Dynamic in vitro measurement of patellar movement after total knee arthroplasty: an in vitro study
BACKGROUND: Changing the kinematic behaviour of patellar movement could be one of the reasons for anterior knee pain after implantation of a total knee arthroplasty (TKA). The aim of the current study was to measure the potential influence on patellar kinematics of patellar resurfacing during TKA. METHODS: Patellar movement before and after TKA with and without patellar resurfacing was measured under dynamic conditions in an in vitro cadaver simulation. Physiologic Musculus quadriceps forces were applied to five physiologic human knee specimens undergoing simulated isokinetic extension motions, patellar movement was measured using an ultrasonic measurement system. Thereafter, the Interax(® )I.S.A.-prosthesis system was implanted without and with resurfacing the patella, and patellar movement was again measured. RESULTS: The physiologic patella center moved on a semilunar path up to 6.4 mm (SD 6.4 mm) medially during extension. After TKA, the unresurfaced patella showed significantly less medial translation (p = 0.04) than the resurfaced patella. Subsequent resurfacing of the patella then resulted in a return to mediolateral positioning of the patella similar to the physiological case, whereas the resurfaced patella tilted up to twice as much as physiologic. CONCLUSION: The results of this study suggest that resurfacing of the patella during TKA can result in a restoration of the physiologic mediolateral shift of the patellofemoral joint while angulation of the patella remains unphysiologic
Patient and surgery related factors associated with fatigue type polyethylene wear on 49 PCA and DURACON retrievals at autopsy and revision
<p>Abstract</p> <p>Background</p> <p>Polyethylene wear is an important factor for longevity of total knee arthroplasty. Proven and suspicious factors causing wear can be grouped as material, patient and surgery related. There are more studies correlating design and/or biomaterial factors to in vivo wear than those to patient and surgery related factors. Many retrieval studies just include revision implants and therefore may not be representative. This study is aimed to correlate patient- and surgery- related factors to visual wear score by minimizing design influence and include both autopsy and revision implants. Comparison between the groups was expected to unmask patient and surgery-related factors responsible for wear.</p> <p>Methods</p> <p>The amount of joint side wear on polyethylene retrievals was measured using a modification of an established visual wear score. Fatigue type wear was defined as summation of the most severe wear modes of delamination, pitting and cracks. Analysis of patient and surgery related variables suspicious to cause wear included prospectively sampled patient activity which was measured by self reported walking capacity. Statistical analysis was done by univariate analysis of variance. Activity level and implantation time were merged to an index of use and correlated to the wear score.</p> <p>Results</p> <p>Wear score after comparable implantation time was significantly less in the autopsy group. Even so, fatigue type wear accounted for 84 and 93 % of total wear score on autopsy and revision implants respectively. A highly significant influence on wear score was found in time of implantation (p = 0.002), level of activity (p = 0.025) and inserts belonging to revision group (p = 0.006). No influence was found for the kind of patella replacement (p = 0.483). Body mass index and accuracy of component alignment had no significant influence on visual wear score. Fatigue-type wear in the medial compartment was closely correlated to the index of use in the autopsy (R<sup>2 </sup>= 0.383) and the revision group (R<sup>2 </sup>= 0.813).</p> <p>Conclusion</p> <p>The present study's finding of substantial fatigue type wear in both autopsy and revision retrievals supports the theory that polyethylene fatigue strength is generally exceeded in this type of prosthesis. Furthermore, this study correlated fatigue-type polyethylene wear to an index of use as calculated by activity over time. Future retrieval studies may use activity over time as an important patient related factor correlated to the visual wear score. When evaluating total knee arthroplasty routine follow up, the surgeon must think of substantial wear present even without major clinical signs.</p
- …