184 research outputs found

    Charnley low-friction arthroplasty of the hip. Five to 25 years survivorship in a general hospital

    Get PDF
    <p>Abstract</p> <p>Background</p> <p>Some studies have raised the question about whether the good results obtained with the Charnley prosthesis could be replicated at general hospitals when it comes to the frequency of early complications and failure rates, both of which would be higher than those published by centres devoted to hip arthroplasties.</p> <p>Methods</p> <p>We reviewed the results of 404 Low Friction Arthroplasties of the hip implanted between 1976 and 1993 in a general hospital by general orthopaedic surgeons. For the survival analysis, the end-point chosen would be the chirurgical revision of any of the prosthetic components for whatever reason.</p> <p>Results</p> <p>The complications were 16 dislocations (4%), 14 deep infections (3.5%), 2 neurological injuries (0,5%) and 5 clinical deep venous thromboses (1.2%) (2 pulmonary embolisms). The survival rate at 25 years, both for stem and cup, was 83%. Survival was higher in those arthroplasties implanted in patients older than 60 years, with statistical significance.</p> <p>Conclusion</p> <p>Low Friction Arthroplasty undertaken at general hospitals by general orthopaedic surgeons feature similar outcomes to those found in centres devoted to hip surgery.</p

    A 28-year clinical and radiological follow-up of alumina ceramic-on-crosslinked polyethylene total hip arthroplasty: a follow-up report and analysis of the oxidation of a shelf-aged acetabular component

    Get PDF
    Aims: Our aim in this study was to describe a continuing review of 11 total hip arthroplasties using 22.225 mm Alumina ceramic femoral heads on a Charnley flanged femoral component, articulating against a silane crosslinked polyethylene. Patients and Methods: Nine patients (11 THAs) were reviewed at a mean of 27.5 years (26 to 28) post-operatively. Outcome was assessed using the d’Aubigne and Postel, and Charnley scores and penetration was recorded on radiographs. In addition, the oxidation of a 29-year-old shelf-aged acetabular component was analysed. Results: The mean clinical outcome scores remained excellent at final follow-up. The mean total penetration remained 0.41 mm (0.40 to 0.41). There was no radiographic evidence of acetabular or femoral loosening or osteolysis. There was negligible oxidation in the shelf-aged sample despite gamma irradiation and storage in air. Conclusion: These results highlight the long-term stability and durability of this type of crosslinked, antioxidant containing polyethylene when used in combination with a small diameter alumina ceramic femoral head

    Alumina-on-Polyethylene Bearing Surfaces in Total Hip Arthroplasty

    Get PDF
    The long-term durability of polyethylene lining total hip arthroplasty (THA) mainly depends on periprosthetic osteolysis due to wear particles, especially in young active patients. In hip simulator study, reports revealed significant wear reduction of the alumina ceramic-on-polyethylene articulation of THA compared with metal-on-polyethylene bearing surfaces. However, medium to long-term clinical studies of THA using the alumina ceramic-on-polyethylene are few and the reported wear rate of this articulation is variable. We reviewed the advantages and disadvantages of ceramicon- polyethylene articulation in THA, hip simulator study and retrieval study for polyethylene wear, in vivo clinical results of THA using alumina ceramic-on-polyethylene bearing surfaces in the literature, and new trial alumina ceramic-onhighly cross linked polyethylene bearing surfaces

    Population-Based Rates of Revision of Primary Total Hip Arthroplasty: A Systematic Review

    Get PDF
    Background: Most research on failure leading to revision total hip arthroplasty (THA) is reported from single centers. We searched PubMed between January 2000 and August 2010 to identify population- or community-based studies evaluating ten-year revision risks. We report ten-year revision risk using the Kaplan-Meier method, stratifying by age and fixation technique. Results: Thirteen papers met the inclusion criteria. Cemented prostheses had Kaplan-Meier estimates of revision-free implant survival of ten years ranging from 88 % to 95%; uncemented prostheses had Kaplan-Meier estimates from 80 % to 85%. Estimates ranged from 72 % to 86 % in patients less than 60 years old and from 90 to 96 % in older patients. Conclusion: Data reported from national registries suggest revision risks of 5 to 20 % ten years following primary THA. Revision risks are lower in older THA recipients. Uncemented implants may have higher ten-year rates of revision, regardless of age

    Stem diameter and rotational stability in revision total hip arthroplasty: a biomechanical analysis

    Get PDF
    BACKGROUND: Proximal femoral bone loss during revision hip arthroplasty often requires bypassing the deficient metaphyseal bone to obtain distal fixation. The purpose of this study was to determine the effect of stem diameter and length of diaphyseal contact in achieving rotational stability in revision total hip arthroplasty. METHODS: Twenty-four cadaveric femoral specimens were implanted with a fully porous-coated stem. Two different diameters were tested and the stems were implanted at multiple contact lengths without proximal bone support. Each specimen underwent torsional testing to failure and rotational micromotion was measured at the implant-bone interface. RESULTS: The larger stem diameter demonstrated a greater torsional stability for a given length of cortical contact (p ≤ 0.05). Decreasing length of diaphyseal contact length was associated with less torsional stability. Torsional resistance was inconsistent at 2 cm of depth. CONCLUSION: Larger stem diameters frequently used in revisions may be associated with less diaphyseal contact length to achieve equivalent rotational stability compared to smaller diameter stems. Furthermore, a minimum of 3 cm or 4 cm of diaphyseal contact with a porous-coated stem should be achieved in proximal femoral bone deficiency and will likely be dependent on the stem diameter utilized at the time of surgery

    Outcome and serum ion determination up to 11 years after implantation of a cemented metal-on-metal hip prosthesis

    Get PDF
    Background and purpose Little is known about the long-term outcome of cemented metal-on-metal hip arthroplasties. We evaluated a consecutive series of metal-on-metal polyethylene-backed cemented hip arthroplasties implanted in patients under 60 years of age

    Stem subsidence of polished and rough double-taper stems: In vitro mechanical effects on the cement-bone interface

    Get PDF
    Background and purpose Many clinical reports have indicated that polished hip stems show better clinical results than rough stems of the same geometry. It is still unknown, however, what the mechanical effects are of different surface finishes on the cement at the cement-bone interface. We compared mechanical effects in an in vitro cemented hip arthroplasty model
    • …
    corecore