10 research outputs found

    Results after 562 total elbow replacements: A report from the Norwegian Arthroplasty Register

    Get PDF
    Background: The aim of this study was to give results of elbow arthroplasty for a relatively large population and compare different prosthesis brands and different patient subgroups. Methods: Between 1994 and 2006, 562 total elbow replacement operations were reported to the Norwegian Arthroplasty Register. Revisions of prostheses were shown using Kaplan-Meier failure curves, and risk of revision was calculated using Cox regression analysis. Results: The overall 5-and 10-year failure rates were 8% and 15%, respectively. There were only minor differences between the different implants. Patients who developed traumatic arthritis after fracture had the worst prognosis compared with inflammatory arthritis (P ¼ .005). Risk of revision was also increased when the ulnar component was inserted without cement (P ¼ .02.) Conclusions: Good results in terms of prosthesis survival were obtained with total elbow arthroplasty, although results were worse than for knee-and hip arthroplasties. The best results were achieved in patients with inflammatory arthritis. Level of evidence: Level 2; prospective cohort study

    Wear and Migration of Highly Cross-Linked and Conventional Cemented Polyethylene Cups with Cobalt Chrome or Oxinium Femoral Heads: A Randomized Radiostereometric Study of 150 Patients

    Get PDF
    ABSTRACT: This randomized study was performed to compare wear and migration of five different cemented total hip joint articulations in 150 patients. The patients received either a Charnley femoral stem with a 22.2 mm head or a Spectron EF femoral stem with a 28 mm head. The Charnley articulated with a g-sterilized Charnley Ogee acetabular cup. The Spectron EF was used with either EtOsterilized non-cross-linked polyethylene (Reflection All-Poly) or highly cross-linked (Reflection All-Poly XLPE) cups, combined with either cobalt chrome (CoCr) or Oxinium femoral heads. The patients were followed with repeated RSA measurements for 2 years. After 2 years, the EtO-sterilized non-cross-linked Reflection All-Poly cups had more than four times higher proximal penetration than its highly cross-linked counterpart. Use of Oxinium femoral heads did not affect penetration at 2 years compared to heads made of CoCr. Further follow-up is needed to evaluate the benefits, if any, of Oxinium femoral heads in the clinical setting. The Charnley Ogee was not outperformed by the more recently introduced implants in our study. We conclude that this prostheses still represents a standard against which new implants can be measured. Osteolysis induced by wear debris of ultra-high molecular weight polyethylene (UHMWPE) is considered the most common cause for failure of total hip arthroplasties (THA). 1 Sterilization technique and UHMWPE quality are related to wear. 2 Sterilization with gamma irradiation in air breaks the polymer chains and generates free radicals. Free radicals may combine with each other creating cross-links between adjacent molecules, which reduces abrasive wear. However, they entail the disadvantage of oxidative degradation of polyethylene (PE) when exposed to the body's oxidative environment, which may lead to deterioration of mechanical properties. 3 To address the problem, sterilization methods in inert environments with ethylene oxide (EtO) or gas plasma were introduced. However, these methods do not have the benefits of cross-linking, and an earlier report showed increase in early in vivo wear after EtO sterilization. 4 Highly cross-linked polyethylenes (HXLPE) provides a considerable reduction of wear compared to conventional polyethylene (PE). Concerns of PE wear have also been addressed by introducing alternative femoral head materials. Oxidized zirconium (Oxinium, Smith & Nephew) was introduced in the last decade for use in THA. Radiostereometric analysis (RSA) enables in vivo measurement of relative 3D motion in the range of 0.1 mm and 0.058. Due to its accuracy, few patients are needed to obtain satisfactory statistical power. With RSA the degree of migration during the first years after surgery correlates with the implant's longterm performance. We wanted to evaluate wear and migration patterns of the cemented highly cross-linked Reflection All-Poly XLPE cup and its non-irradiated counterpart, the cemented Reflection All-Poly cup, when articulating with either Oxinium or traditional cobalt chrome (CoCr) femoral heads (Smith & Nephew). Our null hypothesis was that wear and migration were equal to the Charnley Ogee prostheses (DePuy Intl

    Risk factors for revision after shoulder arthroplasty: 1,825 shoulder arthroplasties from the Norwegian Arthroplasty Register

    Get PDF
    Background and purpose Previous studies on shoulder arthroplasty have usually described small patient populations, and few articles have addressed the survival of shoulder implants. We describe the results of shoulder replacement in the Norwegian population (of 4.7 million) during a 12-year period. Trends in the use of shoulder arthroplasty during the study period were also investigated

    Improved survival for anatomic total shoulder prostheses: Results of 4,173 shoulder arthroplasties reported to the Norwegian Arthroplasty Register from 1994 through 2012

    Get PDF
    Background and purpose — Previously, implant survival of total shoulder prostheses was reported to be inferior to that of hemiprostheses. However, the use of total prostheses has increased in Norway due to reported good functional results. On this background, we wanted to study implant survival of 4 major shoulder prosthesis types in Norway between 1994 and 2012. Patients and methods — The study population comprised 4,173 patients with shoulder replacements reported to the Norwegian Arthroplasty Register, including 2,447 hemiprostheses (HPs), 444 anatomic total prostheses (ATPs), 454 resurfacing prostheses (RPs), and 828 reversed total prostheses (RTPs). Three time periods were compared: 1994–1999, 2000–2005, and 2006–2012. Kaplan-Meier failure curves were used to compare implant failure rates for subgroups of patients, and adjusted risks of revision were calculated using Cox regression analysis. Results — For prostheses inserted from 2006 through 2012, the 5-year survival rates were 95% for HPs (as opposed to 94% in 1994–1999), 95% for ATPs (75% in 1994–1999), 87% for RPs (96% in 1994–1999), and 93% for RTPs (91% in 1994–1999). During the study period, the implant survival improved significantly for ATPs (p < 0.001). A tendency of better results with acute fracture and worse results in sequelae after previous fractures was seen in all time periods. Interpretation — The 5-year implant survival rates were good with all prosthesis types, and markedly improved for anatomic total prostheses in the last 2 study periods. The better functional results with total shoulder prostheses than with hemiprostheses support the trend towards increased use of total shoulder prostheses

