Effects of body part thickness on low‐contrast detail detection and radiation dose during adult chest radiography
Introduction
Differences in patient size often provide challenges for radiographers, particularly when choosing the optimum acquisition parameters to obtain radiographs with acceptable image quality (IQ) for diagnosis. This study aimed to assess the effect of body part thickness on IQ in terms of low‐contrast detail (LCD) detection and radiation dose when undertaking adult chest radiography (CXR).
Methods
This investigation made use of a contrast detail (CD) phantom. Polymethyl methacrylate (PMMA) was utilised to approximate varied body part thicknesses (9, 11, 15 and 17 cm) simulating underweight, standard, overweight and obese patients, respectively. Different tube potentials were tested against a fixed 180 cm source to image distance (SID) and automatic exposure control (AEC). IQ was analysed using bespoke software thus providing an image quality figure inverse (IQFinv) value which represents LCD detectability. Dose area product (DAP) was utilised to represent the radiation dose.
Results
IQFinv values decreased statistically (P = 0.0001) with increasing phantom size across all tube potentials studied. The highest IQFinv values were obtained at 80 kVp for all phantom thicknesses (2.29, 2.02, 1.8 and 1.65, respectively). Radiation dose increased statistically (P = 0.0001) again with increasing phantom thicknesses.
Conclusion
Our findings demonstrate that lower tube potentials provide the highest IQFinv scores for various body part thicknesses. This is not consistent with professional practice because radiographers frequently raise the tube potential with increased part thickness. Higher tube potentials did result in radiation dose reductions. Establishing a balance between dose and IQ, which must be acceptable for diagnosis, can prevent the patient from receiving unnecessary additional radiation dose
Recommended from our members
Relationship between the visual evaluation of pathology visibility and the physical measure of low contrast detail detectability in neonatal chest radiography
Introduction: The detectability of low contrast detail (LCD) is a method used to assess image quality (IQ) in neonatal radiography; however, there is a lack of data on the relationship between LCD detectability and visual IQ. The study aims at investigating the relationship between the LCD detectability and visual IQ and pathology visibility (PV). Methods: Several acquisition parameters were employed to obtain a group of images from a neonatal Gammex chest phantom. Three observers applied relative visual grading analysis (VGA) for assessing the IQ and PV. A simulated pneumothorax visibility (PNV) and simulated hyaline membrane disease visibility (HMV) represented PV. Next, a CDRAD 2.0 phantom was radiographed utilising the same acquisition protocols, and several paired images were obtained. With the use of CDRAD analyser software, the detectability of LCD was assessed and expressed by an image quality figure inverse (IQFiinv) metric. The correlation between the IQFinv and each of IQ, PNV and HMV was examined. Results: The physical measure (IQFinv) and the visual assessment of IQ were shown to be strongly correlated (r = 0.95; p < 0.001). Using Pearson's correlation, the IQFinv, PNV, and HMV were found to be strongly correlated (r = 0.94; p < 0.001) and (r = 0.92; p < 0.001), correspondingly. Conclusion: Results of the study show that physical measures of LCD detectability utilising the CDRAD 2.0 phantom is strongly corelated with visual IQ and PV (PNV and HMV) and can be used to evaluate IQ when undertaking neonatal chest radiography (CXR). Implications for practice: This study establishes the feasibility of utilising the physical measure (IQFinv) and the CDRAD 2.0 phantom in routine quality assurance and neonatal CXR optimisation studies