19 research outputs found

    Development, Technical Performance, and Clinical Evaluation of a NucliSens Basic Kit Application for Detection of Enterovirus RNA in Cerebrospinal Fluid

    No full text
    The combination of nucleic acid sequence-based amplification and electrochemiluminescence detection was used to develop an internally controlled, highly sensitive and specific assay for the detection of enterovirus (EV) RNA in cerebrospinal fluid (CSF). The analytical performance of the assay was determined using both in vitro-transcribed EV RNAs and viral culture isolates. The sensitivity of the assay was 10 EV RNA copies per amplification reaction. The assay detected all enteroviral isolates tested with no cross-reactivity to 21 nonenteroviral species, including rhinovirus and parechovirus. The clinical performance of the assay was evaluated by testing 992 CSF specimens collected from adult and pediatric patients. NucliSens EV results from a subset of 327 CSF samples were compared to viral culture of nasopharyngeal specimens and rectal swabs (n = 195) and/or CSF (n = 212). Of the 212 CSF samples, 96 samples were positive by either the NucliSens EV assay (94/96; 97.9%) or culture (63/96; 65.6%), and 61/96 (63.5%) were positive by both methods. The inclusion of an EV-specific internal control monitored the entire process, including the efficiency of nucleic acid extraction, amplification, and detection. In total, only five blood-clotted CSF samples (0.5%) were inhibited. The NucliSens EV assay demonstrated superior sensitivity over viral culture (P < 0.001), excellent specificity, clear delineation of positive samples, and minimal amplification inhibition

    Development and Clinical Evaluation of an Internally Controlled, Single-Tube Multiplex Real-Time PCR Assay for Detection of Legionella pneumophila and Other Legionella Species

    No full text
    A multiplex real-time PCR assay for detection of Legionella pneumophila and Legionella spp. and including an internal control was designed. Legionella species, L. pneumophila, and the internal control were detected simultaneously by probes labeled with 6-carboxy-fluorescein, hexachlorofluorescein, and indodicarbocyanine, respectively. Therefore, no postamplification analysis was required in order to distinguish the targets. The sensitivity of both assays was 2.5 CFU/ml, and from analysis of 10 culture-positive and 74 culture-negative samples from patients investigated for legionellosis, 100% agreement was observed by both assays in comparison to culture. Four additional positives were found by the multiplex real-time PCR assay in the Legionella culture-negative samples

    Comparison and Evaluation of Real-Time PCR, Real-Time Nucleic Acid Sequence-Based Amplification, Conventional PCR, and Serology for Diagnosis of Mycoplasma pneumoniae

    No full text
    Mycoplasma pneumoniae is a common cause of community-acquired pneumonia and lower-respiratory-tract infections. Diagnosis has traditionally been obtained by serological diagnosis, but increasingly, molecular techniques have been applied. However, the number of studies actually comparing these assays is limited. The development of a novel duplex real-time PCR assay for detection of M. pneumoniae in the presence of an internal control real-time PCR is described. In addition, real-time nucleic acid sequence-based amplification (NASBA) on an iCycler apparatus is evaluated. These assays were compared to serology and a conventional PCR assay for 106 clinical samples from patients with lower-respiratory-tract infection. Of the 106 samples, 12 (11.3%) were positive by all the molecular methods whereas serology with acute sample and convalescent samples detected 6 (5.6%) and 9 (8.5%), respectively. Clinical symptoms of the patients with Mycoplasma-positive results were compared to those of the other patients with lower-respiratory-tract infections, and it was found that the results for mean lower age numbers as well as the presence of chills, increased erythrocyte sedimentation rate, and raised C-reactive protein levels showed significant differences. Molecular methods are superior for diagnosis of M. pneumoniae, providing more timely diagnosis. In addition, using real-time methods involves less hands-on time and affords the ability to monitor the reaction in the same tube