    Failure of aseptic revision total knee arthroplasties: 145 Revision failures from the Norwegian Arthroplasty Register, 1994-2011

    Get PDF
    Background and purpose — In Norway, the proportion of revision knee arthroplasties increased from 6.9% in 1994 to 8.5% in 2011. However, there is limited information on the epidemiology and causes of subsequent failure of revision knee arthroplasty. We therefore studied survival rate and determined the modes of failure of aseptic revision total knee arthroplasties. Method — This study was based on 1,016 aseptic revision total knee arthroplasties reported to the Norwegian Arthroplasty Register between 1994 and 2011. Revisions done for infections were not included. Kaplan-Meier and Cox regression analyses were used to assess the survival rate and the relative risk of re-revision with all causes of re-revision as endpoint. Results — 145 knees failed after revision total knee arthroplasty. Deep infection was the most frequent cause of re-revision (28%), followed by instability (26%), loose tibial component (17%), and pain (10%). The cumulative survival rate for revision total knee arthroplasties was 85% at 5 years, 78% at 10 years, and 71% at 15 years. Revision total knee arthroplasties with exchange of the femoral or tibial component exclusively had a higher risk of re-revision (RR = 1.7) than those with exchange of the whole prosthesis. The risk of re-revision was higher for men (RR = 2.0) and for patients aged less than 60 years (RR = 1.6). Interpretation — In terms of implant survival, revision of the whole implant was better than revision of 1 component only. Young age and male sex were risk factors for re-revision. Deep infection was the most frequent cause of failure of revision of aseptic total knee arthroplasties

    Wear and Migration of Highly Cross-Linked and Conventional Cemented Polyethylene Cups with Cobalt Chrome or Oxinium Femoral Heads: A Randomized Radiostereometric Study of 150 Patients

    No full text
    ABSTRACT: This randomized study was performed to compare wear and migration of five different cemented total hip joint articulations in 150 patients. The patients received either a Charnley femoral stem with a 22.2 mm head or a Spectron EF femoral stem with a 28 mm head. The Charnley articulated with a g-sterilized Charnley Ogee acetabular cup. The Spectron EF was used with either EtOsterilized non-cross-linked polyethylene (Reflection All-Poly) or highly cross-linked (Reflection All-Poly XLPE) cups, combined with either cobalt chrome (CoCr) or Oxinium femoral heads. The patients were followed with repeated RSA measurements for 2 years. After 2 years, the EtO-sterilized non-cross-linked Reflection All-Poly cups had more than four times higher proximal penetration than its highly cross-linked counterpart. Use of Oxinium femoral heads did not affect penetration at 2 years compared to heads made of CoCr. Further follow-up is needed to evaluate the benefits, if any, of Oxinium femoral heads in the clinical setting. The Charnley Ogee was not outperformed by the more recently introduced implants in our study. We conclude that this prostheses still represents a standard against which new implants can be measured. ß 2011 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res Keywords: highly cross-linked polyethylene; Oxinium; Charnley Ogee; radiostereometric analysis Osteolysis induced by wear debris of ultra-high molecular weight polyethylene (UHMWPE) is considered the most common cause for failure of total hip arthroplasties (THA). 1 Sterilization technique and UHMWPE quality are related to wear. 2 Sterilization with gamma irradiation in air breaks the polymer chains and generates free radicals. Free radicals may combine with each other creating cross-links between adjacent molecules, which reduces abrasive wear. However, they entail the disadvantage of oxidative degradation of polyethylene (PE) when exposed to the body&apos;s oxidative environment, which may lead to deterioration of mechanical properties. 3 To address the problem, sterilization methods in inert environments with ethylene oxide (EtO) or gas plasma were introduced. However, these methods do not have the benefits of cross-linking, and an earlier report showed increase in early in vivo wear after EtO sterilization. 4 Highly cross-linked polyethylenes (HXLPE) provides a considerable reduction of wear compared to conventional polyethylene (PE). Concerns of PE wear have also been addressed by introducing alternative femoral head materials. Oxidized zirconium (Oxinium, Smith &amp; Nephew) was introduced in the last decade for use in THA. Radiostereometric analysis (RSA) enables in vivo measurement of relative 3D motion in the range of 0.1 mm and 0.058. Due to its accuracy, few patients are needed to obtain satisfactory statistical power. With RSA the degree of migration during the first years after surgery correlates with the implant&apos;s longterm performance. We wanted to evaluate wear and migration patterns of the cemented highly cross-linked Reflection All-Poly XLPE cup and its non-irradiated counterpart, the cemented Reflection All-Poly cup, when articulating with either Oxinium or traditional cobalt chrome (CoCr) femoral heads (Smith &amp; Nephew). Our null hypothesis was that wear and migration were equal to the Charnley Ogee prostheses (DePuy Intl