluation; NCT00279279
    corecore