Measurement of the energy resolution and calibration of hybrid pixel detectors with GaAs:Cr sensor and Timepix readout chip
This paper describes an iterative method of per-pixel energy calibration of
hybrid pixel detectors with GaAs:Cr sensor and Timepix readout chip. A
convolution of precisely measured spectra of characteristic X-rays of different
metals with the resolution and the efficiency of the pixel detector is used for
the calibration. The energy resolution of the detector is also measured during
the calibration. The use of per-pixel calibration allows to achieve a good
energy resolution of the Timepix detector with GaAs:Cr sensor: 8% and 13% at 60
keV and 20 keV, respectively
First results from the LUCID-Timepix spacecraft payload onboard the TechDemoSat-1 satellite in Low Earth Orbit
The Langton Ultimate Cosmic ray Intensity Detector (LUCID) is a payload
onboard the satellite TechDemoSat-1, used to study the radiation environment in
Low Earth Orbit (635km). LUCID operated from 2014 to 2017, collecting
over 2.1 million frames of radiation data from its five Timepix detectors on
board. LUCID is one of the first uses of the Timepix detector technology in
open space, with the data providing useful insight into the performance of this
technology in new environments. It provides high-sensitivity imaging
measurements of the mixed radiation field, with a wide dynamic range in terms
of spectral response, particle type and direction. The data has been analysed
using computing resources provided by GridPP, with a new machine learning
algorithm that uses the Tensorflow framework. This algorithm provides a new
approach to processing Medipix data, using a training set of human labelled
tracks, providing greater particle classification accuracy than other
algorithms. For managing the LUCID data, we have developed an online platform
called Timepix Analysis Platform at School (TAPAS). This provides a swift and
simple way for users to analyse data that they collect using Timepix detectors
from both LUCID and other experiments. We also present some possible future
uses of the LUCID data and Medipix detectors in space.Comment: Accepted for publication in Advances in Space Researc
Exploring transmission Kikuchi diffraction using a Timepix detector
Electron backscatter diffraction (EBSD) is a well-established scanning electron microscope (SEM)-based technique [1]. It allows the non-destructive mapping of the crystal structure, texture, crystal phase and strain with a spatial resolution of tens of nanometers. Conventionally this is performed by placing an electron sensitive screen, typically consisting of a phosphor screen combined with a charge coupled device (CCD) camera, in front of a specimen, usually tilted 70° to the normal of the exciting electron beam. Recently, a number of authors have shown that a significant increase in spatial resolution is achievable when Kikuchi diffraction patterns are acquired in transmission geometry; that is when diffraction patterns are generated by electrons transmitted through an electron-transparent, usually thinned, specimen. The resolution of this technique, called transmission Kikuchi diffraction (TKD), has been demonstrated to be better than 10 nm [2,3]. We have recently demonstrated the advantages of a direct electron detector, Timepix [4,5], for the acquisition of standard EBSD patterns [5]. In this article we will discuss the advantages of Timepix to perform TKD and for acquiring spot diffraction patterns and more generally for acquiring scanning transmission electron microscopy micrographs in the SEM. Particularly relevant for TKD, is its very compact size, which allows much more flexibility in the positioning of the detector in the SEM chamber. We will furthermore show recent results using Timepix as a virtual forward scatter detector, and will illustrate the information derivable on producing images through processing of data acquired from different areas of the detector. We will show results from samples ranging from gold nanoparticles to nitride semiconductor nanorods
Precision scans of the pixel cell response of double sided 3D pixel detectors to pion and x-ray beams
hree-dimensional (3D) silicon sensors offer potential advantages over standard planar sensors for radiation hardness in future high energy physics experiments and reduced charge-sharing for X-ray applications, but may introduce inefficiencies due to the columnar electrodes. These inefficiencies are probed by studying variations in response across a unit pixel cell in a 55μm pitch double-sided 3D pixel sensor bump bonded to TimePix and Medipix2 readout ASICs. Two complementary characterisation techniques are discussed: the first uses a custom built telescope and a 120GeV pion beam from the Super Proton Synchrotron (SPS) at CERN; the second employs a novel technique to illuminate the sensor with a micro-focused synchrotron X-ray beam at the Diamond Light Source, UK. For a pion beam incident perpendicular to the sensor plane an overall pixel efficiency of 93.0±0.5% is measured. After a 10o rotation of the device the effect of the columnar region becomes negligible and the overall efficiency rises to 99.8±0.5%. The double-sided 3D sensor shows significantly reduced charge sharing to neighbouring pixels compared to the planar device. The charge sharing results obtained from the X-ray beam study of the 3D sensor are shown to agree with a simple simulation in which charge diffusion is neglected. The devices tested are found to be compatible with having a region in which no charge is collected centred on the electrode columns and of radius 7.6±0.6μm. Charge collection above and below the columnar electrodes in the double-sided 3D sensor is observed
- …