    Field evaluation of a new antibody-based diagnostic for Schistosoma haematobium and S. mansoni at the point-of-care in northeast Zimbabwe

    Get PDF
    BACKGROUND: Rapid diagnostic tests (RDTs) for use at the point-of-care (POC) are likely to become increasingly useful as large-scale control programmes for schistosomiasis get underway. Given the low sensitivity of the reference standard egg count methods in detecting light infections, more sensitive tests will be required to monitor efforts aimed at eliminating schistosomiasis as advocated by the World Health Assembly Resolution 65.21 passed in 2012. METHODS: A recently developed RDT incorporating Schistosoma mansoni cercarial transformation fluid (SmCTF) for detection of anti-schistosome antibodies in human blood was here evaluated in children (mean age: 7.65 years; age range: 1-12 years) carrying light S. mansoni and S. haematobium infections in a schistosome-endemic area of Zimbabwe by comparison to standard parasitological techniques (i.e. the Kato-Katz faecal smear and urine filtration). Enzyme-linked immunosorbent assays (ELISAs) incorporating S. haematobium antigen preparations were also employed for additional comparison. RESULTS: The sensitivity of the SmCTF-RDT compared to standard parasitological methods was 100% while the specificity was 39.5%. It was found that the sera from RDT “false-positive” children showed significantly higher antibody titres in IgM-cercarial antigen preparation (CAP) and IgM-soluble egg antigen (SEA) ELISA assays than children identified by parasitology as “true-negatives”. CONCLUSIONS: Although further evaluations are necessary using more accurate reference standard tests, these results indicate that the RDT could be a useful tool for the rapid prevalence-mapping of both S. mansoni and S. haematobium in schistosome-endemic areas. It is affordable, user-friendly and allows for diagnosis of both schistosome species at the POC

    A new rapid diagnostic test for detection of anti-Schistosoma mansoni and anti-Schistosoma haematobium antibodies

    Get PDF
    Parasitological methods are widely used for the diagnosis of schistosomiasis. However, they are insensitive, particularly in areas of low endemicity, and labour-intensive. Immunoassays based on detection of anti-schistosome antibodies have the merit of high sensitivity and recently a rapid diagnostic test (RDT), incorporating Schistosoma mansoni cercarial transformation fluid (SmCTF) for detection of anti-schistosome antibodies in blood has been developed. Here, we assessed the diagnostic performance of the SmCTF-RDT for S. mansoni and S. haematobium infections by comparing it with microscopy for egg detection.; A cross-sectional survey was carried out in Azaguié, south Côte d'Ivoire. 118 pre-school-aged children submitted two stool and two urine samples, which were subjected to the Kato-Katz and urine filtration methods for the detection of S. mansoni and S. haematobium eggs, respectively. Urine was also subjected to a commercially available cassette test for S. mansoni, which detects circulating cathodic antigen. A finger-prick blood sample was used for the SmCTF-RDT for detection of anti-S. mansoni and anti-S. haematobium antibodies.; The prevalence of both anti-S. mansoni and anti-S. haematobium antibodies was more than three times higher than the prevalence of infection estimated by egg detection under a microscope. Using quadruplicate Kato-Katz as the reference standard for the diagnosis of S. mansoni infection, the sensitivity, negative predictive value (NPV), and positive predictive value (PPV) of the SmCTF-RDT was 75.0%, 84.2% and 22.5%, respectively. When two urine filtrations were considered as the reference standard for the diagnosis of S. haematobium infection, the sensitivity, NPV and PPV of SmCTF-RDT was 66.7%, 94.9% and 5.1%, respectively. The specificity of SmCTF-RDT, when using egg-detection as the reference standard, was estimated to be 34.4%. This low specificity may be a reflection of the relative insensitivity of the direct diagnostic approaches using microscopy.; The SmCTF-RDT is at least as sensitive as duplicate Kato-Katz and a single urine filtration for detection of S. mansoni and S. haematobium, respectively. Further investigations into the specificity of the test for anti-schistosome antibodies are necessary, but our results suggest that it may be a useful tool for mapping the prevalence of anti-schistosome antibodies in a given population pending intervention
    corecore