Optimization of the linear quantification range of an online trypsin digestion coupled liquid chromatography–tandem mass spectrometry (LC–MS/MS) platform

Abstract

<p>Tandem mass spectrometry (MS/MS)-based proteomic workflows with a bottom-up approach require enzymatic digestion of proteins to peptide analytes, usually by trypsin. Online coupling of trypsin digestion of proteins, using an immobilized enzyme reactor (IMER), with liquid chromatography (LC) and MS/MS is becoming a frequently used approach. However, finding IMER digestion conditions that allow quantitative analysis of multiple proteins with wide range of endogenous concentration requires optimization of multiple interactive parameters: digestion buffer flow rate, injection volume, sample dilution, and surfactant type/concentration. In this report, we present a design of experiment approach for the optimization of an integrated IMER-LC–MS/MS platform. With bovine serum albumin as a model protein, the digestion efficacy and digestion rate were monitored based on LC–MS/MS peak area count versus protein concentration regression. The optimal parameters were determined through multivariate surface response modeling and consideration of diffusion controlled immobilized enzyme kinetics. The results may provide guidance to other users for the development of quantitative IMER-LC–MS/MS methods for other proteins.</p

Similar works

Full text

thumbnail-image

FigShare

redirect
Last time updated on 12/02/2018

This paper was published in FigShare.

Having an issue?