85 research outputs found
After LUX: The LZ Program
The LZ program consists of two stages of direct dark matter searches using
liquid Xe detectors. The first stage will be a 1.5-3 tonne detector, while the
last stage will be a 20 tonne detector. Both devices will benefit tremendously
from research and development performed for the LUX experiment, a 350 kg liquid
Xe dark matter detector currently operating at the Sanford Underground
Laboratory. In particular, the technology used for cryogenics and electrical
feedthroughs, circulation and purification, low-background materials and
shielding techniques, electronics, calibrations, and automated control and
recovery systems are all directly scalable from LUX to the LZ detectors.
Extensive searches for potential background sources have been performed, with
an emphasis on previously undiscovered background sources that may have a
significant impact on tonne-scale detectors. The LZ detectors will probe
spin-independent interaction cross sections as low as 5E-49 cm2 for 100 GeV
WIMPs, which represents the ultimate limit for dark matter detection with
liquid xenon technology.Comment: Conference proceedings from APS DPF 2011. 9 pages, 6 figure
Signal yields, energy resolution, and recombination fluctuations in liquid xenon
This work presents an analysis of monoenergetic electronic recoil peaks in
the dark-matter-search and calibration data from the first underground science
run of the Large Underground Xenon (LUX) detector. Liquid xenon charge and
light yields for electronic recoil energies between 5.2 and 661.7 keV are
measured, as well as the energy resolution for the LUX detector at those same
energies. Additionally, there is an interpretation of existing measurements and
descriptions of electron-ion recombination fluctuations in liquid xenon as
limiting cases of a more general liquid xenon re- combination fluctuation
model. Measurements of the standard deviation of these fluctuations at
monoenergetic electronic recoil peaks exhibit a linear dependence on the number
of ions for energy deposits up to 661.7 keV, consistent with previous LUX
measurements between 2-16 keV with H. We highlight similarities in liquid
xenon recombination for electronic and nuclear recoils with a comparison of
recombination fluctuations measured with low-energy calibration data.Comment: 11 pages, 12 figures, 3 table
An Ultra-Low Background PMT for Liquid Xenon Detectors
Results are presented from radioactivity screening of two models of
photomultiplier tubes designed for use in current and future liquid xenon
experiments. The Hamamatsu 5.6 cm diameter R8778 PMT, used in the LUX dark
matter experiment, has yielded a positive detection of four common radioactive
isotopes: 238U, 232Th, 40K, and 60Co. Screening of LUX materials has rendered
backgrounds from other detector materials subdominant to the R8778
contribution. A prototype Hamamatsu 7.6 cm diameter R11410 MOD PMT has also
been screened, with benchmark isotope counts measured at <0.4 238U / <0.3 232Th
/ <8.3 40K / 2.0+-0.2 60Co mBq/PMT. This represents a large reduction, equal to
a change of \times 1/24 238U / \times 1/9 232Th / \times 1/8 40K per PMT,
between R8778 and R11410 MOD, concurrent with a doubling of the photocathode
surface area (4.5 cm to 6.4 cm diameter). 60Co measurements are comparable
between the PMTs, but can be significantly reduced in future R11410 MOD units
through further material selection. Assuming PMT activity equal to the measured
90% upper limits, Monte Carlo estimates indicate that replacement of R8778 PMTs
with R11410 MOD PMTs will change LUX PMT electron recoil background
contributions by a factor of \times1/25 after further material selection for
60Co reduction, and nuclear recoil backgrounds by a factor of \times 1/36. The
strong reduction in backgrounds below the measured R8778 levels makes the
R11410 MOD a very competitive technology for use in large-scale liquid xenon
detectors.Comment: v2 updated to include content after reviewer comments (Sep 2012
- …