Prostate biopsies guided by three-dimensional real-time (4-D)
transrectal ultrasonography on a phantom: comparative study versus
two-dimensional transrectal ultrasound-guided biopsies
OBJECTIVE: This study evaluated the accuracy in localisation and distribution
of real-time three-dimensional (4-D) ultrasound-guided biopsies on a prostate
phantom. METHODS: A prostate phantom was created. A three-dimensional real-time
ultrasound system with a 5.9MHz probe was used, making it possible to see
several reconstructed orthogonal viewing planes in real time. Fourteen
operators performed biopsies first under 2-D then 4-D transurethral ultrasound
(TRUS) guidance (336 biopsies). The biopsy path was modelled using segmentation
in a 3-D ultrasonographic volume. Special software was used to visualise the
biopsy paths in a reference prostate and assess the sampled area. A comparative
study was performed to examine the accuracy of the entry points and target of
the needle. Distribution was assessed by measuring the volume sampled and a
redundancy ratio of the sampled prostate. RESULTS: A significant increase in
accuracy in hitting the target zone was identified using 4-D ultrasonography as
compared to 2-D. There was no increase in the sampled volume or improvement in
the biopsy distribution with 4-D ultrasonography as compared to 2-D.
CONCLUSION: The 4-D TRUS guidance appears to show, on a synthetic model, an
improvement in location accuracy and in the ability to reproduce a protocol.
The biopsy distribution does not seem improved