7 research outputs found
Periprocedural safety of saccular aneurysm embolization with the Penumbra SMART Coil System: a SMART registry subset analysis
Background Using data from the SMART registry, we report on periprocedural safety of the Penumbra SMART Coil System for endovascular coil embolization of saccular intracranial aneurysms.
Methods The SMART registry was a prospective, multi-center registry of site standard of care endovascular coiling procedures performed using at least 75% Penumbra SMART Coil, PC400, and/or POD coils. This subset analysis reports on the periprocedural safety outcomes of the saccular intracranial aneurysm cohort. Predictors of rupture/re-rupture or perforation (RRP), thromboembolic complications, and device- or procedure-related adverse events (AEs) were determined in univariate and multivariate analysis.
Results Between June 2016 and August 2018, 851 saccular aneurysm patients (31.0%, 264/851 ruptured) were enrolled across 66 North American centers. Clinically significant (ie, a serious adverse event) RRP occurred in 2.0% (17/851) of cases – 1.9% (5/264) for the ruptured cohort and 2.0% (12/587) for the un-ruptured cohort. Clinically significant thromboembolic events occurred in 3.1% (26/851) of cases – 5.3% (14/264) for the ruptured cohort and 2.0% (12/587) for the un-ruptured cohort. Multivariate predictors of periprocedural RRP were increased packing density and adjunctive treatment with a balloon. For periprocedural thromboembolic events, multivariate predictors were bifurcation location and ruptured status. For device- or procedure-related AEs, multivariate predictors were bifurcation location and adjunctive treatment with stent or balloon.
Conclusion The low rates of thromboembolic complications and RRP events demonstrate the adequate safety profile of the SMART Coil System to treat cerebral aneurysms in routine clinical practice
Individual lifetime benefit from low-dose colchicine in patients with chronic coronary artery disease
Aims Low-dose colchicine reduces cardiovascular risk in patients with coronary artery disease (CAD), but absolute benefits ma vary between individuals. This study aimed to assess the range of individual absolute benefits from low-dose colchicine ac cording to patient risk profile. Methods and results The European Society of Cardiology (ESC) guideline–recommended SMART-REACH model was combined with the relative treatment effect of low-dose colchicine and applied to patients with CAD from the Low-Dose Colchicine 2 (LoDoCo2) trial and the Utrecht Cardiovascular Cohort-Second Manifestations of ARTerial disease (UCC-SMART) study (n = 10 830). Individual treatment benefits were expressed as 10-year absolute risk reductions (ARRs) for myocardial infarction, stroke, or cardiovascular death (MACE), and MACE-free life-years gained. Predictions were also performed for MACE plus coronary revascularization (MACE+), using a new lifetime model derived in the REduction of Atherothrombosis for Continued Health (REACH) registry. Colchicine was compared with other ESC guideline–recommended intensified (Step 2) prevention strategies, i.e. LDL cholesterol (LDL-c) reduction to 1.4 mmol/L and systolic blood pressure (SBP) reduction to 130 mmHg. The generalizability to other populations was assessed in patients with CAD from REACH North America and Western Europe (n = 25 812). The median 10-year ARR from low-dose colchicine was 4.6% [interquartile range (IQR) 3.6–6.0%] for MACE and 8.6% (IQR 7.6–9.8%) for MACE+. Lifetime benefit was 2.0 (IQR 1.6–2.5) MACE-free years, and 3.4 (IQR 2.6–4.2) MACE+-free life-years gained. For LDL-c and SBP reduction, respectively, the median 10-year ARR for MACE was 3.0% (IQR 1.5–5.1%) and 1.7% (IQR 0.0–5.7%), and the lifetime benefit was 1.2 (IQR 0.6–2.1) and 0.7 (IQR 0.0–2.3) MACE-free life-years gained. Similar results were obtained for MACE+ and in American and European patients from REACH. Conclusion The absolute benefits of low-dose colchicine vary between individual patients with chronic CAD. They may be expected to be of at least similar magnitude to those of intensified LDL-c and SBP reduction in a majority of patients already on conventional lipid-lowering and blood pressure–lowering therapy
Recommended from our members
The SMART Registry: Long-Term Results on the Utility of the Penumbra SMART COIL System for Treatment of Intracranial Aneurysms and Other Malformations
Introduction: Penumbra SMART COIL® (SMART) System is a novel generation embolic coil with varying stiffness. The study purpose was to report real-world usage of the SMART System in patients with intracranial aneurysms (ICA) and non-aneurysm vascular lesions. Materials and Methods: The SMART Registry is a post-market, prospective, multicenter registry requiring ≥75% Penumbra Coils, including SMART, PC400, and/or POD coils. The primary efficacy endpoint was retreatment rate at 1-year and the primary safety endpoint was the procedural device-related serious adverse event rate. Results: Between June 2016 and August 2018, 995 patients (mean age 59.6 years, 72.1% female) were enrolled at 68 sites in the U.S. and Canada. Target lesions were intracranial aneurysms in 91.0% of patients; 63.5% were wide-neck and 31.8% were ruptured. Adjunctive devices were used in 55.2% of patients. Mean packing density was 32.3%. Procedural device-related serious adverse events occurred in 2.6% of patients. The rate of immediate post-procedure adequate occlusion