Physics goals of the STAR Experiment at RHIC in recent (and future) years
drive the need to operate the STAR TPC at ever higher luminosities, leading to
increased ionization levels in the TPC gas. The resulting ionic space charge
introduces field distortions in the detector which impact tracking performance.
Further complications arise from ionic charge leakage into the main TPC volume
from the high gain anode region. STAR has implemented corrections for these
distortions based on measures of luminosity, which we present here.
Additionally, we highlight a novel approach to applying the corrections on an
event-by-event basis applicable in conditions of rapidly varying ionization
sources.Comment: 6 pages, 7 figures, proceedings of the Workshop on Tracking in High
Multiplicity Environments (TIME 05) in Zurich, Switzerland, submitted to
Nucl. Instr. and Meth.