research

A Readout System for the STAR Time Projection Chamber

Abstract

We describe the readout electronics for the STAR Time Projection Chamber. The system is made up of 136,608 channels of waveform digitizer, each sampling 512 time samples at 6-12 Mega-samples per second. The noise level is about 1000 electrons, and the dynamic range is 800:1, allowing for good energy loss (dE/dxdE/dx) measurement for particles with energy losses up to 40 times minimum ionizing. The system is functioning well, with more than 99% of the channels working within specifications.Comment: 22 pages + 8 separate figures; 2 figures are .jpg photos to appear in Nuclear Instruments and Method