was 97.1% in aneurysms and the rate of complete occlusion was 85.2% in non-aneurysms. At 1-year, the retreatment rate was 6.8%, Raymond Roy Occlusion Classification (RROC) I or II was 90.0% for aneurysms, and Modified Rankin Scale (mRS) 0-2 was achieved in 83.1% of all patients. Predictors of 1-year for RROC III or retreatment (incomplete occlusion) were rupture status (P < 0.0001), balloon-assisted coiling (P = 0.0354), aneurysm size (P = 0.0071), and RROC III immediate post-procedure (P = 0.0086) in a model that also included bifurcation aneurysm (P = 0.7788). Predictors of aneurysm retreatment at 1-year was rupture status (P < 0.0001). Conclusions: Lesions treated with SMART System coils achieved low long-term retreatment rates. Clinical Trial Registration: https://www.clinicaltrials.gov/, identifier NCT02729740. © Copyright © 2021 Spiotta, Park, Bellon, Bohnstedt, Yoo, Schirmer, DeLeacy, Fiorella, Woodward, Hawk, Nanda, Zaidat, Sunenshine, Liu, Kabbani, Snyder, Sivapatham, Dumont, Reeves, Starke and the SMART Registry Investigators.Open access journalThis item from the UA Faculty Publications collection is made available by the University of Arizona with support from the University of Arizona Libraries. If you have questions, please contact us at [email protected]
Identification of vascular patients at very high risk for recurrent cardiovascular events : Validation of the current ACC/AHA very high risk criteria
Aims: To validate and assess performance of the current ACC/AHA very high risk criteria in patients with clinically manifest arterial disease. Methods and results: Data were used from the SMART study (n = 7216) and REACH Registry (n = 48 322), two prospective cohorts of patients with manifest atherosclerotic arterial disease. Prevalence and incidence rates of recurrent major adverse cardiovascular events (MACE) were calculated, according to the ACC/AHA VHR criteria (cardiovascular disease combined with diabetes, smoking, dyslipidaemia, and/or recent recurrent coronary events). Performance of the ACC/AHA criteria was compared with single very high risk factors in terms of C-statistics and Net Reclassification Index. All patients were at VHR according to the ESC guidelines (incidence of recurrent MACE in SMART was 2.4/100PY, with 95% CI 2.3-2.5/100PY and in REACH 5.1/100PY with 95% CI 5.0-5.3/100PY). In SMART 57% of the patients were at VHR according to the ACC/AHA criteria (incidence of recurrent MACE 2.7/100PY, 95% CI 2.5-2.9/100PY) and in REACH this was 64% (5.9/100PY, 95% CI 5.7-6.1/100PY). The C-statistic for the ACC/AHA VHR criteria was 0.53 in REACH and 0.54 in SMART. Very high risk factors with comparable or slightly better performance were eGFR 70 years. Around two third of the patients meeting the ACC/AHA VHR criteria had a predicted 10-year risk of recurrent MACE 70 years performs as well as the ACC/AHA VHR criteria
Individual lifetime benefit from low-dose colchicine in patients with chronic coronary artery disease
BACKGROUND AND AIMS: Low-dose colchicine reduces cardiovascular risk in patients with coronary artery disease (CAD), but absolute benefits may vary between individuals. This study aimed to assess the range of absolute benefit from low-dose colchicine according to individual patient risk profile. METHODS: The ESC guideline-recommended SMART-REACH model was combined with the relative treatment effect of low-dose colchicine, and applied to CAD patients from the LoDoCo2 trial and UCC-SMART cohort (n = 10,830). Individual treatment benefit was expressed as 10-year absolute risk reductions (ARRs) for myocardial infarction, stroke, or cardiovascular death (MACE), and MACE-free life-years gained. Predictions were also performed for MACE plus coronary revascularization (MACE+), using a new lifetime model derived in the REACH registry. Colchicine was compared to other ESC guideline-recommended intensified (step 2) prevention strategies, i.e. low density lipoprotein-cholesterol (LDL-c) reduction to 1.4 mmol/L, and systolic blood pressure (SBP) reduction to 130 mmHg. Generalizability to other populations was assessed in CAD patients from REACH North America and Western Europe (n = 25,812). RESULTS: Median 10-year ARR from low-dose colchicine was 4.6% (interquartile range [IQR] 3.6-6.0%) for MACE, and 8.6% (IQR 7.6-9.8%) for MACE + . Lifetime benefit was 2.0 (IQR 1.6-2.5) MACE-free years, and 3.4 (IQR 2.6-4.2) MACE + -free life-years gained. For LDL-c and SBP reduction respectively, median 10-year ARR for MACE was 3.0% (IQR 1.5-5.1%) and 1.7% (IQR 0.0-5.7%), and lifetime benefit was 1.2 (IQR 0.6-2.1) and 0.7 (IQR 0.0-2.3) MACE-free life-years gained. Similar results were obtained for MACE+, and in American and European patients from REACH. CONCLUSIONS: The absolute benefits of low-dose colchicine vary between individual patients with chronic CAD. They may be expected to be of at least similar magnitude to those of intensified LDL-c and SBP reduction in a majority of patients already on conventional lipid-lowering and blood pressure-lowering therapy
The SMART Registry: Long-Term Results on the Utility of the Penumbra SMART COIL System for Treatment of Intracranial Aneurysms and Other Malformations
Introduction: Penumbra SMART COIL® (SMART) System is a novel generation embolic coil with varying stiffness. The study purpose was to report real-world usage of the SMART System in patients with intracranial aneurysms (ICA) and non-aneurysm vascular lesions.
