    Analysis of the rotor temperature.

    No full text
    <p>(A) Temperature values obtained in different instruments of the spinning rotor, as measured in the iButton at 1,000 rpm after temperature equilibration, while the set point for the console temperature is 20°C (indicated as dotted vertical line). The box-and-whisker plot indicates the central 50% of the data as solid line, with the median displayed as vertical line, and individual circles for data in the upper and lower 25% percentiles. The mean and standard deviation is 19.62°C ± 0.41°C. (B) Correlation between iButton temperature and measured BSA monomer <i>s</i>-values corrected for radial magnification, scan time, scan velocity, but not viscosity (symbols). In addition to the data from the present study as shown in (A) (circles), also shown are measurements from the pilot study [<a href="http://www.plosone.org/article/info:doi/10.1371/journal.pone.0126420#pone.0126420.ref027" target="_blank">27</a>] where the same experiments were carried out on instruments not included in the present study (stars). The dotted line describes the theoretically expected temperature-dependence considering solvent viscosity.</p

    Examples of transient changes in the console temperature reading during the SV experiment, as saved in the scan file data.

    No full text
    <p>For comparison, the maximum adiabatic cooling of -0.3°C would be expected after approximately 300 sec, recovering to the equilibrium temperature after approximately 1,200 s (see Fig 3 in [<a href="http://www.plosone.org/article/info:doi/10.1371/journal.pone.0126420#pone.0126420.ref033" target="_blank">33</a>]).</p

    Histogram and box-and-whisker plot of <i>s</i>-values of the BSA monomer after different corrections: Raw experimental <i>s</i>-values (black, with grey histogram), scan time corrected <i>s</i><sub><i>t</i></sub>-values (blue), rotor temperature corrected <i>s</i><sub><i>20T</i></sub>-values (green), or radial magnification corrected <i>s</i><sub><i>r</i></sub>-values (cyan), and fully corrected <i>s</i><sub><i>20T</i>,<i>t</i>,<i>r</i>,<i>v</i></sub>-values (red with red histogram).

    No full text
    <p>The box-and-whisker plots indicate the central 50% of the data as solid line and draw the smaller and larger 25% percentiles as individual circles. The median for each group is displayed as a vertical line.</p

    Corrected best-fit apparent monomer molecular mass from integration of the <i>c</i>(<i>s</i>) peak when scanned with the absorbance system (green) and the interference system (magenta).

    No full text
    <p>Only data with rmsd less than 0.01 OD or 0.01 fringes were included. The box-and-whisker plot indicates the central 50% of the data as solid line and draws the smaller and larger 25% percentiles as individual circles. The median is displayed as a vertical line.</p

    Absence of a long-term trend in <i>s</i><sub><i>20T</i>,<i>t</i>,<i>r</i>,<i>v</i></sub>-values of the BSA monomer with time of experiment for the three kits (blue, green, and magenta).

    No full text
    <p>Highlighted as bold solid line is the overall average, and the grey area indicates one standard deviation.</p

    Observed fraction of dimer (as a ratio of dimer peak area to the sum of monomer plus dimer peak areas).

    No full text
    <p>The box-and-whisker plot indicates the central 50% of the data as solid line and draws the smaller and larger 25% percentiles as individual circles. The median displayed as vertical line. The mean and standard deviations are 18.5% ± 1.1% for the absorbance system, and 19.0% ± 2.1% for the interference system.</p

    Example for the analysis of absorbance data from the sedimentation velocity experiment of BSA.

    No full text
    <p>(A) Absorbance scans (symbols) and best-fit <i>c</i>(<i>s</i>) model at different points in time indicated by color temperature. (B and C) Bitmap and overlay of the residuals of the fit. (D) <i>c</i>(<i>s</i>) sedimentation coefficient distribution showing peaks for monomer, dimer, trimer, and traces of higher oligomers.</p

    Magnitude of the radial magnification correction obtained with the absorbance system (green) and the interference system (magenta).

    No full text
    <p>The box-and-whisker plot above the histogram indicates the central 50% of the data as solid horizontal line and draws data in the smallest and highest 25% percentiles as individual circles. The median is displayed as vertical line. The mean and standard deviations are -0.43% ±1.36% for the absorbance system, and -0.75% ± 0.82% for the interference system (once the three outliers are excluded).</p

    Correlation between the ratio of uncorrected <i>s</i>-values for the BSA monomer peak from the simultaneously acquired interference and the absorbance data (<i>s</i><sub>IF</sub>/<i>s</i><sub>ABS</sub>) and the ratio of corresponding radial magnification correction factors measured independently with the steel mask (r<sub>IF</sub>/r<sub>ABS</sub>).

    No full text
    <p>Squares are a histogram with frequency as indicated in the sidebar. The dotted line indicates the ideally expected relationship assuming perfect measurements of <i>s</i>-values and radial magnification correction factors.</p