    The GraftConnector experience. Long-term patency and histological work up in an animal model.

    No full text
    BACKGROUND: A device to perform sutureless end-to-side coronary artery anastomosis has been developed by means of stent technology (GraftConnector). The present study assesses the long-term quality of the GraftConnector anastomosis in a sheep model. METHODS: In 8 adult sheep, 40-55 kg in weight, through left anterior thoracotomy, the right internal mammary artery (RIMA) was prepared and connected to the left anterior descending artery (LAD) by means of GraftConnector, on beating heart, without using any stabilizer. Ticlopidine 250 mg/day for anticoagulation for 4 weeks and Aspirin 100 mg/day for 6 months were given. The animals were sacrificed after 6 months and histological examination of anastomoses was carried out after slicing with the connector in situ for morphological analysis. RESULTS: All animals survived at 6 months. All anastomoses were patent and mean luminal width at histology was 1.8 +/- 0.2 mm; mean myotomia hyperplasia thickness was 0.21 +/- 0.1 mm. CONCLUSIONS: Long-term results demonstrate that OPCABGs performed with GraftConnector had 100% patency rate. The mean anastomotic luminal width corresponds to mean LAD's adult sheep diameter. We may speculate that myotomia hyperplasia occurred as a result of local device oversizing

    Is the GraftConnector a valid alternative to running suture in end-to-side coronary arteries anastomoses?

    No full text
    BACKGROUND: An animal study was carried out to compare long-term patency rates of coronary anastomoses performed with the GraftConnector versus running suture technique. METHODS: 10 sheep, 45 to 55 kg, underwent off-pump coronary artery bypass grafting (right internal mammary artery to left anterior descending artery). In 5 animals, the anastomosis was performed with a GraftConnector and in 5 animals with 7-0 running suture. Intraoperative fluoroscopy and a fluoroscopic control at 6 months were performed. After 6 months, the animals were sacrificed and the anastomoses were examined histologically. RESULTS: All animals survived at 6 months with 100% anastomosis patency rates in both groups. In the GraftConnector group, the anastomosis diameter at 6 months fluoroscopy was 118% of native left anterior descending artery versus 97% of the control group. Luminal anastomotic width at histology was 1.7 +/- 0.2 mm in the device group versus 1.6 +/- 0.1 mm in the control group. Mean intimal hyperplasia thickness was 0.21 +/- 0.1 mm in the device group versus 0.01 mm in the control group. CONCLUSIONS: The GraftConnector provides a consistent and reproducible coronary artery anastomosis and reduces technical demand and manual dexterity in coronary operations. Long-term results demonstrate that off-pump coronary artery bypass grafting performed with the GraftConnector had the same patency rate and luminal width as those performed with running suture
    corecore