Materials and Methods: The SMART Registry is a post-market, prospective, multicenter registry requiring ≥75% Penumbra Coils, including SMART, PC400, and/or POD coils. The primary efficacy endpoint was retreatment rate at 1-year and the primary safety endpoint was the procedural device-related serious adverse event rate.
Results: Between June 2016 and August 2018, 995 patients (mean age 59.6 years, 72.1% female) were enrolled at 68 sites in the U.S. and Canada. Target lesions were intracranial aneurysms in 91.0% of patients; 63.5% were wide-neck and 31.8% were ruptured. Adjunctive devices were used in 55.2% of patients. Mean packing density was 32.3%. Procedural device-related serious adverse events occurred in 2.6% of patients. The rate of immediate post-procedure adequate occlusion was 97.1% in aneurysms and the rate of complete occlusion was 85.2% in non-aneurysms. At 1-year, the retreatment rate was 6.8%, Raymond Roy Occlusion Classification (RROC) I or II was 90.0% for aneurysms, and Modified Rankin Scale (mRS) 0-2 was achieved in 83.1% of all patients. Predictors of 1-year for RROC III or retreatment (incomplete occlusion) were rupture status (P < 0.0001), balloon-assisted coiling (P = 0.0354), aneurysm size (P = 0.0071), and RROC III immediate post-procedure (P = 0.0086) in a model that also included bifurcation aneurysm (P = 0.7788). Predictors of aneurysm retreatment at 1-year was rupture status (P < 0.0001).
Conclusions: Lesions treated with SMART System coils achieved low long-term retreatment rates
The SMART Registry: Long-Term Results on the Utility of the Penumbra SMART COIL System for Treatment of Intracranial Aneurysms and Other Malformations
Introduction:
Penumbra SMART COIL® (SMART) System is a novel generation embolic coil with varying stiffness. The study purpose was to report real-world usage of the SMART System in patients with intracranial aneurysms (ICA) and non-aneurysm vascular lesions.
Materials and Methods:
The SMART Registry is a post-market, prospective, multicenter registry requiring ≥75% Penumbra Coils, including SMART, PC400, and/or POD coils. The primary efficacy endpoint was retreatment rate at 1-year and the primary safety endpoint was the procedural device-related serious adverse event rate.
Results:
Between June 2016 and August 2018, 995 patients (mean age 59.6 years, 72.1% female) were enrolled at 68 sites in the U.S. and Canada. Target lesions were intracranial aneurysms in 91.0% of patients; 63.5% were wide-neck and 31.8% were ruptured. Adjunctive devices were used in 55.2% of patients. Mean packing density was 32.3%. Procedural device-related serious adverse events occurred in 2.6% of patients. The rate of immediate post-procedure adequate occlusion was 97.1% in aneurysms and the rate of complete occlusion was 85.2% in non-aneurysms. At 1-year, the retreatment rate was 6.8%, Raymond Roy Occlusion Classification (RROC) I or II was 90.0% for aneurysms, and Modified Rankin Scale (mRS) 0-2 was achieved in 83.1% of all patients. Predictors of 1-year for RROC III or retreatment (incomplete occlusion) were rupture status (
P
< 0.0001), balloon-assisted coiling (
P
= 0.0354), aneurysm size (
P
= 0.0071), and RROC III immediate post-procedure (
P
= 0.0086) in a model that also included bifurcation aneurysm (
P
= 0.7788). Predictors of aneurysm retreatment at 1-year was rupture status (
P
< 0.0001).
Conclusions:
Lesions treated with SMART System coils achieved low long-term retreatment rates.
Clinical Trial Registration:
https://www.clinicaltrials.gov/